Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study

被引:0
|
作者
Dabli, Djamel [1 ]
Salvat, Cecile [2 ]
Fitton, Isabelle [3 ]
Ty, Claire Van Ngoc [3 ]
Palanchon, Peggy [4 ]
Beregi, Jean-Paul [1 ]
Greffier, Joel [1 ]
Hadid-Beurrier, Lama [2 ]
机构
[1] Montpellier Univ, Nimes Univ Hosp, IMAGINE UR UM 103, Dept Med Imaging, F-30029 Nimes 9, France
[2] Lariboisiere Univ Hosp, Med Phys & Radiat Protect Dept, AP HP, F-75010 Paris, France
[3] Paris City Univ, Georges Pompidou European Hosp, Dept Radiol, AP HP, F-75015 Paris, France
[4] CHU Angers, Dept Radiodiagnost, 4 Rue Larrey, F-49933 Angers, France
关键词
3D imaging; spine surgery; noise power spectrum; task transfer function; BEAM COMPUTED-TOMOGRAPHY; CT; NAVIGATION; OPTIMIZATION; FLUOROSCOPY;
D O I
10.3390/s24216883
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An image-quality CT phantom was scanned with three different 3D X-ray imaging guidance devices in the operating theatre: O-Arm, Loop-X, and Airo TruCT. Default acquisition and reconstruction parameters for lumbar spine procedures were used on each device. The tube current was set to a dose level of around 27 mGy. A task-based image quality assessment was performed by calculating the noise power spectrum (NPS) and task transfer function (TTF). A detectability index (d') was calculated for three simulated bone lesions. The noise magnitude of the O-Arm was higher than the Airo TruCT, and the Loop-X had higher noise than the Airo TruCT. The highest average NPS frequency was for the O-Arm images, and the lowest was for the Loop-X. The TTFs at 50% values were similar for the Airo TruCT and Loop-X devices. Compared to Airo TruCT, the TTF at 50% value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. Compared to Airo TruCT, the d' value was lower with Loop-X by -26.73%, -27.02%, and -23.95% for lytic lesions, sclerotic lesions, and high-density bone, respectively. Each 3D-imaging spine surgery guidance device has its own strengths and weaknesses in terms of image quality. Cone-beam CT systems apparently offer the best compromise between noise and spatial resolution for spine surgery.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Quantitative analysis of 3D coronary modeling in 3D rotational X-ray imaging
    Movassaghi, B
    Rasche, V
    Viergever, MA
    Niessen, W
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 878 - 880
  • [42] Additional Imaging Guidance Dose to Patient Organs Resulting From X-Ray Tubes Used in CyberKnife Image Guidance System
    Sullivan, A.
    Ding, G.
    MEDICAL PHYSICS, 2015, 42 (06) : 3264 - 3264
  • [43] IMAGE QUALITY STUDY IN 3D X-RAY ANGIOGRAPHY - A FIRST APPROACH USING THE EXPERIMENTAL-DESIGN STRATEGY
    RENAUDIN, C
    MAGNIN, IE
    PICARD, C
    TROUSSET, Y
    SERGENT, M
    AMIEL, M
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1993, 17 (4-5) : 309 - 322
  • [44] 3D interventional imaging with 2D x-ray detectors
    Desbat, L
    Champleboux, G
    Fleute, M
    Komarek, P
    Mennessier, C
    Monteil, B
    Rodet, T
    Bessou, P
    Coulomb, M
    Ferretti, G
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION, MICCAI'99, PROCEEDINGS, 1999, 1679 : 973 - 980
  • [45] 3D registration through pseudo x-ray image generation
    Viant, WJ
    Barnel, F
    MEDICINE MEETS VIRTUAL REALITY 2001: OUTER SPACE, INNER SPACE, VIRTUAL SPACE, 2001, 81 : 541 - 543
  • [46] ARTICULATED REGISTRATION OF 3D HUMAN GEOMETRY TO X-RAY IMAGE
    Le, Sang N.
    Karlekar, Jayashree
    Fang, Anthony C.
    2008 15TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, VOLS 1-5, 2008, : 1108 - 1111
  • [47] 3D simulation of the image formation in soft x-ray microscopes
    Selin, Marten
    Fogelqvist, Emelie
    Holmberg, Anders
    Guttmann, Peter
    Vogt, Ulrich
    Hertz, Hans M.
    OPTICS EXPRESS, 2014, 22 (25): : 30756 - 30768
  • [48] 3D Image Reconstruction from X-Ray Measurements with Overlap
    Klodt, Maria
    Hauser, Raphael
    COMPUTER VISION - ECCV 2016, PT VI, 2016, 9910 : 19 - 33
  • [49] Towards a Quality Assurance Protocol for 2D and 3D Breast X-Ray Imaging Techniques
    Bosmans, H.
    Vollmar, S.
    Lemmens, K.
    Kalender, W.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 647 - 649
  • [50] Alternative in 3D X-ray imaging brings dynamic solutions
    不详
    INSIGHT, 2017, 59 (08) : 406 - 406