Longitudinal Observation of Micromotion upon Loading of Implant-Abutment Connection

被引:0
|
作者
Yamashita, Kohei [1 ]
Kataoka, Yu [2 ,3 ]
Munakata, Motohiro [4 ]
Yamaguchi, Kikue [4 ]
Hayashi, Myu [5 ]
Baba, Daisuke [6 ]
机构
[1] Showa Univ, Grad Sch Dent, Dept Implant Dent, 2-1-1 Kita Senzoku,Ota Ku, Tokyo 1458515, Japan
[2] Showa Univ, Sch Dent, Dept Dent Educ, 1-5-8 Hatanodai,Shinagawa Ku, Tokyo 1428555, Japan
[3] Showa Univ, Dept Biomat & Engn, Sch Dent, 1-5-8 Hatanodai,Shinagawa Ku, Tokyo 1428555, Japan
[4] Showa Univ, Sch Dent, Dept Implant Dent, 2-1-1 Kita Senzoku,Ohta Ku, Tokyo 1458515, Japan
[5] Showa Univ, Sch Dent, Dept Oral Hlth Management, Div Oral Funct Management, 2-1-1 Kita Senzoku,Ota Ku, Tokyo 1458515, Japan
[6] Showa Univ, Sch Dent, Dept Dent Lab, 2-1-1 Kita Senzoku,Ota Ku, Tokyo 1458515, Japan
来源
BIOENGINEERING-BASEL | 2024年 / 11卷 / 06期
关键词
implant loading; micromotion; peri-implantitis; implant-abutment connection; removal torque;
D O I
10.3390/bioengineering11060582
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
While technological advances have made implants a good treatment option with a good long-term prognosis, peri-implantitis, which results in alveolar bone resorption around implants, has been observed in some cases. Micromotion at the implant abutment connection can cause peri-implantitis. However, the temporal progression of micromotion upon loading remains unclear. Therefore, we aimed to longitudinally measure micromotion upon loading application on an implant. Implants with Morse-tapered connections were prepared. Custom titanium abutments were fabricated and tightened onto implant bodies at 35 N. A 100 N vertical load was applied for 200,000 cycles. Micromotion was measured when the load was applied, as was the total implant length and removal torque before and after loading. The micromotion was measured from the position data of the jig of the testing machine during loading. The average removal torque was 30.67 N after 10 min of tightening and 27.95 N after loading, indicating a decrease due to loading. The implant length reduced by 3.6 mu m under the load. The average micromotion was 0.018 mm at 2 cycles, 0.016 mm at 100,000 cycles, and 0.0157 mm at 200,000 cycles, indicating implant length reduction under the load but not reaching 0. The micromotion between the implant and abutment under a cyclic load decreased over time but did not completely cease. These results highlight the relationship between micromotion and loading, underscoring the importance of careful monitoring and management to mitigate potential complications, such as peri-implantitis, and ensure optimal performance and durability of the implant.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] The survival of dental implants with different implant-abutment connection systems
    Szpak, Piotr
    Szymanska, Jolanta
    CURRENT ISSUES IN PHARMACY AND MEDICAL SCIENCES, 2016, 29 (01) : 11 - 13
  • [22] THE ROLE OF IMPLANT-ABUTMENT CONNECTION IN PREVENTING BACTERIAL LEAKAGE: A REVIEW
    Candotto, V
    Gabrione, F.
    Oberti, L.
    Lento, D.
    Severino, M.
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2019, 33 (03): : 129 - 134
  • [23] Essay 3: Biologic and mechanical principals of the implant-abutment connection
    不详
    INTERNATIONAL JOURNAL OF ESTHETIC DENTISTRY, 2011, 6 (02): : 246 - 251
  • [24] Correlation between microleakage and screw loosening at implant-abutment connection
    Sahin, Cem
    Ayyildiz, Simel
    JOURNAL OF ADVANCED PROSTHODONTICS, 2014, 6 (01): : 35 - 38
  • [25] Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection
    Semper-Hogg, Wiebke
    Kraft, Silvan
    Stiller, Sebastian
    Mehrhof, Juergen
    Nelson, Katja
    CLINICAL ORAL INVESTIGATIONS, 2013, 17 (03) : 1017 - 1023
  • [26] Micro-CT Evaluation of Microgaps at Implant-Abutment Connection
    Kowalski, Jakub
    Puszkarz, Adam K.
    Radwanski, Mateusz
    Sokolowski, Jerzy
    Cichomski, Michal
    Bourgi, Rim
    Hardan, Louis
    Sauro, Salvatore
    Lukomska-Szymanska, Monika
    MATERIALS, 2023, 16 (12)
  • [27] Effects of different implant-abutment connections on micromotion and stress distribution: Prediction of microgap formation
    Saidin, Syafiqah
    Kadir, Mohammed Rafiq Abdul
    Sulaiman, Eshamsul
    Abu Kasim, Noor Hayaty
    JOURNAL OF DENTISTRY, 2012, 40 (06) : 467 - 474
  • [28] MICROSCOPE EVALUATION OF MICRO-GAPS AT IMPLANT-ABUTMENT CONNECTION
    Tartea, Daniel Adrian
    Popescu, Sanda Mihaela
    Manolea, Horia Octavian
    Ionescu, Mihaela
    Amarascu, Olimpia Marina
    Marasescu, Petre Costin
    Gingu, Oana
    ROMANIAN JOURNAL OF ORAL REHABILITATION, 2024, 16 (02): : 411 - 423
  • [29] Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems
    Jo, Jae-Young
    Yang, Dong-Seok
    Huh, Jung-Bo
    Heo, Jae-Chan
    Yun, Mi-Jung
    Jeong, Chang-Mo
    JOURNAL OF ADVANCED PROSTHODONTICS, 2014, 6 (06): : 491 - 497
  • [30] The behavior of a zirconia or metal abutment on the implant-abutment interface during cyclic loading
    Queiroz, Daher Antonio
    Hagee, Natalia
    Lee, Damian J.
    Zheng, Fengyuan
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (02): : 211 - 216