Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques

被引:25
|
作者
Clemente-Gutierrez, Francisco [1 ]
Perez-Vara, Consuelo [1 ]
机构
[1] Hosp Cent Def Gomez Ulla, Secc Radiofis, Serv Oncol Radioterap, Madrid 28047, Spain
来源
关键词
VMAT QA; pretreatment verifications; Mobius3D; COMPASS; CARLO-BASED PHOTON; RADIATION-THERAPY; IMRT QA; DETECTOR ARRAY; PASSING RATES; VMAT QA; BEAM; CONVOLUTION/SUPERPOSITION; CHAMBER; PHANTOM;
D O I
10.1120/jacmp.v16i2.5190
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/3 mm criteria). The clinical use of both systems was successfully evaluated.
引用
收藏
页码:198 / 217
页数:20
相关论文
共 50 条
  • [1] 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry
    Mans, Anton
    Remeijer, Peter
    Olaciregui-Ruiz, Igor
    Wendling, Markus
    Sonke, Jan-Jakob
    Mijnheer, Ben
    van Herk, Marcel
    Stroom, Joep C.
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) : 181 - 187
  • [2] Clinical Implementation and dosimetric Verification of Volumetric Modulated Arc Therapy (VMAT)
    Wolff, D.
    Barrois, V
    Vogelgesang, L.
    Bosold, S.
    Tsogtbaatar, L.
    Sparenberg, J.
    Engenhart-Cabillic, R.
    Zink, K.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 : 157 - 157
  • [3] Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT)
    Li, Guangjun
    Bai, Sen
    Chen, Nianyong
    Henderson, Lansdale
    Wu, Kui
    Xiao, Jianghong
    Zhang, Yingjie
    Jiang, Qingfeng
    Jiang, Xiaoqin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05): : 13 - 24
  • [4] A novel method for routine quality assurance of volumetric-modulated arc therapy
    Wang, Qingxin
    Dai, Jianrong
    Zhang, Ke
    MEDICAL PHYSICS, 2013, 40 (10)
  • [5] Quality assurance of volumetric modulated arc therapy: Evaluation and comparison of different dosimetric systems
    Masi, L.
    Casamassima, F.
    Doro, R.
    Francescon, P.
    MEDICAL PHYSICS, 2011, 38 (02) : 612 - 621
  • [6] Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    Letourneau, Daniel
    Publicover, Julia
    Kozelka, Jakub
    Moseley, Douglas J.
    Jaffray, David A.
    MEDICAL PHYSICS, 2009, 36 (05) : 1813 - 1821
  • [7] Geometric and Dosimetric Quality Assurance in Volumetric Modulated Arc Therapy (VMAT)
    Qian, J.
    Ma, Y.
    Lee, L.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S639 - S639
  • [8] Dose verification of volumetric-modulated arc therapy using one-dimensional and two-dimensional dosimeters
    Rehman, Jalil ur
    Syed, Zahra
    Hussain, Ghulam
    Ahmad, Nisar
    Asghar, H. M. Noor ul Huda Khan
    Gilani, Zaheer Abbas
    Nasar, Gulfam
    Akhter, Malik M.
    Isa, Muhammad
    Iqbal, Tahir
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (03) : 304 - 308
  • [9] A retrospective analysis for patient-specific quality assurance of volumetric-modulated arc therapy plans
    Li, Guangjun
    Wu, Kui
    Peng, Guang
    Zhang, Yingjie
    Bai, Sen
    MEDICAL DOSIMETRY, 2014, 39 (04) : 309 - 313
  • [10] First Clinical Results of 3D Verification of Volumetric-Modulated Arc Therapy (VMAT) of Prostate and Lung Cancer Patients Using EPID Dosimetry
    Mijnheer, B.
    Mans, A.
    Olaciregui-Ruiz, I.
    Sonke, J.
    Stroom, J.
    Tielenburg, R.
    Herk, M.
    Vijlbrief, R.
    MEDICAL PHYSICS, 2010, 37 (06) : 3351 - +