Impact Assessment of Systemic Geometric Distortion in 1.5T Magnetic Resonance Imaging Simulation through Three-dimensional Geometric Distortion Phantom on Dosimetric Accuracy for Magnetic Resonance Imaging-only Prostate Treatment Planning

被引:0
|
作者
Chaknam, Korawig [1 ]
Worapruekjaru, Ladawan [2 ]
Suphaphong, Sithiphong [2 ]
Stansook, Nualjun [2 ]
Sodkokkruad, Prapa [1 ]
Asavaphatiboon, Sawwanee [1 ]
机构
[1] Mahidol Univ, Fac Med, Dept Diagnost & Therapeut Radiol, Div Diagnost Radiol,Ramathibodi Hosp, 270 Rama Vi Rd, Bangkok 10400, Thailand
[2] Mahidol Univ, Dept Diagnost & Therapeut Radiol, Div Radiat Oncol, Fac Med,Ramathibodi Hosp, Bangkok, Thailand
关键词
Dosimetric accuracy; magnetic resonance imaging-only treatment planning; magnetic resonance imaging-simulation; systemic geometric distortion; RADIATION-THERAPY; CHEMICAL-SHIFT; RADIOTHERAPY; REGISTRATION; MRI; VALIDATION;
D O I
10.4103/jmp.jmp_62_24
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aims: Magnetic resonance imaging (MRI)-only radiotherapy has emerged as a solution to address registration errors that can lead to missed dose delivery. However, the presence of systemic geometric distortion (SGD) stemming from gradient nonlinearity (GNL) and inhomogeneity of the main magnetic field (B0) necessitates consideration. This study aimed to quantitatively assess residual SGD in 1.5T MRI simulation using a three-dimensional (3D) geometric distortion phantom and evaluate its impact on dosimetric accuracy for retrospective prostate cancer patients.Materials and Methods: Ten retrospective cases of prostate cancer patients treated with volumetric modulated arc radiotherapy (VMAT) were randomly selected. A geometric distortion phantom was scanned on a 1.5T MRI simulation using a 3D T1 volumetric interpolated breath-hold examination sequence, varying bandwidth (BW), and two-phase-encoding directions. Distortion maps were generated and applied to the original computed tomography (oriCT) plan to create a distorted computed tomography plan (dCT), and a dice similarity coefficient (DSC) was observed. Dosimetric accuracy was evaluated by recalculating radiation dose for dCT plans using identical beam parameters as oriCT.Results: The SGD increased with distance from the isocenter in all series. DSC exceeded 0.95 for all plans except the rectum. Regarding GNL's impact on dosimetric accuracy, most mean percentage errors for clinical target volume, planning target volume, and both femurs were under 2% in all plans, except for the bladder and rectum.Conclusion: SGD pre-evaluation is crucial and should be incorporated into a quality assurance program to ensure effective MRI-simulation performance before MRI-only treatment planning for prostate cancer.
引用
收藏
页码:356 / 362
页数:7
相关论文
共 50 条
  • [1] Design and construction of a customizable phantom for the characterization of the three-dimensional magnetic resonance imaging geometric distortion
    Torfeh, Tarraf
    Hammoud, Rabih
    Paloor, Satheesh
    Arunachalam, Yoganathan
    Aouadi, Souha
    Al-Hammadi, Noora
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 149 - 157
  • [2] Reproducibility of geometric distortion in magnetic resonance imaging based on phantom studies
    Mizowaki, T
    Nagata, Y
    Okajima, K
    Kokubo, M
    Negoro, Y
    Araki, N
    Hiraoka, M
    RADIOTHERAPY AND ONCOLOGY, 2000, 57 (02) : 237 - 242
  • [3] A proposed scheme for comprehensive characterization of the measured geometric distortion in magnetic resonance imaging using a three-dimensional phantom
    Wang, DM
    Doddrell, DM
    MEDICAL PHYSICS, 2004, 31 (08) : 2212 - 2218
  • [4] Quantification of Geometric Distortion in Magnetic Resonance Imaging for Radiation Therapy Treatment Planning
    Slagowski, J. M.
    Ding, Y.
    Wen, Z.
    Fuller, C. D.
    Chung, C.
    Kadbi, M.
    Ibbott, G. S.
    Wang, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E547 - E547
  • [5] Evaluation of the Geometric and Dosimetric Accuracy of Synthetic Computed Tomography Images for Magnetic Resonance Imaging-only Stereotactic Radiosurgery
    Fatemi, Ali
    Kanakamedala, Madhava R.
    Yang, Claus Chunli
    Morris, Bart
    Duggar, William N.
    Vijayakumar, Srinivasan
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (04)
  • [6] Evaluation of system-related magnetic resonance imaging geometric distortion in radiation therapy treatment planning: two approaches and effectiveness of three-dimensional distortion correction
    Kato, Yutaka
    Okudaira, Kuniyasu
    Kamomae, Takeshi
    Kumagai, Motoki
    Nagai, Youta
    Taoka, Toshiaki
    Itoh, Yoshiyuki
    Naganawa, Shinji
    NAGOYA JOURNAL OF MEDICAL SCIENCE, 2022, 84 (01): : 29 - 41
  • [7] An Anthropomorphic Head Phantom Prototype for the Measurement of Geometric Distortion in Magnetic Resonance Imaging System
    Shurche, Sadegh
    Sooteh, Mohammad Yousefi
    Iranian Journal of Medical Physics, 2024, 21 (01) : 53 - 63
  • [8] A Preliminary Study of the Geometric Distortion in Diffusion-Weighted Imaging of Magnetic Resonance Imaging Simulation
    Liu, X.
    Qiu, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E328 - E329
  • [9] A Preliminary Study of the Geometric Distortion in Diffusion-Weighted Imaging of Magnetic Resonance Imaging Simulation
    Liu, X.
    Qiu, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E510 - E510
  • [10] Distortion inherent to magnetic resonance imaging can lead to geometric miss in radiosurgery planning
    Seibert, Tyler M.
    White, Nathan S.
    Kim, Gwe-Ya
    Moiseenko, Vitali
    McDonald, Carrie R.
    Farid, Nikdokht
    Bartsch, Hauke
    Kuperman, Joshua
    Karunamuni, Roshan
    Marshall, Deborah
    Holland, Dominic
    Sanghvi, Parag
    Simpson, Daniel R.
    Mundt, Arno J.
    Dale, Anders M.
    Hattangadi-Gluth, Jona A.
    PRACTICAL RADIATION ONCOLOGY, 2016, 6 (06) : E319 - E328