Measuring system of absorbed dose standard for electron beam radiation processing

被引:0
|
作者
Zhang, Limin [1 ]
Gu, Jiqing [1 ]
Xu, Peiling [1 ]
Bai, Yihua [1 ]
Ye, Dongsheng [1 ]
Wu, Bin [1 ]
机构
[1] Shanghai Inst of Metrological, Technology, Shanghai, China
来源
He Jishu/Nuclear Techniques | 1995年 / 18卷 / 04期
关键词
Absorption - Dosimetry - Electron beams - Measurements - Processing - Radiation;
D O I
暂无
中图分类号
学科分类号
摘要
The design and experiments of an absorbed dose calorimeter in quasi-adiabatic mode of operation and its measuring system were described. The absorption disk and outer shell of the calorimeter were made of anisotropic pyrolytic graphite. The graphite itself was used as heating resistor of the electrical heating unit, for the purpose of check and simulation. The ratio of calculated value and experimental one for temperature rising is 1.008. The standard deviation is less than 0.6%. The total uncertainty of absorbed dose determination with the graphite calorimeter approximates to 2.7% (K=2) for the 0.5-3.0 MeV electron beam in the dose range of 1-5×102kGy. the powder alanine/ESR dosimeter and plastic thin-film dosimeter can be calibrated by the calorimeter. The passivation of temperature rising curve, heat equilibrium, measurement of calorimetric ratio and its error were also discussed.
引用
收藏
页码:238 / 246
相关论文
共 50 条
  • [1] Calorimeters for absorbed dose standard of electron beam radiation processing
    Zong, YD
    Zhang, YL
    Gao, JC
    Liu, ZM
    Xia, X
    Li, CH
    Xie, LQ
    RADIATION PHYSICS AND CHEMISTRY, 1998, 53 (05): : 549 - 553
  • [2] MEASUREMENTS AND CALCULATIONS OF ABSORBED DOSE IN ELECTRON-BEAM RADIATION PROCESSING
    ZHOU, Y
    ZHOU, YY
    LUO, ZM
    RADIATION PHYSICS AND CHEMISTRY, 1993, 42 (4-6): : 817 - 820
  • [3] SIMPLE METHOD OF EVALUATING ABSORBED DOSE IN ELECTRON-BEAM PROCESSING
    TABATA, T
    ITO, R
    KURIYAMA, I
    MORIUCHI, Y
    RADIATION PHYSICS AND CHEMISTRY, 1989, 33 (05) : 411 - 416
  • [4] POWDER STATE ALANINE ESR DOSIMETER FOR MEASURING THE ABSORBED DOSE OF ELECTRON-BEAM
    BAI, YH
    YE, X
    ZHANG, LM
    ZHANG, WH
    RADIATION PHYSICS AND CHEMISTRY, 1993, 42 (4-6): : 821 - 822
  • [5] COMPUTATION OF DISTRIBUTION OF ABSORBED DOSE AND ABSORBED DOSE RATE FROM A SCANNING ELECTRON BEAM
    ROZENFEL.M
    LANZL, LH
    NEWTON, CM
    SKAGGS, LS
    PHYSICS IN MEDICINE AND BIOLOGY, 1970, 15 (01): : 140 - &
  • [6] COMPUTATION OF DISTRIBUTION OF ABSORBED DOSE AND ABSORBED DOSE RATE FROM A SCANNING ELECTRON BEAM
    ROZENFELD, M
    LANZL, LH
    NEWTON, CM
    SKAGGS, LS
    STRAHLENTHERAPIE, 1969, 138 (06) : 651 - +
  • [7] Graphite calorimeter for measurement of absorbed dose of electron beam
    Chen, Ke-Sheng
    Lin, Min
    Xu, Li-Jun
    Ye, Hong-Sheng
    Chen, Yi-Zhen
    Xia, Wen
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2009, 43 (02): : 113 - 116
  • [8] SYSTEM FOR DETERMINATION OF ABSORBED DOSE AND ABSORBED DOSE DISTRIBUTION FOR HIGH-ENERGY ELECTRON-RADIATION - IONIZATION AND CHARGE MEASUREMENT
    HEGEWALD, H
    WULFF, W
    RADIOBIOLOGIA-RADIOTHERAPIA, 1977, 18 (04) : 477 - 482
  • [9] DEVELOPMENT OF A METHOD OF ABSORBED DOSE ON-LINE MONITORING AT PRODUCT PROCESSING BY SCANNED ELECTRON BEAM
    Pomatsalyuk, R. I.
    Shevchenko, V. A.
    Tenishev, A. Eh.
    Titov, D. V.
    Uvarov, V. L.
    Zakharchenko, A. A.
    PROBLEMS OF ATOMIC SCIENCE AND TECHNOLOGY, 2016, (03): : 149 - 153
  • [10] STATE STANDARD FOR UNIT OF ABSORBED DOSE OF BETA AND NEUTRON RADIATION
    YUDIN, MF
    URYAEV, IA
    YANOVSKII, AP
    MEASUREMENT TECHNIQUES, 1972, 15 (12) : 1772 - 1776