Factors affecting the accuracy of TMA measurements

被引:0
|
作者
Foreman, Jonathon
Kelsey, Mark
Widmann, Georg
机构
[1] Mettler-Toledo, Inc., Columbus, OH 43240, United States
[2] Mettler-Toledo GmbH, CH-8603 Schwerzenbach, Switzerland
关键词
D O I
暂无
中图分类号
学科分类号
摘要
This study investigates factors that cause differences between reported temperatures in thermomechanical analysis (TMA) instruments and actual sample temperatures. Observations of the melting of indium are made under several sets of conditions. These include varying the indium specimen size, altering the position of the thermal sensor, changing the heating rate, and placing pieces of indium atop and below glass and copper specimens of varying thickness. A TMA instrument with direct-contact thermal sensor and single differential thermal analysis capability yields better sample temperature measurements and allows direct examination of the effects of heat capacity. In a second set of experiments, the glass transition of polycarbonate film is measured in the tension mode using different forces and specimen lengths. The results of this study show that low material thermal conductivity and high specimen heat capacity cause the sample temperature to be lower than that of the surrounding atmosphere. Temperature gradients along the length of the furnace can intensify the temperature difference. Furthermore, stress applied to specimens in tension can lead to erroneous Tg values. Specimen and calibration procedural changes are recommended.
引用
收藏
页码:181 / 196
相关论文
共 50 条
  • [41] FACTORS AFFECTING THE ACCURACY OF BBT FOR OVULATION DETECTION
    BARSOUM, D
    CONTRACEPTIVE DELIVERY SYSTEMS, 1984, 5 (03) : 43 - 43
  • [42] The Factors Affecting Positioning Accuracy of Geared Servodrives
    Slapak, Viktor
    Pajkos, Michal
    Hric, Matus
    2019 19TH INTERNATIONAL CONFERENCE ON ELECTRICAL DRIVES & POWER ELECTRONICS (EDPE), 2019, : 155 - 159
  • [43] Factors affecting the accuracy of pregnancy scanning in ewes
    Fridlund, C.
    Humblot, P.
    Bage, R.
    Soderquist, L.
    VETERINARY RECORD, 2013, 173 (24) : 606 - +
  • [44] FACTORS AFFECTING ACCURACY OF PREGNANCY DIAGNOSIS IN CATTLE
    GOWAN, EW
    ETCHES, RJ
    BRYDEN, C
    KING, GJ
    JOURNAL OF DAIRY SCIENCE, 1982, 65 (07) : 1294 - 1302
  • [45] DIRECT OBSERVATION - FACTORS AFFECTING THE ACCURACY OF OBSERVERS
    REPP, AC
    NIEMINEN, GS
    OLINGER, E
    BRUSCA, R
    EXCEPTIONAL CHILDREN, 1988, 55 (01) : 29 - 36
  • [46] FACTORS AFFECTING ACCURACY OF AEROTRIANGULATION WITH AUXILIARY DATA
    FAIG, W
    ELHAKIM, S
    PHOTOGRAMMETRIC ENGINEERING AND REMOTE SENSING, 1976, 42 (06): : 826 - 826
  • [47] Factors affecting the accuracy of an active vision head
    Gasteratos, A
    Sandini, G
    METHODS AND APPLICATIONS OF ARTIFICIAL INTELLIGENCE, 2002, 2308 : 413 - 422
  • [48] FACTORS AFFECTING ACCURACY OF PRODUCING TIME INTERVALS
    MARMARAS, N
    VASSILAKIS, P
    DOUNIAS, G
    PERCEPTUAL AND MOTOR SKILLS, 1995, 80 (03) : 1043 - 1056
  • [49] Factors affecting the accuracy of elastometric impression materials
    Chen, SY
    Liang, WM
    Chen, FN
    JOURNAL OF DENTISTRY, 2004, 32 (08) : 603 - 609
  • [50] An examination of factors affecting accuracy in technology forecasts
    Fye, Shannon R.
    Charbonneau, Steven M.
    Hay, Jason W.
    Mullins, Carie A.
    TECHNOLOGICAL FORECASTING AND SOCIAL CHANGE, 2013, 80 (06) : 1222 - 1231