Dosimetric characteristics of transparent bolus for external beam radiotherapy

被引:0
|
作者
Govindaraj K. [1 ,2 ]
Senthilkumar S. [2 ,3 ]
Jeevan Ram R.K. [4 ]
机构
[1] Department of Radiotherapy, Vadamalayan Integrated Cancer Centre, Madurai
[2] Research and Development Centre, Bharathiar University, Coimbatore
[3] Department of Radiotherapy, Madurai Medical College and Government Rajaji Hospital, Madurai
[4] Department of Medical Physics, Bharathidasan University, Tiruchirapalli
关键词
Bolus; Radiotherapy; Surface dose;
D O I
10.22038/ijmp.2019.37129.1470
中图分类号
学科分类号
摘要
Introduction: In radiotherapy, the bolus is often used while treating the tumor under the uneven surfaces of the patients for correcting the anatomical irregularities and increasing skin dose. Wet cotton and wet gauze are still used in developing countries, since the use of wet cotton and wet gauze has certain disadvantages, there is a need for transparent bolus which should be similar to a universally accepted bolus in terms of properties with a lower cost (50% less expensive). Material and Methods: The present study was conducted to investigate the characteristics of transparent bolus (Senflab) material, such as transmission factor, percentage depth dose (PDD), stability over time and high dose, homogeneity and transparency using 6 and 15 MV photons beam and 12, 15, and 18 MeV electrons beam. Moreover, the new bolus material was compared with those of the commercially available Superflab and RW3 slab. Results: The percentage difference in the transmission factor of Senflab was less than ±1.9%, compared with Superflab and RW3. For PDD, the percentage difference was ±2.88% and ±1.26% for photons and electron beams, respectively. The performance of bolus remained constant both physically and dosimetrically after higher dose exposure. The percentage standard deviation was 0.0002% for a period of one month, and 0.0003% for the homogeneity. The transparency of the bolus material was good enough to display the set radiation treatment field. Conclusion: This study shows the suitability of the new bolus for routine use in radiotherapy. © 2019, Mashhad University of Medical Sciences.
引用
收藏
页码:410 / 415
页数:5
相关论文
共 50 条
  • [1] Evaluation of dosimetric properties of 3D printed flat bolus for external beam radiotherapy
    Ricotti, R.
    Vavassori, A.
    Spoto, R.
    Ciardo, D.
    Pansini, F.
    Bazani, A.
    Noris, S.
    Cattani, F.
    Orecchia, R.
    Jereczek-Fossa, B. A.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S923 - S924
  • [2] Dosimetric characterization of 3D printed bolus at different infill percentage for external photon beam radiotherapy
    Ricotti, Rosalinda
    Ciardo, Delia
    Pansini, Floriana
    Bazani, Alessia
    Comi, Stefania
    Spoto, Ruggero
    Noris, Samuele
    Cattani, Federica
    Baroni, Guido
    Orecchia, Roberto
    Vavassori, Andrea
    Jereczek-Fossa, Barbara Alicja
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 39 : 25 - 32
  • [3] The dosimetric consequences of delineation variation for cervical external beam radiotherapy
    Eminowicz, G.
    Rompokos, V.
    Stacey, C.
    McCormack, M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S455 - S456
  • [4] Dosimetric effect of respiratory motion in external beam radiotherapy of the lung
    Mechalakos, J
    Yorke, E
    Mageras, GS
    Hertanto, A
    Jackson, A
    Obcemea, C
    Rosenzweig, K
    Ling, CC
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 (02) : 191 - 200
  • [5] Dosimetric assessment of bolus for postmastectomy radiotherapy
    Kawamoto, Terufumi
    Shikama, Naoto
    Kurokawa, Chie
    Hara, Naoya
    Oshima, Masaki
    Sasai, Keisuke
    MEDICAL DOSIMETRY, 2021, 46 (01) : e1 - e4
  • [6] Physical and Dosimetric Characterization of Silicone Rubber Bolus for Head Photon-Beam Radiotherapy
    Chen, Xuemei
    Wang, Wei
    Huang, Xiaotong
    Diao, Wenchao
    Xu, Chendi
    Jia, Lecheng
    Li, Hua
    Li, Bin
    Jiang, Xiaobo
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2024, 23
  • [7] DOSIMETRIC IMPLICATIONS OF UNCORRECTED ROTATION OF THE PROSTATE DURING EXTERNAL BEAM RADIOTHERAPY
    Burke, K.
    Lamb, C.
    Rosenfelder, N.
    Bidmead, M.
    van As, N.
    Khoo, V.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S473 - S474
  • [8] External beam radiotherapy and brachytherapy for partial breast irradiation: dosimetric comparison
    Irina, A.
    Sergey, N.
    Zhanna, B.
    Julia, M.
    Sergey, K.
    RADIOTHERAPY AND ONCOLOGY, 2021, 158 : S131 - S131
  • [9] Dosimetric feasibility of MRI-guided external beam radiotherapy of the kidney
    Stam, Mette K.
    van Vulpen, Marco
    Barendrecht, Maurits M.
    Zonnenberg, Bernard A.
    Crijns, Sjoerd P. M.
    Lagendijk, Jan Jw
    Raaymakers, Bas W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (14): : 4933 - 4941
  • [10] Dosimetric characteristics of self-expandable metallic and plastic stents for transpapillary biliary decompression in external beam radiotherapy
    Ueda, Yoshihiro
    Ikezawa, Kenji
    Sagawa, Tomohiro
    Isono, Masaru
    Ohira, Shingo
    Miyazaki, Masayoshi
    Takada, Ryoji
    Yamai, Takuo
    Ohkawa, Kazuyoshi
    Teshima, Teruki
    Konishi, Koji
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (04) : 1323 - 1335