Preliminary Values of Diagnostic Reference Level for Selected X-Ray Examinations in Indonesia

被引:0
|
作者
Hiswara, E. [1 ]
Prasetio, H. [1 ]
Kartikasari, D. [1 ]
Nuraeni, N. [1 ]
Sandy, K. Y. P. [1 ]
机构
[1] Natl Nucl Energy Agcy, Ctr Radiat Safety Technol & Metrol, Jl Lebak Bulus Raya 49, Jakarta 12070, Indonesia
关键词
Diagnostic radiology; Diagnostic reference levels; Adult patient doses; Entrance surface doses; Children patient doses;
D O I
10.17146/aij.2016.515
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The diagnostic reference levels form an efficient, concise, and powerful standard for optimizing the radiation protection of a patient. With an aim to contribute toward the establishment of the Indonesian National Diagnostic Reference Levels (NDRLs), a nationwide survey of the entrance surface doses received by adult patients during the most typical X-ray examinations has been performed. A number of 44 hospitals in 21 cities located in Java, Bali, Sumatera, Kalimantan, and Sulawesi islands were selected randomly to participate in this survey. Eight most common adult X-ray examinations in 13 projections, as well as four children X-ray examinations in six projections, were included in the list of procedures under consideration. Hospitals of different sizes and levels using different X-ray machines were represented in the survey. Standard thermoluminescence dosemeters were applied to measure entrance surface dose (ESD). A total of 1493 patients, consisting of 1208 adults and 285 children, were included in this study. The data were analyzed statistically and the minimum, median, mean, maximum, first quartile, and third quartile values of ESDs were reported. The ESDs calculated (third quartile) for adults varied from 0.18 mGy (for extremities AP) to 5.84 mGy (for lumbar spine LAT), and for children they varied from 0.16 mGy (for chest AP/PA) to 1.46 mGy (for skull AP/PA). Considering the geographic spread and size of Indonesia, those third quartile values calculated can only be regarded as preliminary DRL values for Indonesia. Compared with data from other countries, the calculated ESDs in this study are in general lower than the ESDs in those countries. (C) 2016 Atom Indonesia. All rights reserved
引用
下载
收藏
页码:99 / 104
页数:6
相关论文
共 50 条
  • [31] UNNECESSARY X-RAY EXAMINATIONS
    BRANSBYZACHARY, MAP
    SUTHERLAND, GR
    BRITISH MEDICAL JOURNAL, 1989, 298 (6683): : 1294 - 1294
  • [32] ANNUAL X-RAY EXAMINATIONS
    不详
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1951, 146 (06): : 608 - 608
  • [33] X-RAY EXAMINATIONS IN PREGNANCY
    不详
    BMJ-BRITISH MEDICAL JOURNAL, 1965, 2 (5463): : 692 - +
  • [34] EXAMINATIONS ON EFFECTIVE FOCAL SPOT SIZE IN DIAGNOSTIC X-RAY TUBES
    ANDERSSON, O
    RADIOLOGIA DIAGNOSTICA, 1974, 15 (04) : 535 - 542
  • [35] QUALITY OF DISTURBED RADIATION TO OVARIES DURING DIAGNOSTIC X-RAY EXAMINATIONS
    EWEN, K
    FIEBACH, BJO
    FISCHER, PG
    LOHR, E
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1977, 126 (02): : 165 - 168
  • [36] UNNECESSARY X-RAY EXAMINATIONS
    CRICHLOW, TV
    BRITISH MEDICAL JOURNAL, 1951, 2 (4734): : 797 - 797
  • [37] ESTIMATION OF SKIN ENTRANCE DOSES (SEDs) FOR COMMON MEDICAL X-RAY DIAGNOSTIC EXAMINATIONS IN INDIA AND PROPOSED DIAGNOSTIC REFERENCE LEVELS (DRLs)
    Sonawane, A. U.
    Shirva, V. K.
    Pradhan, A. S.
    RADIATION PROTECTION DOSIMETRY, 2010, 138 (02) : 129 - 136
  • [38] Normalized organ doses and effective doses to a reference Indian adult male in conventional medical diagnostic x-ray examinations
    Biju, K
    Nagarajan, PS
    HEALTH PHYSICS, 2006, 90 (03): : 217 - 222
  • [39] Local diagnostic reference levels for routine chest X-ray examinations at a public sector hospital in central South Africa
    Junda, Maurice
    Muller, Henra
    Friedrich-Nel, Hesta
    HEALTH SA GESONDHEID, 2021, 26
  • [40] Diagnostic Reference Levels for Common X-ray Procedures in Peru
    Portocarrero Bonifaz, Andres
    Camarena Rodriguez, Caterina S.
    Palma Esparza, Ricardo
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (10)