Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system

被引:0
|
作者
Lin, Yuting [1 ]
Li, Wangyao [1 ]
Wang, Aoxiang [1 ,2 ]
Johnson, Daniel [1 ]
Gan, Gregory N. [1 ]
Gao, Hao [1 ]
机构
[1] Univ Kansas, Med Ctr, Dept Radiat Oncol, Kansas City, KS 66103 USA
[2] Huazhong Univ Sci & Technol, Dept Biomed Engn, Wuhan, Peoples R China
来源
FRONTIERS IN ONCOLOGY | 2024年 / 14卷
关键词
proton therapy; spatially fractionated radiotherapy; proton mini-beam; small field dosimetry; film dosimetry; proton dosimetry; FRACTIONATED RADIATION-THERAPY;
D O I
10.3389/fonc.2024.1421869
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Proton minibeam radiation therapy (pMBRT) can deliver spatially fractionated dose distributions with submillimeter resolution. These dose distributions exhibit significant heterogeneity in both depth and lateral directions. Accurate characterization of pMBRT doses requires dosimetry devices with high spatial resolution and a wide dynamic range. Furthermore, the dependency of dosimetric measurements on Linear Energy Transfer (LET), as observed in conventional proton therapy, is also present in pMBRT depth dose measurements.Purpose This work demonstrates the process of performing comprehensive dosimetric measurements to characterize the pMBRT collimator on a clinical single-gantry proton machine, utilizing commercially available dosimetry devices.Methods The minibeam collimator is designed to be mounted on the clinical nozzle as a beam-modifying accessory. Three collimators, each with a slit opening of 0.4 mm, are thoroughly evaluated. The center-to-center (c-t-c) distances of the slits for these collimators are 2.8 mm, 3.2 mm, and 4.0 mm, respectively. High spatial resolution dosimetry devices are essential for PMBRT dose characterizations. To meet this requirement, two-dimensional (2D) dose measurement devices, Gafchromic films, are used to measure lateral profiles at various depths. Films are also used for depth dose profile measurements in solid water. Additionally, high-resolution point dose detectors, microDiamond, and Razor diode detectors are employed for lateral profile measurements at various depths. Percent depth dose (PDD) measurements of pMBRT in solid water, with various proton energies, collimators, and air gaps, are performed using Gafchromic films. The film's LET dependency for proton beams is corrected to ensure accurate pMBRT PDD measurements. The Monte Carlo simulation tool TOPAS is utilized to compare and validate all experimental measurements.Results At depths where LET is not a concern, film dose measurements were consistent with microDiamond and Razor diode point measurements. The point detectors need to be orientated with the thin side aligned to the incoming beam. Comparison of the lateral dose profiles extracted from TOPAS simulations, films, microDiamond, and Razor diode detectors shows a passing rate exceeding 98% in 1D gamma analysis at 3% 0.1 mm criteria.However, when the microDiamond detector is orientated to face the pMBRT beam, its spatial resolution may not be sufficient to capture the peak and valley dose accurately. Nevertheless, an accuracy within 2% can still be achieved when comparing the average dose. The PDD measurements show that the peak valley dose ratio (PVDR) of pMBRT can be altered at different depths with different air gaps using the same collimator or different collimators of different c-t-c distances.Conclusion Our study demonstrates that comprehensive dose measurements for pMBRT can be conducted using standard clinical dose measurement devices. These measurements are indispensable for guiding and ensuring accurate dose reporting in pre-clinical studies using the pMBRT technique.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Dosimetric characteristics of the ocular beam line and commissioning data for an ocular proton therapy planning system at the Proton Therapy Center Houston
    Titt, Uwe
    Suzuki, Kazumichi
    Li, Yupeng
    Sahoo, Narayan
    Gillin, Michael T.
    Zhu, Xiaorong R.
