Background: Volumetric-modulated arc therapy (VMAT) is an efficient method of administering intensity-modulated radiotherapy beams. The Delta(4) device was employed to examine patient data. Aims and Objectives: The utility of the Delta(4) device in identifying errors for patient-specific quality assurance of VMAT plans was studied in this research. Materials and Methods: Intentional errors were purposely created in the collimator rotation, gantry rotation, multileaf collimator (MLC) position displacement, and increase in the number of monitor units (MU). Results: The results show that when the characteristics of the treatment plans were changed, the gamma passing rate (GPR) decreased. The largest percentage of erroneous detection was seen in the increasing number of MU, with a GPR ranging from 41 to 92. Gamma analysis was used to compare the dose distributions of the original and intentional error designs using the 2%/2 mm criteria. The percentage of dose errors (DEs) in the dose-volume histogram (DVH) was also analyzed, and the statistical association was assessed using logistic regression. A modest association (Pearson's R-values: 0.12-0.67) was seen between the DE and GPR in all intentional plans. The findings indicated a moderate association between DVH and GPR. The data reveal that Delta(4) is effective in detecting mistakes in treatment regimens for head-and-neck cancer as well as lung cancer. Conclusion: The study results also imply that Delta(4) can detect errors in VMAT plans, depending on the details of the defects and the treatment plans employed.
引用
收藏
页码:56 / 63
页数:8
相关论文
共 50 条
  • [1] Intensity-modulated radiotherapy, coplanar volumetric-modulated arc, therapy, and noncoplanar volumetric-modulated arc therapy in, glioblastoma: A dosimetric comparison
    Hou, Yong
    Zhang, Yan
    Liu, Zhen
    Yv, Lili
    Liu, Kun
    Tian, Xiufang
    Lv, Yajuan
    [J]. CLINICAL NEUROLOGY AND NEUROSURGERY, 2019, 187
  • [2] Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy
    Thongsawad, Sangutid
    Khamfongkhruea, Chirasak
    Tannanonta, Chirapha
    [J]. JOURNAL OF MEDICAL PHYSICS, 2018, 43 (01) : 52 - 57
  • [3] Dosimetric comparison of tomotherapy and volumetric-modulated arc therapy for children with neuroblastoma
    Liu, Xia
    Liu, Zhikai
    Pang, Tingtian
    Dong, Tingting
    Qiu, Jie
    [J]. PEDIATRIC INVESTIGATION, 2020, 4 (03) : 186 - 191
  • [4] Dosimetric comparison of tomotherapy and volumetric-modulated arc therapy for children with neuroblastoma
    Liu Xia
    Liu Zhikai
    Pang Tingtian
    Dong Tingting
    Qiu Jie
    [J]. 儿科学研究(英文), 2020, 04 (03) : 186 - 191
  • [5] The importance of collimator angle error in volumetric-modulated arc therapy
    Puchades-Puchades, Vicente
    Serna-Berna, Alfredo
    Mata-Colodro, Fernando
    Ramos-Amores, David
    Alcaraz-Banos, Miguel
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (03) : 323 - 330
  • [6] A dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT
    Stathakis, Sotirios
    Narayanasamy, Ganesh
    Licon, Anna Laura
    Myers, Pamela
    Li, Ying
    Crownover, Richard
    Papanikolaou, Niko
    [J]. JOURNAL OF BUON, 2019, 24 (02): : 838 - 843
  • [7] Dosimetric Evaluation of Volumetric-modulated Arc Therapy (Rapid Arc) for Primary Leiomyosarcoma in the Spine
    Wu, Yong-Zhong
    Jin, Fu
    Wang, Ying
    Zhou, Juan
    Wu, Li-Hua
    He, Guang-Lei
    Qiu, Da
    He, Ya-Nan
    Liu, Xian-Feng
    Zhong, Ming-Song
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2014, 13 (05) : 387 - 394
  • [8] Dosimetric comparison of hybrid volumetric-modulated arc therapy, volumetric-modulated arc therapy, and intensity-modulated radiation therapy for left-sided early breast cancer
    Lin, Jia-Fu
    Yeh, Dah-Cherng
    Yeh, Hui-Ling
    Chang, Chen-Fa
    Lin, Jin-Ching
    [J]. MEDICAL DOSIMETRY, 2015, 40 (03) : 262 - 267
  • [9] Impact of Plan Complexity On the Delivery Accuracy and Dosimetric Accuracy of Volumetric-Modulated Arc Therapy
    Li, G.
    Li, Y.
    Wang, Q.
    Xiao, J.
    Bai, S.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2892 - 2892
  • [10] 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry
    Mans, Anton
    Remeijer, Peter
    Olaciregui-Ruiz, Igor
    Wendling, Markus
    Sonke, Jan-Jakob
    Mijnheer, Ben
    van Herk, Marcel
    Stroom, Joep C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) : 181 - 187