Clinical implementation of a commercial synthetic computed tomography solution for radiotherapy treatment of glioblastoma

被引:2
|
作者
Emin, Sevgi [1 ]
Rossi, Elia [2 ]
Rooth, Elisabeth Myrvold [2 ]
Dorniok, Torsten [1 ]
Hedman, Mattias [2 ,3 ]
Gagliardi, Giovanna [1 ,3 ]
Villegas, Fernanda [1 ,3 ]
机构
[1] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Radiat Oncol, S-17176 Stockholm, Sweden
[3] Karolinska Inst, Dept Oncol Pathol, S-17177 Stockholm, Sweden
关键词
MR-only; Glioblastoma; Radiotherapy; Synthetic-CT; ACCURACY; MRI;
D O I
10.1016/j.phro.2024.100589
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and Purpose: Magnetic resonance (MR)-only radiotherapy (RT) workflow eliminates uncertainties due to computed tomography (CT)-MR image registration, by using synthetic CT (sCT) images generated from MR. This study describes the clinical implementation process, from retrospective commissioning to prospective validation stage of a commercial artificial intelligence (AI)-based sCT product. Evaluation of the dosimetric performance of the sCT is presented, with emphasis on the impact of voxel size differences between image modalities. Materials and methods: sCT performance was assessed in glioblastoma RT planning. Dose differences for 30 patients in both commissioning and validation cohorts were calculated at various dose-volume-histogram (DVH) points for target and organs-at-risk (OAR). A gamma analysis was conducted on regridded image plans. Quality assurance (QA) guidelines were established based on commissioning phase results. Results: Mean dose difference to target structures was found to be within +/- 0.7 % regardless of image resolution and cohort. OARs' mean dose differences were within +/- 1.3 % for plans calculated on regridded images for both cohorts, while differences were higher for plans with original voxel size, reaching up to -4.2 % for chiasma D2% in the commissioning cohort. Gamma passing rates for the brain structure using the criteria 1 %/1mm, 2 %/2mm and 3 %/3mm were 93.6 %/99.8 %/100 % and 96.6 %/99.9 %/100 % for commissioning and validation cohorts, respectively. Conclusions: Dosimetric outcomes in both commissioning and validation stages confirmed sCT's equivalence to CT. The large patient cohort in this study aided in establishing a robust QA program for the MR-only workflow, now applied in glioblastoma RT at our center.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Computed Tomography/Magnetic Resonance Imaging (CT/MRI) Image Registration and Fusion Assessment for Accurate Glioblastoma Radiotherapy Treatment Planning
    Touabti, Khalil Mohamed Mokhtar
    Kharfi, Faycal
    Benkahila, Karim
    Merouane, Sid-Ali
    INTERNATIONAL JOURNAL OF CANCER MANAGEMENT, 2020, 13 (09) : 1 - 8
  • [22] RADIOTHERAPY IN THE MANAGEMENT OF THYROID ORBITOPATHY - COMPUTED-TOMOGRAPHY AND CLINICAL OUTCOMES
    KAO, SCS
    KENDLER, DL
    NUGENT, RA
    ADLER, JS
    ROOTMAN, J
    ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (06) : 819 - 823
  • [23] Applying a commercial atlas-based synthetic Computed Tomography algorithm to patients with hip prostheses for prostate Magnetic Resonance-only radiotherapy
    Wyatt, Jonathan
    McCallum, Hazel
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : 100 - 105
  • [24] Technical Note: Volumetric computed tomography for radiotherapy simulation and treatment planning
    Young, Heather M.
    Park, Claire K.
    Chau, Oi-Wai
    Lee, Ting-Yim
    Gaede, Stewart
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 295 - 302
  • [25] COMPUTED-TOMOGRAPHY APPLIED TO RADIOTHERAPY TREATMENT PLANNING - TECHNIQUES AND RESULTS
    HOBDAY, P
    HODSON, NJ
    HUSBAND, J
    PARKER, RP
    MACDONALD, JS
    RADIOLOGY, 1979, 133 (02) : 477 - 482
  • [26] UTILIZATION OF COMPUTED-TOMOGRAPHY FOR QUANTITATIVE DESCRIPTION OF RADIOTHERAPY TREATMENT PLANS
    PRASAD, SC
    PILEPICH, MV
    PEREZ, CA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1980, 6 (10): : 1446 - 1446
  • [27] Clinical applications of positron emission tomography/computed tomography treatment planning
    Macapinlac, Homer A.
    SEMINARS IN NUCLEAR MEDICINE, 2008, 38 (02) : 137 - 140
  • [28] Recommendations of megavoltage computed tomography settings for the implementation of adaptive radiotherapy on helical tomotherapy units
    Velten, Christian
    Boyd, Robert
    Jeong, Kyoungkeun
    Garg, Madhur K.
    Tome, Wolfgang A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (05): : 87 - 92
  • [29] Recommendations of Megavoltage Computed Tomography Settings for the Implementation of Adaptive Radiotherapy on Helical Tomotherapy Units
    Velten, C.
    Boyd, R.
    Jeong, K.
    Kalnicki, S.
    Garg, M. K.
    Tome, W. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E789 - E789
  • [30] Magnetic Resonance-Based Synthetic Computed Tomography Using Generative Adversarial Networks for Intracranial Tumor Radiotherapy Treatment Planning
    Wang, Chun-Chieh
    Wu, Pei-Huan
    Lin, Gigin
    Huang, Yen-Ling
    Lin, Yu-Chun
    Chang, Yi-Peng
    Weng, Jun-Cheng
    JOURNAL OF PERSONALIZED MEDICINE, 2022, 12 (03):