Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imaging

被引:33
|
作者
Torfeh, Tarraf [1 ]
Hammoud, Rabih [1 ]
McGarry, Maeve [1 ]
Al-Hammadi, Noora [1 ]
Perkins, Gregory [1 ]
机构
[1] Hamad Med Corp, Natl Ctr Canc Care & Res, Dept Radiat Oncol, Doha, Qatar
关键词
Quality control; Magnetic resonance imaging; Geometric distortion; Software tools; CLINICAL MRI SYSTEMS; SCHEME;
D O I
10.1016/j.mri.2015.04.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To develop and validate a large field of view phantom and quality assurance software tool for the assessment and characterization of geometric distortion in MRI scanners commissioned for radiation therapy planning. Materials and Methods: A purpose built phantom was developed consisting of 357 rods (6 mm in diameter) of polymethyl-methacrylat separated by 20 mm intervals, providing a three dimensional array of control points at known spatial locations covering a large field of view up to a diameter of 420 mm. An in-house software module was developed to allow automatic geometric distortion assessment. This software module was validated against a virtual dataset of the phantom that reproduced the exact geometry of the physical phantom, but with known translational and rotational displacements and warping. For validation experiments, clinical MRI sequences were acquired with and without the application of a commercial 3D distortion correction algorithm (Gradwarp (TM)). The software module was used to characterize and assess system-related geometric distortion in the sequences relative to a benchmark a dataset, and the efficacy of the vendor geometric distortion correction algorithms (GDC) was also assessed. Results: Results issued from the validation of the software against virtual images demonstrate the algorithm's ability to accurately calculate geometric distortion with sub-pixel precision by the extraction of rods and quantization of displacements. Geometric distortion was assessed for the typical sequences used in radiotherapy applications and over a clinically relevant 420 mm field of view (FOV). As expected and towards the edges of the field of view (FOV), distortion increased with increasing FOV. For all assessed sequences, the vendor GDC was able to reduce the mean distortion to below 1 mm over a field of view of 5, 10, 15 and 20 cm radius respectively. Conclusion: Results issued from the application of the developed phantoms and algorithms demonstrate a high level of precision. The results indicate that this platform represents an important, robust and objective tool to perform routine quality assurance of MR-guided therapeutic applications, where spatial accuracy is paramount. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:939 / 949
页数:11
相关论文
共 50 条
  • [21] Impact Assessment of Systemic Geometric Distortion in 1.5T Magnetic Resonance Imaging Simulation through Three-dimensional Geometric Distortion Phantom on Dosimetric Accuracy for Magnetic Resonance Imaging-only Prostate Treatment Planning
    Chaknam, Korawig
    Worapruekjaru, Ladawan
    Suphaphong, Sithiphong
    Stansook, Nualjun
    Sodkokkruad, Prapa
    Asavaphatiboon, Sawwanee
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 356 - 362
  • [22] Development of a temperature-controlled phantom for magnetic resonance quality assurance of diffusion, dynamic, and relaxometry measurements
    Jerome, Neil P.
    Papoutsaki, Marianthi-Vasiliki
    Orton, Matthew R.
    Parkes, Harold G.
    Winfield, Jessica M.
    Boss, Michael A.
    Leach, Martin O.
    deSouza, Nandita M.
    Collins, David J.
    MEDICAL PHYSICS, 2016, 43 (06) : 2998 - 3007
  • [23] Quality assurance measurements of geometric accuracy for magnetic resonance imaging-based radiotherapy treatment planning
    Ranta, Iiro
    Kemppainen, Reko
    Keyrilainen, Jani
    Suilamo, Sami
    Heikkinen, Samuli
    Kapanen, Mika
    Saunavaara, Jani
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 62 : 47 - 52
  • [24] Quality assurance of magnetic resonance imaging for a polymer gel dosimeter using a 3D-printed phantom
    Tachibana, Hidenobu
    Hoshino, Yoshihiko
    Watanabe, Yusuke
    Usui, Keisuke
    Mizukami, Shinya
    Shibukawa, Shuhei
    Kodama, Takashi
    Tachibana, Rie
    RADIATION PHYSICS AND CHEMISTRY, 2025, 226
  • [25] A Novel Magnetic Resonance Quality Assurance Phantom (KMRP-4) Multi-Site Comparison With the : American College of Radiology Phantom
    Kumar, Suchit
    Kim, Jong-Min
    Cha, Jang Gyu
    Hwang, Ji Young
    Jung, Seung Eun
    Lee, Chulhyun
    Oh, Chang-Hyun
    INVESTIGATIVE MAGNETIC RESONANCE IMAGING, 2023, 27 (01) : 10 - 20
  • [26] Design of a fused phantom for quantitative evaluation of brain metabolites and enhanced quality assurance testing for magnetic resonance imaging and spectroscopy
    Song, Kyu-Ho
    Kim, Sang-Young
    Lee, Do-Wan
    Jung, Jin-Young
    Lee, Jung-Hoon
    Baek, Hyeon-Man
    Choe, Bo-Young
    JOURNAL OF NEUROSCIENCE METHODS, 2015, 255 : 75 - 84
  • [27] QA Procedures for the Assessment of Geometric Accuracy for MRI-SIM Using a Novel Large Field of View Phantom: System Performance Tests and Validation of Clinical Protocols
    Hammond, R. W.
    Perkins, G.
    Paloor, S.
    Al Hammadi, N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S734 - S735
  • [28] Comparison of a Wide Magnet with a Conventional Magnet at 3 Tesla Magnetic Resonance Imaging Based on Signal Intensity Uniformity with Large Field of View: A Phantom Study
    Lee, Ho-Beom
    Jang, Ji-Sung
    Han, Yong-Soo
    Kim, Sung-Min
    JOURNAL OF MAGNETICS, 2019, 24 (03) : 471 - 476
  • [29] Development of a standard phantom for diffusion-weighted magnetic resonance imaging quality control studies: A review
    Manson, Eric Naab
    Mumuni, Abdul Nashirudeen
    Shirazu, Issahaku
    Hasford, Francis
    Inkoom, Stephen
    Sosu, Edem
    Aikins, Mark Pokoo
    Mohammed, Gedel Ahmed
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2022, 28 (04): : 169 - 179
  • [30] Development of a QA phantom and automated analysis tool for geometric quality assurance of on-board MV and kV x-ray imaging systems
    Mao, Weihua
    Lee, Louis
    Xing, Lei
    MEDICAL PHYSICS, 2008, 35 (04) : 1497 - 1506