Significance and feasibility of air kerma length product and air kerma area product comparisons

被引:0
|
作者
Zivanovic, Milos [1 ]
Kortmis, Maja Vojnic [2 ]
Krzanovic, Nikola [1 ]
Daletic, Milos [1 ]
Komatina, Ivana [1 ,3 ]
机构
[1] Univ Belgrade, Vinca Inst Nucl Sci, Natl Inst Republ Serbia, Dept Radiat & Environm Protect, Mike Petrovica Alasa 12-14, Belgrade 11351, Serbia
[2] IBA Dosimetry GmbH, Secondary Stand Dosimetry Lab, Banhofstr 5, Schwarzenbruck, Germany
[3] Univ Belgrade, Fac Phys Chem, Studentski Trg 12-16, Belgrade 11158, Serbia
关键词
Diagnostic radiology; Intercomparison; Air kerma area product; Air kerma length product; KAP-meter; CT-chamber; CALIBRATION; QUALITY;
D O I
10.1016/j.radmeas.2024.107291
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Air kerma length product and air kerma area product are special quantities used in diagnostic radiology. They are measured using special measurement devices - CT-chambers and KAP-meters, in order to calculate quantities related to patient exposure. Appropriate calibration of all measurement devices is of vital importance, and comparisons between calibration laboratories are necessary to prove competence. It is usually considered adequate to participate in air kerma comparisons to prove capabilities for special quantities, but the literature shows that some problems in calibration procedure can remain unknown. The comparisons directly in terms of special quantities provide additional burden to laboratories, and require special transfer instruments, but they allow checking the whole calibration procedures. This paper describes a comparison between two calibration laboratories in terms of both air kerma length product and air kerma area product. Both laboratories achieved good results for all radiation qualities, considering the measurement uncertainty. Transfer instruments' linearity, field size dependence and energy dependence were investigated. Even though the metrological properties of the transfer instruments are worse than the ionization chambers, they can be taken into account by introducing additional measurement uncertainty, performing appropriate corrections or choosing calibration points for the comparison for the values of influence quantities where the transfer instrument response is relatively flat. These comparisons provide additional value to calibration laboratories, but there are still several challenges related to their organization and execution.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Changes to the VNIIM air kerma primary standard
    Kharitonov, I. A.
    Oborin, A. V.
    Villevalde, A. Y.
    METROLOGIA, 2007, 44 (06) : L71 - L72
  • [42] Seasonal variation of air kerma in the "Vulcano Porto" area (Aeolian Islands, Italy)
    Bellia, S
    Basile, S
    Brai, M
    Hauser, S
    Puccio, P
    Rizzo, S
    APPLIED RADIATION AND ISOTOPES, 2001, 54 (04) : 701 - 706
  • [43] AUDIENCE PARTICIPATION PRESENTATION - DISCUSSION AND VOTE ON EXPOSURE VS AIR KERMA VS WATER KERMA
    LOEVINGER, R
    MEDICAL PHYSICS, 1982, 9 (04) : 646 - 646
  • [44] Kerma-area product is not a useful indictor of potential tissue reactions in interventional radiology
    O'Hora, L.
    RADIOGRAPHY, 2016, 22 (03) : E196 - E200
  • [45] Noises Depression Based on Statistic Analysis Algorithm for Kerma Area Product Meter Design
    Hsieh, Chi-Wen
    Chen, Chih-Yen
    2018 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC): DISCOVERING NEW HORIZONS IN INSTRUMENTATION AND MEASUREMENT, 2018, : 959 - 963
  • [46] An international calibration of Kerma-Area Product meters for patient dose optimisation study
    Jankowski, J.
    Domienik, J.
    Papierz, S.
    Padovani, R.
    Vano, E.
    Faulkner, K.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 328 - 332
  • [47] Conversion factors for effective dose and organ doses with the air kerma area product in patients undergoing percutaneous transhepatic biliary drainage and trans arterial chemoembolization
    Brambilla, M.
    Cannillo, B.
    Guzzardi, G.
    D'Alessio, Andrea
    Galbiati, Andrea
    Matheoud, Roberta
    Carriero, A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 72 : 7 - 15
  • [48] Effect of equalization filters on measurements with kerma-area product meter in a cardiovascular angiography system
    Ichikawa, Nao
    Fukuda, Atsushi
    Hayashi, Takuma
    Matsubara, Kosuke
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 177 - 185
  • [49] Better Kerma-Area-Product (KAP) Estimation Using the System Parameters in Radiography and Fluoroscopy
    Kim, D.
    MacDougall, R.
    MEDICAL PHYSICS, 2016, 43 (06) : 3666 - 3667
  • [50] Review of Kerma-Area Product and total energy incident on patients in radiography, mammography and CT
    Yao, Hai
    Huda, Walter
    Mah, Eugene
    He, Wenjun
    RADIATION PROTECTION DOSIMETRY, 2015, 163 (02) : 251 - 260