Utilization of a two-material decomposition from a single-source, dual-energy CT in acute traumatic vertebral fractures

被引:0
|
作者
Tivnan, Patrick [1 ]
Kaliaev, Artem [1 ]
Anderson, Stephan W. [1 ]
LeBedis, Christina A. [1 ]
Li, Baojun [1 ]
Andreu-Arasa, V. Carlota [1 ]
机构
[1] Boston Univ, Sch Med, Dept Radiol, Boston, MA 02215 USA
来源
FRONTIERS IN RADIOLOGY | 2023年 / 3卷
关键词
fracture; bone marrow edema; dual-energy computed tomography; magnetic resonance imaging; calcium; fat density; water density; BONE-MARROW EDEMA; MULTIDETECTOR CT; COMPUTED-TOMOGRAPHY; DIAGNOSTIC-ACCURACY; INJURIES; DISPLAY;
D O I
10.3389/fradi.2023.1187449
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study is to utilize a two-material decomposition to quantify bone marrow edema on a dual-energy computed tomography (DECT) scanner at the cervical, thoracic, and lumbar spine acute fractures in correlation with short tau inversion recovery (STIR) hyperintensity on magnetic resonance imaging (MRI) in comparison with the normal bone marrow. Materials and methods: This retrospective institutional review board-approved study gathered patients over 18 years old who had acute cervical, thoracic, or lumbar spinal fractures scanned on a DECT scanner. Those who had a spinal MRI done with bone marrow STIR hyperintensity within 3 weeks of the DECT were included. The water (calcium) and fat (calcium) density (mg/cm(3)) measurements of the region of interest of the bone marrow were obtained at a normal anatomic equivalent site and at the fracture site where STIR hyperintensity was noted on MRI. A statistical analysis was performed using the paired t-test and Wilcoxon signed rank test (p > 0.05). Results: A total of 20 patients met the inclusion criteria (males n = 17 males, females n = 3). A total of 32 fractures were analyzed: 19 cervical and 13 thoracolumbar. There were statistically significant differences in the water (43 +/- 24 mg/cm(3)) and fat (36 +/- 31 mg/cm(3)) density (mg/cm(3)) at the acute thoracic and lumbar spine fractures in correlation with edema on STIR images (both paired t-test <0.001, both Wilcoxon signed ranked test p < 0.01). There were no significant differences in the water (-10 +/- 46 mg/cm(3)) or fat (+7 +/- 50 mg/cm(3)) density (mg/cm(3)) at the cervical spine fractures. Conclusion: The DECT two-material decomposition using water (calcium) and fat (calcium) analyses has the ability to quantify a bone marrow edema at the acute fracture site in the thoracic and lumbar spine.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Two-Material Decomposition Algorithm of Dual-Energy CT Based on Gradient Descent Method
    Teng Y.-Y.
    Zheng S.-Y.
    Lu Z.-P.
    Kang Y.
    Dongbei Daxue Xuebao/Journal of Northeastern University, 2017, 38 (07): : 936 - 940
  • [3] Comparison of Dual- and Single-Source Dual-Energy CT for Diagnosis of Acute Pulmonary Artery Embolism
    Petritsch, Bernhard
    Pannenbecker, Pauline
    Weng, Andreas Max
    Veldhoen, Simon
    Grunz, Jan-Peter
    Bley, Thorsten Alexander
    Kosmala, Aleksander
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2021, 193 (04): : 427 - 436
  • [4] Clinical Implementation of Single-Source, Dual-Energy CT Simulation in Radiotherapy
    Smith, A.
    Hyun, M.
    Zheng, D.
    MEDICAL PHYSICS, 2018, 45 (06) : E148 - E148
  • [5] Dual-Energy CT of Material Decomposition Analysis for Detection with Bone Marrow Edema in Patients with Vertebral Compression Fractures
    Akisato, Kyouhei
    Nishihara, Reisuke
    Okazaki, Hajime
    Masuda, Takanori
    Hironobe, Ayako
    Ishizaki, Hiromi
    Shota, Kondo
    Yamaguchi, Hiroshi
    Funama, Yoshinori
    ACADEMIC RADIOLOGY, 2020, 27 (02) : 227 - 232
  • [6] Comparison of dual- and single-source dual-energy CT in head and neck imaging
    Matthias Stefan May
    Marco Wiesmueller
    Rafael Heiss
    Michael Brand
    Joscha Bruegel
    Michael Uder
    Wolfgang Wuest
    European Radiology, 2019, 29 : 4207 - 4214
  • [7] Comparison of dual- and single-source dual-energy CT in head and neck imaging
    May, Matthias Stefan
    Wiesmueller, Marco
    Heiss, Rafael
    Brand, Michael
    Bruegel, Joscha
    Uder, Michael
    Wuest, Wolfgang
    EUROPEAN RADIOLOGY, 2019, 29 (08) : 4207 - 4214
  • [8] Calculation of Two-Material Tissue Parameterizations for Dual-Energy Imaging
    Jennings, R.
    MEDICAL PHYSICS, 2016, 43 (06) : 3328 - 3329
  • [9] Material differentiation in forensic radiology with single-source dual-energy computed tomography
    Thomas D. Ruder
    Yannick Thali
    Stephan A. Bolliger
    Sandra Somaini-Mathier
    Michael J. Thali
    Gary M. Hatch
    Sebastian T. Schindera
    Forensic Science, Medicine, and Pathology, 2013, 9 : 163 - 169
  • [10] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Nakhostin, Dominik
    Sartoretti, Thomas
    Eberhard, Matthias
    Krauss, Bernhard
    Mueller, Daniel
    Alkadhi, Hatem
    Euler, Andre
    ABDOMINAL RADIOLOGY, 2021, 46 (05) : 2079 - 2089