Development of a novel automated algorithm for patient dosimetry in computed tomography: a step towards the facilitation of size-specific dose estimates and organs dosimetry estimations in a busy clinical workflow

被引:2
|
作者
Sekkat, Hamza [1 ]
Khallouqi, Abdellah [1 ,2 ]
Madkouri, Youssef [3 ]
El Atifi, Wiam [1 ,4 ]
Rhouch, Imane [1 ]
El Rhazouani, Omar [1 ]
Tahiri, Zakaria [5 ]
Talbi, Mohammed [6 ,7 ]
Halimi, Abdellah [1 ]
机构
[1] Hassan 1st Univ, Higher Inst Hlth Sci, Lab Hlth Sci & Technol, Settat, Morocco
[2] Publ Hosp Mediouna, Dept radiol, Casablanca, Morocco
[3] Ibn Tofail Kenitra Univ, Fac Sci, Lab Elect Syst Informat Proc Mech & Energet, Kenitra, Morocco
[4] Hassan II Univ, Hosp Ctr Ibn Rochd, Fac Med & Pharm, Dept Radiotherapy, Casablanca, Morocco
[5] Abdelmalek Essaadi Univ, Fac Med & Pharm Tangier, Biophys Lab, Life & Hlth Sci Res Lab, Tetouan, Morocco
[6] Minist Hlth & Social Protect, Higher Inst Nursing Profess & Hlth Tech, Rabat, Morocco
[7] Ibn Tofail Univ, Fac Sci, Lab Mat & Subatom Phys, Kenitra, Morocco
关键词
computed tomography; automated algorithm; dosimetry; size-specific dose estimate; organ dose estimation;
D O I
10.1088/1402-4896/ad8b86
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Accurate dosimetry in computed tomography ( CT ) is essential for patient safety and effective radiation management. This study presents the development of an automated algorithm designed to enhance patient dosimetry by facilitating size-specific dose estimates ( SSDE ) and organ dose estimations. Utilizing a Python-based script, the proposed method integrates advanced image preprocessing, contour detection, and mathematical calculations to quantify key metrics from CT images. This automated approach addresses the limitations of manual measurement techniques. A retrospective analysis was conducted on CT axial images from examinations acquired with an 80-detector scanner. The algorithm processes DICOM images, converts pixel values to Hounsfield Units, applies Gaussian smoothing, windowing, and thresholding, followed by morphological operations to refine segmentation. It measures the water equivalent diameter ( Dw ) and estimates both region SSDE and organ doses, incorporating tissue attenuation. Validation was performed using an adult anthropomorphic ATOM phantom, with organ doses measured by optically stimulated luminescence dosimeters. The results demonstrated the algorithm's potential in estimating SSDE and organ doses. Validation of the automated method revealed strong correlations for Dw and SSDE between the proposed method and manual measurements of fi ve expert reviewers ranging from 0.86 to 0.99 for determination coefficient. Comparative analysis of organ doses showed close agreement between results from experimental setup against the proposed algorithm. The automated algorithm estimated brain dose with a mean of 21.8 mGy, while measurements from the ATOM phantom and CT Expo indicated 19.74 mGy and 23.05 mGy, respectively. For lung doses, the automated algorithm estimated 12.5 mGy compared to 11.0 mGy from the ATOM phantom and 13.1 mGy from CT Expo. Liver doses were measured at 12.7 mGy by the automated method, versus 12.1 mGy from the ATOM phantom and 11.1 mGy from CT Expo. This study shows the potential of automated image analysis techniques in enhancing dosimetry accuracy in CT examinations.
引用
收藏
页数:17
相关论文
共 11 条
  • [1] A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans
    Abdulkadir, Muhammad Kabir
    Osman, Noor Diyana
    Achuthan, Anusha
    Nasirudin, Radin A.
    Ahmad, Muhammad Zabidi
    Zain, Noor Hasyima Mat
    Shuaib, Ibrahim Lutfi
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 456 - 463
  • [2] Development of size-specific dose estimates for common computed tomography examinations: a study in Ghana
    Ohene-Botwe, Benard
    Anim-Sampong, Samuel
    Nkansah, Josephine
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (01)
  • [3] Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Hanaoka, Shinsuke
    Matsubara, Kosuke
    Takata, Tadanori
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04): : 246 - 251
  • [4] SIMPLE METHOD OF SIZE-SPECIFIC DOSE ESTIMATES CALCULATION FROM PATIENT WEIGHT ON COMPUTED TOMOGRAPHY
    Iriuchijima, Akiko
    Fukushima, Yasuhiro
    Nakajima, Takahito
    Tsushima, Yoshito
    Ogura, Akio
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (02) : 208 - 212
  • [5] Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction
    Ueno, Hiroyuki
    Matsubara, Kosuke
    Bou, Sayuri
    Hizume, Masato
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [6] Novel method for calculating the effective dose using size-specific dose estimates conversion factors in abdomen–pelvis computed tomography
    Kentaro Funashima
    Shigeru Abiko
    Kazuhiro Sato
    Radiological Physics and Technology, 2023, 16 (4) : 506 - 515
  • [7] Factors influencing size-specific dose estimates of selected computed tomography protocols at two clinical practices in South Africa
    Sebelego, Ida-Keshia
    Acho, Sussan
    van der Merwe, Belinda
    Rae, William I. D.
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (07) : 588 - 602
  • [8] Novel method for calculating the effective dose using size-specific dose estimates conversion factors in abdomen-pelvis computed tomography
    Funashima, Kentaro
    Abiko, Shigeru
    Sato, Kazuhiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (04) : 506 - 515
  • [9] Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs
    Kuriyama, Kazuki
    Matsubara, Kosuke
    Hisahara, Shu
    Nagata, Yukie
    Nosaka, Rumiko
    Goto, Rika
    Yanano, Nobutaka
    Shimizu, Kanichiro
    Shoji, Tomokazu
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 665 - 672
  • [10] Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs
    Kazuki Kuriyama
    Kosuke Matsubara
    Shu Hisahara
    Yukie Nagata
    Rumiko Nosaka
    Rika Goto
    Nobutaka Yanano
    Kanichiro Shimizu
    Tomokazu Shoji
    Physical and Engineering Sciences in Medicine, 2020, 43 : 665 - 672