Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques

被引:25
|
作者
Clemente-Gutierrez, Francisco [1 ]
Perez-Vara, Consuelo [1 ]
机构
[1] Hosp Cent Def Gomez Ulla, Secc Radiofis, Serv Oncol Radioterap, Madrid 28047, Spain
来源
关键词
VMAT QA; pretreatment verifications; Mobius3D; COMPASS; CARLO-BASED PHOTON; RADIATION-THERAPY; IMRT QA; DETECTOR ARRAY; PASSING RATES; VMAT QA; BEAM; CONVOLUTION/SUPERPOSITION; CHAMBER; PHANTOM;
D O I
10.1120/jacmp.v16i2.5190
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/3 mm criteria). The clinical use of both systems was successfully evaluated.
引用
收藏
页码:198 / 217
页数:20
相关论文
共 50 条
  • [41] Transition From Manual to Automated Planning and Delivery of Volumetric Modulated Arc Therapy Stereotactic Radiosurgery: Clinical, Dosimetric, and Quality Assurance Results
    Popple, Richard A.
    Brown, Matthew H.
    Thomas, Evan M.
    Willey, Christopher D.
    Cardan, Rex A.
    Covington, Elizabeth L.
    Riley, Kristen O.
    Markert, James M.
    Bredel, Markus
    Fiveash, John B.
    PRACTICAL RADIATION ONCOLOGY, 2021, 11 (02) : E163 - E171
  • [42] 3D DenseNet for Improving Dose Prediction of Volumetric Modulated Arc Therapy in Prostate Cancer
    Fu, J.
    Singhrao, K.
    Wang, Z.
    Ruan, D.
    Low, D.
    Lewis, J.
    Qi, X.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [43] Retrospective evaluation of dosimetric quality for prostate carcinomas treated with 3D conformal, intensity modulated and volumetric modulated arc radiotherapy
    Crowe, Scott B.
    Kairn, Tanya
    Middlebrook, Nigel
    Hill, Brendan
    Christie, David R. H.
    Knight, Richard T.
    Kenny, John
    Langton, Christian M.
    Trapp, Jamie V.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2013, 60 (04) : 131 - 138
  • [44] Comparison of rectal dose reduction by a hydrogel spacer among 3D conformal radiotherapy, volumetric-modulated arc therapy, helical tomotherapy, CyberKnife and proton therapy
    Saito, Masahide
    Suzuki, Toshihiro
    Sugama, Yuya
    Marino, Kan
    Sano, Naoki
    Komiyama, Takafumi
    Aoki, Shinichi
    Maehata, Yoshiyasu
    Yoshizawa, Kazuya
    Ashizawa, Kazunari
    Suzuki, Hidekazu
    Ueda, Koji
    Miyasaka, Yosuke
    Araya, Masayuki
    Takahashi, Hiroshi
    Onishi, Hiroshi
    JOURNAL OF RADIATION RESEARCH, 2020, 61 (03) : 487 - 493
  • [45] Automatic Selection of An Absolute-Dose Verification Point for Volumetric-Modulated Arc Therapy Via Dose Uncertainty Potential Accumulation Model
    Saito, A.
    Nakashima, T.
    Kawahara, D.
    Shiba, E.
    Matsuura, T.
    Fukata, K.
    Miki, K.
    Tsuneda, M.
    Nishio, T.
    Nagata, Y.
    MEDICAL PHYSICS, 2020, 47 (06) : E708 - E708
  • [46] IMPLEMENTATION OF VOLUMETRIC INTENSITY MODULATED ARC THERAPY IN OROPHARINGEAL CANCER. A COMPARISON WITH CLASSICAL TECHNIQUES OF 3D CONFORMAL RADIOTHERAPY
    Popa, R.
    Suditu, M.
    Adam, D.
    Ciocaltei, V.
    Duliu, O. G.
    Popescu, A. I.
    ROMANIAN REPORTS IN PHYSICS, 2015, 67 (03) : 966 - 977
  • [47] Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS
    Vikraman, Subramani
    Manigandan, Durai
    Karrthick, Karukkupalayam Palaniappan
    Sambasivaselli, Raju
    Senniandavar, Vellaingiri
    Ramu, Mahendran
    Rajesh, Thiyagarajan
    Lutz, Muller
    Muthukumaran, Manavalan
    Karthikeyan, Nithyanantham
    Tejinder, Kataria
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01): : 192 - 207
  • [48] Treatment planning and dosimetric comparison study on two different volumetric modulated arc therapy delivery techniques
    Kumar, S. A. Syam
    Holla, Raghavendra
    Sukumar, Prabakar
    Padmanaban, Sriram
    Vivekanandan, Nagarajan
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2013, 18 (02) : 87 - 94
  • [49] Jaw Tracking Impacts on Treatment Plan Quality in Small Volumetric Modulated Arc Therapy Fields: A Dosimetric Verification
    Swinnen, A.
    Ollers, M.
    Roijen, E.
    Nijsten, S.
    Verhaegen, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E658 - E659
  • [50] Predicting the complexity of head-and-neck volumetric-modulated arc therapy planning using a radiation therapy planning quality assurance software
    Sasaki, Motoharu
    Nakaguchi, Yuji
    Kamomae, Takeshi
    Ueda, Shoji
    Endo, Yuto
    Sato, Daisuke
    Ikushima, Hitoshi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (06) : 963 - 972