Quantitative and qualitative performance evaluation of commercial metal artifact reduction methods: Dosimetric effects on the treatment planning

被引:0
|
作者
Ghorbanzadeh, Mohammad [1 ]
Hosseini, Seyed Abolfazl [1 ]
Vahdat, Bijan Vosoughi [2 ]
Mirzaiy, Hamed [3 ]
Akhavanallaf, Azadeh [4 ]
Arabi, Hossein [4 ]
机构
[1] Sharif Univ Technol, Dept Energy Engn, Tehran 863911365, Iran
[2] Sharif Univ Technol, Dept Elect Engn, Tehran 863911365, Iran
[3] Erfan Radiat Oncol Ctr EROC, Tehran 57176, Iran
[4] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, CH-1211 Geneva, Switzerland
关键词
Computerized tomography; Metal artifact reduction; Radiation therapy; CLINICAL-EVALUATION; CT; RESOLUTION; RADIATION; IMPLANTS; DESIGN;
D O I
10.1016/j.radphyschem.2024.112140
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The presence of metal implants within CT imaging causes severe attenuation of the X-ray beam. Due to the incomplete information recorded by CT detectors, artifacts in the form of streaks and dark bands would appear in the resulting CT images. The metal-induced artifacts would firstly affect the quantitative accuracy of CT imaging, and consequently, the radiation treatment planning and dose estimation in radiation therapy. To address this issue, CT scanner vendors have implemented metal artifact reduction (MAR) algorithms to avoid such artifacts and enhance the overall quality of CT images. The orthopedic-MAR (OMAR) and normalized MAR (NMAR) algorithms are the most well-known metal artifact reduction (MAR) algorithms, used worldwide. These algorithms have been implemented on Philips and Siemens scanners, respectively. In this study, we set out to quantitatively and qualitatively evaluate the effectiveness of these two MAR algorithms and their impact on accurate radiation treatment planning and CT-based dosimetry. The quantitative metrics measured on the simulated metal artifact dataset demonstrated superior performance of the OMAR technique over the NMAR one in metal artifact reduction. The analysis of radiation treatment planning using the OMAR and NMAR techniques in the corrected CT images showed that the OMAR technique reduced the toxicity of healthy tissues by 10% compared to the uncorrected CT images.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Evaluation of Three Commercial Metal Artifact Reduction Methods for CT Simulations in Radiation Therapy Treatment Planning
    Huang, J.
    Kerns, J.
    Nute, J.
    Liu, X.
    Stingo, F.
    Followill, D.
    Mirkovic, D.
    Howell, R.
    Kry, S.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 498 - 498
  • [2] Dosimetric Impact of Metal Artifact Reduction in Treatment Planning: Phantom and Clinical Studies
    Chen, D.
    Chetty, I.
    Glide-Hurst, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [3] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08): : R49 - R80
  • [4] Quantitative evaluation of metal artifact reduction techniques
    Kolind, SH
    MacKay, AL
    Munk, PL
    Xiang, QS
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (03) : 487 - 495
  • [5] Evaluation of two commercial CT metal artifact reduction algorithms for use in proton radiotherapy treatment planning in the head and neck area
    Andersson, Karin M.
    Dahlgren, Christina Vallhagen
    Reizenstein, Johan
    Cao, Yang
    Ahnesjoe, Anders
    Thunberg, Per
    [J]. MEDICAL PHYSICS, 2018, 45 (10) : 4329 - 4344
  • [6] A dosimetric evaluation of interface effects using two commercial electron treatment planning algorithms
    Catt, B.
    Yudelev, M.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 456 - 459
  • [7] Evaluation of Metal Artifact Reduction Technique for the Radiation Therapy Planning
    Jeong, K.
    Kuo, H.
    Ritter, J.
    Shen, J.
    Basavatia, A.
    Yaparpalvi, R.
    Kalnicki, S.
    Tome, W.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3645 - 3645
  • [8] Quantitative Comparison of Commercial and Non-Commercial Metal Artifact Reduction Techniques in Computed Tomography
    Wagenaar, Dirk
    van der Graaf, Emiel R.
    van der Schaaf, Arjen
    Greuter, Marcel J. W.
    [J]. PLOS ONE, 2015, 10 (06):
  • [9] Technical note: evaluation of dosimetric performance in a commercial 3D treatment planning system
    Sandilos, P
    Seferlis, S
    Antypas, C
    Karaiskos, P
    Dardoufas, C
    Vlahos, L
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (934): : 899 - 905
  • [10] Metal Artifact Reduction in Photon-Counting Detector CT Quantitative Evaluation of Artifact Reduction Techniques
    Skornitzke, Stephan
    Mergen, Victor
    Biederer, Juergen
    Alkadhi, Hatem
    Do, Thuy D.
    Stiller, Wolfram
    Frauenfelder, Thomas
    Kauczor, Hans-Ulrich
    Euler, Andre
    [J]. INVESTIGATIVE RADIOLOGY, 2024, 59 (06) : 442 - 449