Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imaging

被引:33
|
作者
Torfeh, Tarraf [1 ]
Hammoud, Rabih [1 ]
McGarry, Maeve [1 ]
Al-Hammadi, Noora [1 ]
Perkins, Gregory [1 ]
机构
[1] Hamad Med Corp, Natl Ctr Canc Care & Res, Dept Radiat Oncol, Doha, Qatar
关键词
Quality control; Magnetic resonance imaging; Geometric distortion; Software tools; CLINICAL MRI SYSTEMS; SCHEME;
D O I
10.1016/j.mri.2015.04.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To develop and validate a large field of view phantom and quality assurance software tool for the assessment and characterization of geometric distortion in MRI scanners commissioned for radiation therapy planning. Materials and Methods: A purpose built phantom was developed consisting of 357 rods (6 mm in diameter) of polymethyl-methacrylat separated by 20 mm intervals, providing a three dimensional array of control points at known spatial locations covering a large field of view up to a diameter of 420 mm. An in-house software module was developed to allow automatic geometric distortion assessment. This software module was validated against a virtual dataset of the phantom that reproduced the exact geometry of the physical phantom, but with known translational and rotational displacements and warping. For validation experiments, clinical MRI sequences were acquired with and without the application of a commercial 3D distortion correction algorithm (Gradwarp (TM)). The software module was used to characterize and assess system-related geometric distortion in the sequences relative to a benchmark a dataset, and the efficacy of the vendor geometric distortion correction algorithms (GDC) was also assessed. Results: Results issued from the validation of the software against virtual images demonstrate the algorithm's ability to accurately calculate geometric distortion with sub-pixel precision by the extraction of rods and quantization of displacements. Geometric distortion was assessed for the typical sequences used in radiotherapy applications and over a clinically relevant 420 mm field of view (FOV). As expected and towards the edges of the field of view (FOV), distortion increased with increasing FOV. For all assessed sequences, the vendor GDC was able to reduce the mean distortion to below 1 mm over a field of view of 5, 10, 15 and 20 cm radius respectively. Conclusion: Results issued from the application of the developed phantoms and algorithms demonstrate a high level of precision. The results indicate that this platform represents an important, robust and objective tool to perform routine quality assurance of MR-guided therapeutic applications, where spatial accuracy is paramount. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:939 / 949
页数:11
相关论文
共 50 条
  • [1] Geometric distortion evaluation of magnetic resonance images by a new large field of view phantom for magnetic resonance based radiotherapy purposes
    Rostami, A.
    Naseri, Sh
    Momennezhad, M.
    Zare, H.
    Anvari, K.
    Badkhor, H. R. Sayah
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 18 (04): : 733 - 742
  • [2] Reproducibility of geometric distortion in magnetic resonance imaging based on phantom studies
    Mizowaki, T
    Nagata, Y
    Okajima, K
    Kokubo, M
    Negoro, Y
    Araki, N
    Hiraoka, M
    RADIOTHERAPY AND ONCOLOGY, 2000, 57 (02) : 237 - 242
  • [3] Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
    Wyatt, Jonathan
    Hedley, Stephen
    Johnstone, Emily
    Speight, Richard
    Kelly, Charles
    Henry, Ann
    Short, Susan
    Murray, Louise
    Sebag-Montefiore, David
    McCallum, Hazel
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 6 : 31 - 38
  • [4] A novel phantom and method for comprehensive 3-dimensional measurement and correction of geometric distortion in magnetic resonance imaging
    Wang, DM
    Doddrell, DM
    Cowin, G
    MAGNETIC RESONANCE IMAGING, 2004, 22 (04) : 529 - 542
  • [5] An Anthropomorphic Head Phantom Prototype for the Measurement of Geometric Distortion in Magnetic Resonance Imaging System
    Shurche, Sadegh
    Sooteh, Mohammad Yousefi
    Iranian Journal of Medical Physics, 2024, 21 (01) : 53 - 63
  • [6] Assessment of a novel commercial large field of view phantom for comprehensive MR imaging quality assurance of a 0.35T MRgRT system
    Lewis, Benjamin C.
    Shin, Jaeik
    Maraghechi, Borna
    Quinn, Benjamin
    Cole, Mike
    Barberi, Enzo
    Kim, Jin Sung
    Green, Olga
    Kim, Taeho
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (04):
  • [7] A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy
    Alzahrani, Meshal
    Broadbent, David A.
    Teh, Irvin
    Al-Qaisieh, Bashar
    Johnstone, Emily
    Speight, Richard
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2025, 33
  • [8] Quality Assurance Testing Methods and Customized Phantom for Magnetic Resonance Imaging and Spectroscopy
    Song, K-H
    Lee, D-W
    Choe, B-Y
    MEDICAL PHYSICS, 2015, 42 (06) : 3255 - 3255
  • [9] Validation Of A Novel Software For Correcting Distortion In Cranial Magnetic Resonance (Mr) Images
    Calvo Ortega, J. F.
    Mateos, J.
    Alberich, A.
    Moragues, S.
    Casals, J.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1188 - S1188
  • [10] Phantom for assessment of fat suppression in large field-of-view diffusion-weighted magnetic resonance imaging
    Winfield, J. M.
    Douglas, N. H. M.
    deSouza, N. M.
    Collins, D. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09): : 2235 - 2248