Measuring system of absorbed dose standard for electron beam radiation processing

被引:0
|
作者
Zhang, Limin [1 ]
Gu, Jiqing [1 ]
Xu, Peiling [1 ]
Bai, Yihua [1 ]
Ye, Dongsheng [1 ]
Wu, Bin [1 ]
机构
[1] Shanghai Inst of Metrological, Technology, Shanghai, China
来源
He Jishu/Nuclear Techniques | 1995年 / 18卷 / 04期
关键词
Absorption - Dosimetry - Electron beams - Measurements - Processing - Radiation;
D O I
暂无
中图分类号
学科分类号
摘要
The design and experiments of an absorbed dose calorimeter in quasi-adiabatic mode of operation and its measuring system were described. The absorption disk and outer shell of the calorimeter were made of anisotropic pyrolytic graphite. The graphite itself was used as heating resistor of the electrical heating unit, for the purpose of check and simulation. The ratio of calculated value and experimental one for temperature rising is 1.008. The standard deviation is less than 0.6%. The total uncertainty of absorbed dose determination with the graphite calorimeter approximates to 2.7% (K=2) for the 0.5-3.0 MeV electron beam in the dose range of 1-5×102kGy. the powder alanine/ESR dosimeter and plastic thin-film dosimeter can be calibrated by the calorimeter. The passivation of temperature rising curve, heat equilibrium, measurement of calorimetric ratio and its error were also discussed.
引用
收藏
页码:238 / 246
相关论文
共 50 条
  • [31] INTERCOMPARISONS OF ABSORBED DOSE DETERMINATIONS IN 10 TO 35 MEV ELECTRON RADIATION
    SVENSSON, H
    HETTINGER, G
    FROST, D
    POHLIT, W
    ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1971, 10 (01): : 56 - +
  • [32] FILM ABSORBED-DOSE DETECTOR OF PHOTON AND ELECTRON-RADIATION
    GENERALOVA, VV
    PONOMAREV, AN
    MEASUREMENT TECHNIQUES, 1979, 22 (07) : 859 - 861
  • [33] ELECTRON-BEAM DOSIMETERS FOR RADIATION PROCESSING
    THALACKER, VP
    SIMPSON, JT
    POSTMA, NB
    RADIATION PHYSICS AND CHEMISTRY, 1988, 31 (4-6): : 473 - 479
  • [34] Stability of a water calorimetry system as a primary standard for absorbed dose to water
    Cojocaru, Claudiu
    Muir, Bryan
    McEwen, Malcolm
    Ross, Carl
    Klassen, Norman
    Marchington, David
    MEDICAL PHYSICS, 2016, 43 (08) : 4960 - 4961
  • [35] The electron dose monitoring and control system in EB radiation processing for wires and cables
    Zhou, XZ
    Zhou, Y
    Zhou, YY
    Tang, Q
    RADIATION PHYSICS AND CHEMISTRY, 2002, 63 (3-6) : 267 - 270
  • [36] CONCERNING COMPARISONS OF ELECTRON-BEAM ABSORBED DOSE MEASUREMENTS IN WATER AND POLYSTYRENE
    MATTSSON, LO
    NAHUM, AE
    MEDICAL PHYSICS, 1984, 11 (01) : 85 - 86
  • [37] GRAPHITE CALORIMETER FOR ABSORBED DOSE STANDARDIZATION OF GAMMA-RADIATION PROCESSING PLANTS
    YOUSSEF, SK
    RAO, ISS
    VOHRA, KG
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1981, 32 (12): : 869 - 875
  • [38] Calculation of absorbed dose of low-energy electron beam by an approximate method
    Vázquez-Polo, G
    Valdivia, HL
    Abrego, HC
    Mijangos, RR
    García, R
    RADIATION PHYSICS AND CHEMISTRY, 2002, 64 (03) : 181 - 187
  • [39] ABSORBED RADIATION-DOSE IN RADIATION SYNOVECTOMY
    BOWRING, CS
    KEELING, DH
    BRITISH JOURNAL OF RADIOLOGY, 1978, 51 (610): : 836 - 837
  • [40] GOVERNMENTAL PRIMARY STANDARD OF UNIT OF POWER OF ABSORBED DOSE OF PHOTON IONIZING-RADIATION
    BREGADZE, YI
    ISAEV, BM
    TULTAEV, AV
    SILCHENK.AP
    KORSHIKOV, AV
    MEASUREMENT TECHNIQUES, 1973, 16 (03) : 313 - 318