3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry

被引:153
|
作者
Mans, Anton [1 ]
Remeijer, Peter [1 ]
Olaciregui-Ruiz, Igor [1 ]
Wendling, Markus [1 ]
Sonke, Jan-Jakob [1 ]
Mijnheer, Ben [1 ]
van Herk, Marcel [1 ]
Stroom, Joep C. [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
关键词
Portal dosimetry; Volumetric-modulated arc therapy; In vivo dosimetry; VIVO EPID DOSIMETRY; CONE-BEAM CT; TREATMENT DELIVERY; DOSE VERIFICATION; IMRT TREATMENTS; PROSTATE IMRT; CANCER; RADIOTHERAPY; ALGORITHM; GUIDANCE;
D O I
10.1016/j.radonc.2009.12.020
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: To demonstrate the feasibility of back-projection portal dosimetry for accurate 3D dosimetric verification of volumetric-modulated arc therapy (VMAT), pre-treatment as well as in vivo. Materials and methods: Several modifications to our existing approach were implemented to make the method applicable to VMAT: (i) gantry angle-resolved data acquisition, (ii) calculation of the patient transmission, (iii) compensation for detector 'flex' and (iv) 3D dose reconstruction and evaluation. Results: Planned and EPID-(Electronic Portal Image Detector)-reconstructed close distributions show good agreement for pre-treatment verification of two prostate, a stereotactic lung and a head-and-neck VMAT plan and for in vivo verification of VMAT treatments of prostate and lung cancer. Averaged over pre-treatment verifications, planned and measured isocentre dose ratios were -1.2% (range [-4.7%,1.8%]). 3D gamma analysis (3% maximum dose, 3 mm) revealed mean gamma <gamma(mean)> = 0.37 [0.34,0.39], maximum 1% gamma (gamma(1%)) = 0.72 [0.66,0.81] and percentage of points with -gamma <= 1 < P(gamma <= 1)> = 99% [97%,100%.] For in vivo verification, the average isocentre dose ratio was -1.2% [-0.8%,-1.7%], <gamma(mean)> = 0.52 [0.40,0.64], <gamma(1%)> = 0.92 [0.76,1.08] and < P(gamma <= 1)> = 96% [93%,100%]. Conclusions: Our portal dosimetry method was successfully adapted for verification of VMAT treatments, pre-treatment as well as in vivo. (C) 2010 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 94 (2010) 181-187
引用
收藏
页码:181 / 187
页数:7
相关论文
共 50 条
  • [1] The Dosimetric Verification Analysis of ArcCHECK and Portal Dosimetry for Volumetric-Modulated Arc Therapy Plans in Halcyon Radiation Delivery
    Li, C.
    Lu, J.
    Zhu, J.
    Yin, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E550 - E551
  • [2] Intensity-modulated radiotherapy, coplanar volumetric-modulated arc, therapy, and noncoplanar volumetric-modulated arc therapy in, glioblastoma: A dosimetric comparison
    Hou, Yong
    Zhang, Yan
    Liu, Zhen
    Yv, Lili
    Liu, Kun
    Tian, Xiufang
    Lv, Yajuan
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2019, 187
  • [3] Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques
    Clemente-Gutierrez, Francisco
    Perez-Vara, Consuelo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 198 - 217
  • [4] Dosimetric precision of 3D gel dosimetry compared with radiochromic films in volumetric modulated arc therapy
    Skyt, P. S.
    Hoye, E. M.
    Yates, E. S.
    Schmidt, M. L.
    Petersen, J. B. B.
    Balling, P.
    Muren, L. P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S409 - S410
  • [5] Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy
    Thongsawad, Sangutid
    Khamfongkhruea, Chirasak
    Tannanonta, Chirapha
    JOURNAL OF MEDICAL PHYSICS, 2018, 43 (01) : 52 - 57
  • [6] 3D DOSIMETRIC VERIFICATION OF VMAT BY PORTAL DOSIMETRY
    Mans, A.
    Wendling, M.
    Stroom, J.
    Remeijer, P.
    Sonke, J. J.
    Mijnheer, B.
    van Herk, M.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S25 - S26
  • [7] First Clinical Results of 3D Verification of Volumetric-Modulated Arc Therapy (VMAT) of Prostate and Lung Cancer Patients Using EPID Dosimetry
    Mijnheer, B.
    Mans, A.
    Olaciregui-Ruiz, I.
    Sonke, J.
    Stroom, J.
    Tielenburg, R.
    Herk, M.
    Vijlbrief, R.
    MEDICAL PHYSICS, 2010, 37 (06) : 3351 - +
  • [8] Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS
    Vikraman, Subramani
    Manigandan, Durai
    Karrthick, Karukkupalayam Palaniappan
    Sambasivaselli, Raju
    Senniandavar, Vellaingiri
    Ramu, Mahendran
    Rajesh, Thiyagarajan
    Lutz, Muller
    Muthukumaran, Manavalan
    Karthikeyan, Nithyanantham
    Tejinder, Kataria
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 192 - 207
  • [9] DOSIMETRIC VERIFICATION OF RAPIDARC™ VOLUMETRIC MODULATED ARC THERAPY
    Medin, J.
    Kjaer-Kristoffersen, F.
    Ceberg, S.
    Back, S.
    Gustavsson, H.
    Archambault, Y.
    Bocanek, J.
    Cozzi, L.
    Eberhard, V.
    Gomola, I.
    Wiberg, I.
    Korreman, S.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S47 - S48
  • [10] Commissioning and dosimetric verification of volumetric modulated arc therapy for multiple modalities using electronic portal imaging device-based 3D dosimetry system: a novel approach
    Hajare, Raghavendra
    Sreelakshmi, K. K.
    Kumar, Anil
    Kalita, Rituraj
    Kaginelli, Shanmukhappa
    Mahantshetty, Umesh
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) : 412 - 424