    MEDICAL PHYSICS, 2017, 44 (12) : 6661 - 6671
  • [32] Commissioning and beam characterization of the first gantry-mounted accelerator pencil beam scanning proton system
    Kang, M.
    Cessac, Rob
    Pang, D.
    MEDICAL PHYSICS, 2020, 47 (08) : 3496 - 3510
  • [33] Multi-collimator proton minibeam radiotherapy with joint dose and PVDR optimization
    Zhang, Weijie
    Hong, Xue
    Wu, Wei
    Wang, Chao
    Johnson, Daniel
    Gan, Gregory N.
    Lin, Yuting
    Gao, Hao
    MEDICAL PHYSICS, 2025, 52 (02) : 1182 - 1192
  • [34] The Beam Commissioning of SC Cyclotron-Based Proton Therapy System With Fast Energy Selection and Precise Gantry
    Zhang, Tianjue
    Wang, Chuan
    Yin, Zhiguo
    Jia, Xianlu
    Wang, Yang
    Fu, Wei
    Liu, Jingyuan
    Jin, Zhaojun
    Zhou, Hongji
    Zhang, Suping
    Cai, Hongru
    Pan, Gaofeng
    Zhu, Xiaofeng
    Lin, Jun
    Zhao, Bohan
    Guo, Qiankun
    Liu, Chuanye
    Xiong, Rui
    He, Aolai
    Jiang, Tianyi
    Hu, Mingzhi
    Song, Qiqi
    Ding, Aoxuan
    Jiang, Shangmeng
    Liu, Ping
    Yang, Fangdi
    Huang, Xiesi
    Chu, Zhichao
    Xia, Xiaoxue
    Zhang, Zheyuan
    Qin, Ling
    Ma, Yongjun
    Cheng, Ziyi
    Xu, Zheng
    Le, Hao
    Zhang, Youwen
    Wang, Xi
    Liu, Zhan
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2025, 72 (03) : 438 - 445
  • [35] Proton radiography for the clinical commissioning of the new Gantry2 head support at PSI
    Placidi, L.
    Konig, S.
    Van der Meer, R.
    Gagnon-Moisan, F.
    Lomax, A. J.
    Weber, D. C.
    Bolsi, A.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S380 - S380
  • [36] Commissioning of a Single-Room Double-Scattering Proton Therapy System
    Jin, H.
    Ahmad, S.
    Chen, Y.
    Keeling, V.
    Lau, A.
    Islam, M.
    Ferreira, C.
    Ferguson, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3498 - 3499
  • [37] Unlocking a closed system: dosimetric commissioning of a ring gantry linear accelerator in a multivendor environment
    Saini, Amarjit
    Tichacek, Chris
    Johansson, William
    Redler, Gage
    Zhang, Geoffrey
    Moros, Eduardo G.
    Qayyum, Muqeem
    Feygelman, Vladimir
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (02): : 21 - 34
  • [38] Commissioning of the First Hyperscan PBS Proton System
    Kang, M.
    Chen, H.
    Pang, D.
    MEDICAL PHYSICS, 2018, 45 (06) : E432 - E432
  • [39] Commissioning and performance of the CMS calorimeter system with proton-proton collisions at the LHC
    Gras, Philippe
    35TH INTERNATIONAL CONFERENCE OF HIGH ENERGY PHYSICS (ICHEP 2010), 2010,
  • [40] Adaptation and dosimetric commissioning of a synchrotron-based proton beamline for FLASH experiments
    Yang, Ming
    Wang, Xiaochun
    Guan, Fada
    Titt, Uwe
    Iga, Kiminori
    Jiang, Dadi
    Takaoka, Takeshi
    Tootake, Satoshi
    Katayose, Tadashi
    Umezawa, Masumi
    Schueler, Emil
    Frank, Steven
    Lin, Steven H.
    Sahoo, Narayan
    Koong, Albert C.
    Mohan, Radhe
    Zhu, X. Ronald
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (16):