Impact Assessment of Systemic Geometric Distortion in 1.5T Magnetic Resonance Imaging Simulation through Three-dimensional Geometric Distortion Phantom on Dosimetric Accuracy for Magnetic Resonance Imaging-only Prostate Treatment Planning

被引:0
|
作者
Chaknam, Korawig [1 ]
Worapruekjaru, Ladawan [2 ]
Suphaphong, Sithiphong [2 ]
Stansook, Nualjun [2 ]
Sodkokkruad, Prapa [1 ]
Asavaphatiboon, Sawwanee [1 ]
机构
[1] Mahidol Univ, Fac Med, Dept Diagnost & Therapeut Radiol, Div Diagnost Radiol,Ramathibodi Hosp, 270 Rama Vi Rd, Bangkok 10400, Thailand
[2] Mahidol Univ, Dept Diagnost & Therapeut Radiol, Div Radiat Oncol, Fac Med,Ramathibodi Hosp, Bangkok, Thailand
关键词
Dosimetric accuracy; magnetic resonance imaging-only treatment planning; magnetic resonance imaging-simulation; systemic geometric distortion; RADIATION-THERAPY; CHEMICAL-SHIFT; RADIOTHERAPY; REGISTRATION; MRI; VALIDATION;
D O I
10.4103/jmp.jmp_62_24
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aims: Magnetic resonance imaging (MRI)-only radiotherapy has emerged as a solution to address registration errors that can lead to missed dose delivery. However, the presence of systemic geometric distortion (SGD) stemming from gradient nonlinearity (GNL) and inhomogeneity of the main magnetic field (B0) necessitates consideration. This study aimed to quantitatively assess residual SGD in 1.5T MRI simulation using a three-dimensional (3D) geometric distortion phantom and evaluate its impact on dosimetric accuracy for retrospective prostate cancer patients.Materials and Methods: Ten retrospective cases of prostate cancer patients treated with volumetric modulated arc radiotherapy (VMAT) were randomly selected. A geometric distortion phantom was scanned on a 1.5T MRI simulation using a 3D T1 volumetric interpolated breath-hold examination sequence, varying bandwidth (BW), and two-phase-encoding directions. Distortion maps were generated and applied to the original computed tomography (oriCT) plan to create a distorted computed tomography plan (dCT), and a dice similarity coefficient (DSC) was observed. Dosimetric accuracy was evaluated by recalculating radiation dose for dCT plans using identical beam parameters as oriCT.Results: The SGD increased with distance from the isocenter in all series. DSC exceeded 0.95 for all plans except the rectum. Regarding GNL's impact on dosimetric accuracy, most mean percentage errors for clinical target volume, planning target volume, and both femurs were under 2% in all plans, except for the bladder and rectum.Conclusion: SGD pre-evaluation is crucial and should be incorporated into a quality assurance program to ensure effective MRI-simulation performance before MRI-only treatment planning for prostate cancer.
引用
收藏
页码:356 / 362
页数:7
相关论文
共 50 条
  • [31] Feasibility of three-dimensional ultrashort echo time magnetic resonance imaging at 1.5 T for the diagnosis of skull fractures
    Hao Wu
    Yu-min Zhong
    Quan-min Nie
    Wei-bo Chen
    Lie-mei Guo
    Xi Yang
    Hong Zhang
    Yi Lin
    Jian-rong Xu
    Yong-ming Dai
    Ming Zhu
    European Radiology, 2016, 26 : 138 - 146
  • [32] Feasibility of three-dimensional ultrashort echo time magnetic resonance imaging at 1.5 T for the diagnosis of skull fractures
    Wu, Hao
    Zhong, Yu-min
    Nie, Quan-min
    Chen, Wei-bo
    Guo, Lie-mei
    Yang, Xi
    Zhang, Hong
    Lin, Yi
    Xu, Jian-rong
    Dai, Yong-ming
    Zhu, Ming
    EUROPEAN RADIOLOGY, 2016, 26 (01) : 138 - 146
  • [33] Geometric determinants of chronic functional ischemic mitral regurgitation: Insights from three-dimensional cardiac magnetic resonance imaging
    Jensen, Henrik
    Jensen, Morten O.
    Ringgaard, Steffen
    Smerup, Morten H.
    Sorensen, Thomas S.
    Kim, Won Y.
    Sloth, Erik
    Wierup, P.
    Hasenkam, J. Michael
    Nielsen, Sten L.
    JOURNAL OF HEART VALVE DISEASE, 2008, 17 (01): : 16 - 22
  • [34] Three-dimensional nuclear magnetic resonance spectroscopy: a complementary tool to multiparametric magnetic resonance imaging in the identification of aggressive prostate cancer at 3.0T
    Deal, Michael
    Bardet, Florian
    Walker, Paul-Michael
    de la Vega, Mathilde Funes
    Cochet, Alexandre
    Cormier, Luc
    Bentellis, Imad
    Loffroy, Romaric
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2021, 11 (08) : 3749 - 3766
  • [35] Geometric determinants of chronic ischemic mitral regurgitation in patients with inferior myocardial infarction: Three-dimensional magnetic resonance imaging study
    Kaji, S
    Yamamuro, A
    Tamita, K
    Kinoshita, M
    Tani, T
    Nagai, K
    Tanabe, K
    Kihara, Y
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (04) : 278A - 278A
  • [36] Accuracy in local staging of prostate cancer by adding a three-dimensional T2-weighted sequence with radial reconstructions in magnetic resonance imaging
    Jaderling, Fredrik
    Nyberg, Tommy
    Oberg, Michael
    Carlsson, Stefan
    Skorpil, Mikael
    Blomqvist, Lennart
    ACTA RADIOLOGICA OPEN, 2018, 7 (02):
  • [37] Stereotactic localization with magnetic resonance imaging: A phantom study to compare the accuracy obtained using two-dimensional and three-dimensional data acquisitions
    Walton, L
    Hampshire, A
    Forster, DMC
    Kemeny, AA
    NEUROSURGERY, 1997, 41 (01) : 131 - 137
  • [38] Development of fetal brain sulci and gyri: assessment through two and three-dimensional ultrasound and magnetic resonance imaging
    Liliam Cristine Rolo
    Edward Araujo
    Luciano Marcondes Machado Nardozza
    Patrícia Soares de Oliveira
    Sérgio Aron Ajzen
    Antonio Fernandes Moron
    Archives of Gynecology and Obstetrics, 2011, 283 : 149 - 158
  • [39] Development of fetal brain sulci and gyri: assessment through two and three-dimensional ultrasound and magnetic resonance imaging
    Rolo, Liliam Cristine
    Araujo, Edward, Jr.
    Machado Nardozza, Luciano Marcondes
    de Oliveira, Patricia Soares
    Ajzen, Sergio Aron
    Moron, Antonio Fernandes
    ARCHIVES OF GYNECOLOGY AND OBSTETRICS, 2011, 283 (02) : 149 - 158
  • [40] Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test
    Wong, Oi Lei
    Yuan, Jing
    Yu, Siu Ki
    Cheung, Kin Yin
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2017, 7 (02) : 205 - 214