摘要
The aim of this work was to evaluate the dosimetric performance of a commercial treatment planning system (TPS) which employs a three-dimensional calculation algorithm (Nucletron Plato version 2.2.3), following the guidelines of the AAPM Task Group 23 (TG23). Seven test cases were used to test the TPS dosimetric performance in homogeneous water. These cases involved absolute dose measurements on central as well as off-axis points situated at various depths, using simple field arrangements, and comparison with corresponding TPS calculations. This comparison yielded differences within +/- 2% at all points, for all test cases. To test the ability of the TPS to account for tissue inhomogeneities, corresponding comparisons were performed with the presence of a low-density material in the beam to resemble an air inhomogeneity. Absolute dose measurements and corresponding TPS calculations showed a mean deviation of the order of +/- 3.5%, reaching a maximum of 11.5% for small field sizes (5 cm x 5 cm). In summary, observed deviations are well within the set tolerance levels while comparison with previous TPS versions showed that Plato version 2.2.3 is significantly improved, especially in dose calculations in the presence of low density inhomogeneities.
引用
收藏
页码:899 / 905
页数:7
相关论文
共 50 条
  • [1] Dosimetric evaluation of a commercial 3D treatment planning system using the AAPM Task Group 23 test package
    Borca, VC
    Pasquino, M
    Bresciani, S
    Catuzzo, P
    Ozzello, F
    Tofani, S
    [J]. MEDICAL PHYSICS, 2005, 32 (03) : 744 - 751
  • [2] Dosimetric verification of a commercial 3D treatment planning system for conformal radiotherapy with a dynamic multileaf collimator
    Pignoli, E
    Serretiello, S
    Somigliana, A
    Zonca, G
    Pellegrini, R
    Mongioj, V
    Marchesini, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08): : N77 - N84
  • [3] Dosimetric evaluation of a commercial inverse treatment planning system: UAB experience
    Duan, J
    Brezovich, IA
    Pareek, PN
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1162 - 1164
  • [4] Modeling and dosimetric performance evaluation of the RayStation treatment planning system
    Mzenda, Bongile
    Mugabe, Koki V.
    Sims, Rick
    Godwin, Guy
    Loria, Dayan
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05): : 29 - 46
  • [5] Dosimetric evaluation of a commercial 3-D treatment planning system using Report 55 by AAPM Task Group 23
    Declich, F
    Fumasoni, K
    Mangili, P
    Cattaneo, GM
    Iori, M
    [J]. RADIOTHERAPY AND ONCOLOGY, 1999, 52 (01) : 69 - 77
  • [6] Dosimetric verification of a dedicated 3D treatment planning system for episcleral plaque therapy
    Knutsen, S
    Hafslund, R
    Monge, OR
    Valen, H
    Muren, LP
    Rekstad, BL
    Krohn, J
    Dahl, O
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04): : 1159 - 1166
  • [7] Dosimetric verification of a commercial inverse treatment planning system
    Lei, X
    Curran, B
    Hill, R
    Holmes, T
    Ma, LJ
    Forster, KM
    Boyer, AL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02): : 463 - 478
  • [8] Commissioning and Comprehensive Quality Assurance of commercial 3D Treatment Planning System using IAEA Technical Report Series — 430
    S. V. Jamema
    R. R. Upreti
    S. Sharma
    D. D. Deshpande
    [J]. Australasian Physics & Engineering Sciences in Medicine, 2008, 31 : 207 - 215
  • [9] Commissioning and comprehensive quality assurance of commercial 3D treatment planning system using IAEA technical report series 430
    Jamema, S.
    Upreti, R.
    Sharma, S.
    Shanna, D.
    Kinhikar, R.
    Patkar, S.
    Sharma, P.
    Deshpande, D.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2426 - 2426
  • [10] STANFORD 3D HYPERTHERMIA TREATMENT PLANNING SYSTEM - TECHNICAL REVIEW AND CLINICAL SUMMARY
    SULLIVAN, DM
    BENYOSEF, R
    KAPP, DS
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1993, 9 (05) : 627 - 643