Impact Assessment of Systemic Geometric Distortion in 1.5T Magnetic Resonance Imaging Simulation through Three-dimensional Geometric Distortion Phantom on Dosimetric Accuracy for Magnetic Resonance Imaging-only Prostate Treatment Planning

被引:0
|
作者
Chaknam, Korawig [1 ]
Worapruekjaru, Ladawan [2 ]
Suphaphong, Sithiphong [2 ]
Stansook, Nualjun [2 ]
Sodkokkruad, Prapa [1 ]
Asavaphatiboon, Sawwanee [1 ]
机构
[1] Mahidol Univ, Fac Med, Dept Diagnost & Therapeut Radiol, Div Diagnost Radiol,Ramathibodi Hosp, 270 Rama Vi Rd, Bangkok 10400, Thailand
[2] Mahidol Univ, Dept Diagnost & Therapeut Radiol, Div Radiat Oncol, Fac Med,Ramathibodi Hosp, Bangkok, Thailand
关键词
Dosimetric accuracy; magnetic resonance imaging-only treatment planning; magnetic resonance imaging-simulation; systemic geometric distortion; RADIATION-THERAPY; CHEMICAL-SHIFT; RADIOTHERAPY; REGISTRATION; MRI; VALIDATION;
D O I
10.4103/jmp.jmp_62_24
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aims: Magnetic resonance imaging (MRI)-only radiotherapy has emerged as a solution to address registration errors that can lead to missed dose delivery. However, the presence of systemic geometric distortion (SGD) stemming from gradient nonlinearity (GNL) and inhomogeneity of the main magnetic field (B0) necessitates consideration. This study aimed to quantitatively assess residual SGD in 1.5T MRI simulation using a three-dimensional (3D) geometric distortion phantom and evaluate its impact on dosimetric accuracy for retrospective prostate cancer patients.Materials and Methods: Ten retrospective cases of prostate cancer patients treated with volumetric modulated arc radiotherapy (VMAT) were randomly selected. A geometric distortion phantom was scanned on a 1.5T MRI simulation using a 3D T1 volumetric interpolated breath-hold examination sequence, varying bandwidth (BW), and two-phase-encoding directions. Distortion maps were generated and applied to the original computed tomography (oriCT) plan to create a distorted computed tomography plan (dCT), and a dice similarity coefficient (DSC) was observed. Dosimetric accuracy was evaluated by recalculating radiation dose for dCT plans using identical beam parameters as oriCT.Results: The SGD increased with distance from the isocenter in all series. DSC exceeded 0.95 for all plans except the rectum. Regarding GNL's impact on dosimetric accuracy, most mean percentage errors for clinical target volume, planning target volume, and both femurs were under 2% in all plans, except for the bladder and rectum.Conclusion: SGD pre-evaluation is crucial and should be incorporated into a quality assurance program to ensure effective MRI-simulation performance before MRI-only treatment planning for prostate cancer.
引用
收藏
页码:356 / 362
页数:7
相关论文
共 50 条
  • [21] Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imaging
    Torfeh, Tarraf
    Hammoud, Rabih
    McGarry, Maeve
    Al-Hammadi, Noora
    Perkins, Gregory
    MAGNETIC RESONANCE IMAGING, 2015, 33 (07) : 939 - 949
  • [22] Geometric impact and dose estimation of on-patient placement of a lightweight receiver coil in a clinical magnetic resonance imaging-only radiotherapy workflow for prostate cancer
    Scherman, Jonas
    af Wetterstedt, Sacha
    Persson, Emilia
    Olsson, Lars E.
    Gustafsson, Christian Jamtheim
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 26
  • [23] Quality assurance measurements of geometric accuracy for magnetic resonance imaging-based radiotherapy treatment planning
    Ranta, Iiro
    Kemppainen, Reko
    Keyrilainen, Jani
    Suilamo, Sami
    Heikkinen, Samuli
    Kapanen, Mika
    Saunavaara, Jani
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 62 : 47 - 52
  • [24] Three-dimensional proton magnetic resonance spectroscopic imaging with and without an endorectal coil: a prostate phantom study
    Ma, Chao
    Chen, Luguang
    Scheenen, Tom W. J.
    Lu, Jianping
    Wang, Jian
    ACTA RADIOLOGICA, 2015, 56 (11) : 1342 - 1349
  • [25] Genome wide association studies on three-dimensional aortic geometric indices measured with magnetic resonance imaging
    Beeche, C.
    Dib, M. J.
    Azzo, J. D.
    Maynard, H.
    Salman, O.
    Witschey, W.
    Chirinos, J.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [26] Geometric Reproducibility of Three-Dimensional Oral Implant Planning Based on Magnetic Resonance Imaging and Cone-Beam Computed Tomography
    Schwindling, Franz Sebastian
    Boehm, Sophia
    Herpel, Christopher
    Kronsteiner, Dorothea
    Vogel, Lorenz
    Juerchott, Alexander
    Heiland, Sabine
    Bendszus, Martin
    Rammelsberg, Peter
    Hilgenfeld, Tim
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (23)
  • [27] Impact of magnetic resonance imaging-related geometric distortion of dose distribution in fractionated stereotactic radiotherapy in patients with brain metastases
    Ohira, Shingo
    Suzuki, Yuta
    Washio, Hayate
    Yamamoto, Yuki
    Tateishi, Soichiro
    Inui, Shoki
    Kanayama, Naoyuki
    Kawamata, Minoru
    Miyazaki, Masayoshi
    Nishio, Teiji
    Koizumi, Masahiko
    Nakanishi, Katsuyuki
    Konishi, Koji
    STRAHLENTHERAPIE UND ONKOLOGIE, 2024, 200 (01) : 39 - 48
  • [28] Impact of magnetic resonance imaging-related geometric distortion of dose distribution in fractionated stereotactic radiotherapy in patients with brain metastases
    Shingo Ohira
    Yuta Suzuki
    Hayate Washio
    Yuki Yamamoto
    Soichiro Tateishi
    Shoki Inui
    Naoyuki Kanayama
    Minoru Kawamata
    Masayoshi Miyazaki
    Teiji Nishio
    Masahiko Koizumi
    Katsuyuki Nakanishi
    Koji Konishi
    Strahlentherapie und Onkologie, 2024, 200 : 39 - 48
  • [29] Cartilage Morphology at 3.0T: Assessment of Three-Dimensional Magnetic Resonance Imaging Techniques
    Chen, Christina A.
    Kijowski, Richard
    Shapiro, Lauren M.
    Tuite, Michael J.
    Davis, Kirkland W.
    Klaers, Jessica L.
    Block, Walter F.
    Reeder, Scott B.
    Gold, Garry E.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 32 (01) : 173 - 183
  • [30] Magnetic resonance imaging based digitally reconstructed radiographs, virtual simulation, and three-dimensional treatment planning for brain neoplasms
    Ramsey, CR
    Oliver, AL
    MEDICAL PHYSICS, 1998, 25 (10) : 1928 - 1934