Dose assessment in petrous bone CT: Effect of tube current modulation and scan length

被引:0
|
作者
Saga, Zouhir [1 ]
Rahmouni, Abdelaali [1 ]
Talbi, Mohammed [2 ]
Anam, Choirul [3 ]
Rezzouk, Abdellah [1 ]
机构
[1] Sidi Mohamed Ben Abdellah Univ, Fac Sci Dhar El Mahraz, Dept Phys, Lab Solid State Phys, Fes, Morocco
[2] Ibn Tofail Univ, Higher Inst Nursing Profess & Hlth Tech, Fac Sci, Rabat Dept Phys,Lab Mat & Subatom Phys, Kenitra, Morocco
[3] Diponegoro Univ, Fac Math & Sci, Dept Phys, Semarang, Indonesia
关键词
Petrous bone CT; TCM; Scan length; DLP; Effective dose;
D O I
10.1016/j.radphyschem.2024.112125
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The aim of this study is to compare the dosimetric performance of two CT acquisition protocols for the petrous bone: a standard fixed mAs protocol and a protocol using tube current modulation (TCM). By analyzing the scan length, dose-length product (DLP) and effective dose (ED) obtained with each protocol, we investigate whether TCM offers an advantage in terms of dose reduction while maintaining comparable image quality. To this end, 249 patients (women: 52%; men: 48%) were included in this study. The first center used TCM (120 kVp, 150 mA reference) for image acquisition, while the other centers opted for fixed mAs values. Acceptability of image quality was assessed using a standardized phantom (model 610 AAPM) by measuring noise uniformity and accuracy, contrast-to-noise ratio (CNR) and spatial resolution. The CTDIvol was not significantly different between center A and the other centers (p > 0.05), while DLP and ED were significantly different (p < 0.001), being reduced by a factor of 2 and 2.23 for scan length. We concluded that a scan with a 14.28% decrease in tube voltage from 140 kVp (136-170-210) mAs fixed to 120 kVp with TCM to 150 mA reference in helical scan mode can be recommended as an optimized protocol during CT examination of petrous bone while reducing DLP and ED to 50% without loss of diagnostic accuracy.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Is It Better to Enter a Volume CT Dose Index Value before or after Scan Range Adjustment for Radiation Dose Optimization of Pediatric Cardiothoracic CT with Tube Current Modulation?
    Goo, Hyun Woo
    KOREAN JOURNAL OF RADIOLOGY, 2018, 19 (04) : 692 - 703
  • [22] Effect of CT Localizer Radiographs on Radiation Dose Associated With Automatic Tube Current Modulation: A Multivendor Study
    Paolicchi, Fabio
    Bastiani, Luca
    Negri, Jacopo
    Caramella, Davide
    CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2020, 49 (01) : 34 - 41
  • [23] Dose reduction in CT examinations of children by anatomically adapted tube current modulation
    Greess, H
    Noemayr, A
    Wolf, H
    Janka, R
    Kalender, WA
    Bautz, WA
    RADIOLOGY, 1999, 213P : 294 - 294
  • [24] Add Tube Current Modulation to a Low Dose Simulation Tool for CT Systems
    Ding, Y.
    Wen, G.
    Brown, K.
    Klahr, P.
    Dhanantwari, A.
    MEDICAL PHYSICS, 2015, 42 (06) : 3621 - 3621
  • [25] Dose conversion coefficients for CT examinations of adults with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20): : 6243 - 6261
  • [26] Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6309 - 6326
  • [27] The impact on CT dose of the variability in tube current modulation technology: a theoretical investigation
    Li, Xiang
    Segars, W. Paul
    Samei, Ehsan
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (16): : 4525 - 4548
  • [28] Head CT dose reduction with organ-based tube current modulation
    Ota, Joji
    Yokota, Hajime
    Kobayashi, Takenori
    Ogata, Yuki
    Kubo, Takumi
    Chida, Koichi
    Masuda, Yoshitada
    Uno, Takashi
    MEDICAL PHYSICS, 2022, 49 (03) : 1964 - 1971
  • [29] X-Ray Tube Current Modulation and Effective Dose Per Unit Dose-Length Product Conversion Factors in CT Dosimetry
    Tavrides, E.
    He, W.
    Huda, W.
    Yao, H.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [30] Dose equations for tube current modulation in CT scanning and the interpretation of the associated CTDIvol
    Dixon, Robert L.
    Boone, John M.
    MEDICAL PHYSICS, 2013, 40 (11)