A rare case of endotracheal tube cuff leakage with no detectable decrease in cuff pressure

被引:0
|
作者
Nagasawa, Keiichi [1 ]
Nishibata, Masayuki [1 ]
Luthe, Sarah Kyuragi [1 ]
Kawamata, Tomoyuki [1 ]
机构
[1] Wakayama Med Univ, Dept Anesthesiol, Sch Med, Wakayama 6418510, Japan
来源
JA CLINICAL REPORTS | 2024年 / 10卷 / 01期
关键词
Endotracheal tube; Cuff leak; Pilot balloon tubing; Fiber-optic light; Surgical fire; Ignition source;
D O I
10.1186/s40981-024-00754-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundCommon causes of air leakage around an endotracheal tube include insufficient endotracheal tube cuff inflation and damage to the cuff, while damage to the pilot balloon or pilot balloon tubing is relatively rare.Case presentationA 74-year-old female with vertebral osteomyelitis was scheduled for an extreme lateral interbody fusion followed posterior fixation. A fiber-optic light was utilized as part of the surgical illuminator. A sudden decrease in tidal volume and airway pressure was noted intraoperatively. We suspected leakage around the endotracheal tube cuff; however, no decrease in cuff pressure was detected. Despite the normal cuff pressure, we decided to inject a small amount of air which led to a significant increase in the cuff pressure. Upon careful inspection of the endotracheal tube, we discovered that the pilot balloon tubing was damaged as a result of thermal energy emitted by the fiber-optic light, which had ignited the surgical drape. The pilot balloon tubing was partially severed in which the section proximal to the endotracheal tube cuff was burned and punctured, causing the cuff leak. Meanwhile, the section proximal to the pilot balloon had melted and occluded the lumen, resulting in a falsely normal cuff pressure reading followed by an elevated cuff pressure when a small amount of air was injected into the pilot balloon during troubleshooting. Appropriate ventilation was resumed after extubation and re-intubation with a new endotracheal tube.ConclusionsWe experienced an endotracheal tube cuff leakage caused by a damaged pilot balloon tubing due to thermal energy of the fiber-optic light. Our case report emphasizes the importance of suspecting damage to the endotracheal tube cuff and inflation system despite a normal cuff pressure reading, given that the measurement may be falsely elevated depending on the specific location of the damage. In addition, all operating personnel should be familiarized with safety warnings and cautions related to handling.
引用
收藏
页数:4
相关论文
共 50 条
  • [21] ENDOTRACHEAL TUBE CUFF DISPLACEMENT
    COPPEN, JE
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1967, 200 (05): : 420 - &
  • [22] Elongated cuff-endotracheal tube to decrease the chance of pulmonary aspiration
    Dabbagh, Ali
    Beihaghi, Hadi
    JOURNAL OF RESEARCH IN MEDICAL SCIENCES, 2008, 13 (04): : 222 - 222
  • [23] AN UNUSUAL CASE OF ENDOTRACHEAL-TUBE CUFF DYSFUNCTION
    BEVACQUA, BK
    CLEARY, WF
    JOURNAL OF CLINICAL ANESTHESIA, 1993, 5 (03) : 237 - 239
  • [24] Monitoring endotracheal tube cuff pressure using a blood pressure manometer
    Bloria, Summit
    Chauhan, Rajeev
    Luthra, Ankur
    Bloria, Pallavi
    Kataria, Ketan
    JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2022, 38 (02) : 318 - 319
  • [25] Evaluation of Endotracheal Tube Cuff Pressure and the Use of Three Cuff Inflation Syringe Devices in Dogs
    Hung, Wan-Chu
    Ko, Jeff C.
    Weil, Ann B.
    Weng, Hsin-Yi
    FRONTIERS IN VETERINARY SCIENCE, 2020, 7
  • [26] Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure
    Guyton, DC
    Barlow, MR
    Besselievre, TR
    CRITICAL CARE MEDICINE, 1997, 25 (01) : 91 - 94
  • [27] The change of endotracheal tube cuff pressure during laparoscopic surgery
    Kwon, Youngsuk
    Jang, Ji Su
    Hwang, Sung Mi
    Lee, Jae Jun
    Hong, Seok Jun
    Hong, Sung Jun
    Kang, Byung Yong
    Lee, Ho Seok
    OPEN MEDICINE, 2019, 14 (01): : 431 - 436
  • [28] TRACHEOMALACIA FROM EXCESSIVE CUFF PRESSURE OF AN ENDOTRACHEAL-TUBE
    SILVA, LU
    WOOD, GJ
    KING FAISAL SPECIALIST HOSPITAL MEDICAL JOURNAL, 1984, 4 (02): : 201 - 203
  • [29] A WIRELESS SENSOR NETWORK FOR ENDOTRACHEAL TUBE CUFF PRESSURE MONITORING
    Cipolletti, M.
    Martin Pirchio, F. N.
    Sanudo, S.
    Julian, P.
    Villalba, W.
    Masson, F.
    Mandolesi, P. S.
    LATIN AMERICAN APPLIED RESEARCH, 2009, 39 (01) : 27 - 32
  • [30] Negative intracuff pressure inside the deflated cuff of an endotracheal tube
    Kumar, Sanjeev
    Agarwal, Rishabh
    INDIAN JOURNAL OF ANAESTHESIA, 2023, 67 (03) : 315 - 317