EXPERIMENTAL-DETERMINATION OF FRICTION CHARACTERISTICS AT THE TRABECULAR BONE POROUS-COATED METAL INTERFACE IN CEMENTLESS IMPLANTS

被引:199
|
作者
SHIRAZIADL, A [1 ]
DAMMAK, M [1 ]
PAIEMENT, G [1 ]
机构
[1] SACRE COEUR HOSP,MONTREAL H3C 3A7,PQ,CANADA
来源
关键词
D O I
10.1002/jbm.820270205
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An apparatus was developed to measure load-displacement friction properties at the cancellous bone/porous-coated metal plate interface. Bone cubes were obtained from different proximal regions of four resurfaced cadaveric tibiae. Three different porous-surfaced metal plates (one fiber mesh and two bead) and a smooth-surfaced metal plate were used. In the presence of a constant normal contact pressure (0.10, 0.15, or 0.25 MPa), a variable tangential load up to the maximum resistance of the interface was applied and both relative normal and tangential displacements were recorded. Repetitive and fatigue loadings were also considered. Measured results show that the interface friction curve is highly nonlinear, exhibiting large relative tangential displacements in the range of 50-400 mum before the maximum load is reached. Relative displacements in the normal direction remain below 10 mum. The maximum resistance in friction is independent of the bone excision site, type of porous-surfaced metal plate, magnitude of normal load, placement of bone cubes on metal plates or vice versa, repetition of applied load, and conservation period of bone cubes in saline solution. The smooth-surfaced metal plate has significantly smaller friction resistance than porous-coated ones. The fatigue loading ot up to 4000 cycles at 1 Hz, in the presence of 0.25 MPa contact pressure, slightly decreases the interface friction coefficient. Finally, the initial secant stiffness of the interface at 75% of the maximum resistance load is found to be larger for the bone cubes from the lateral and medial regions and for the metal plate with smooth surface.
引用
收藏
页码:167 / 175
页数:9
相关论文
共 47 条
  • [21] FLUORAPATITE-COATED IMPLANTS IN EXPERIMENTAL ARTHRITIS - THE RESPONSE OF RABBIT TRABECULAR BONE
    DHERT, WJA
    THOMSEN, P
    KLEIN, CPAT
    DEGROOT, K
    ROZING, PM
    ERICSON, LE
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (02) : 59 - 66
  • [22] BONE INGROWTH AND WEAR DEBRIS IN WELL-FIXED CEMENTLESS POROUS-COATED TIBIAL COMPONENTS REMOVED FROM PATIENTS
    SUMNER, DR
    KIENAPFEL, H
    JACOBS, JJ
    URBAN, RM
    TURNER, TM
    GALANTE, JO
    JOURNAL OF ARTHROPLASTY, 1995, 10 (02): : 157 - 167
  • [23] Porous-coated glassy carbon bone substitute. Experimental study on rats and rabbits
    Tarvainen, T
    ANNALES CHIRURGIAE ET GYNAECOLOGIAE, 1998, 87 (04) : 332 - 333
  • [24] Cementless Total Hip Arthroplasty Using a Spongy Metal Surface Hip Prosthesis With a Collarless, Proximally Porous-Coated Stem
    Ogawa, Hiroyasu
    Ito, Yoshiki
    Itokazu, Mansho
    Mori, Nobuyuki
    Terabayashi, Nobuo
    Shimizu, Katsuji
    JOURNAL OF ARTHROPLASTY, 2010, 25 (03): : 375 - 380
  • [25] A comparison of radiographic bone height and probing attachment level measurements adjacent to porous-coated dental implants in humans
    Levy, D
    Deporter, DA
    Pharoah, M
    Tomlinson, G
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1997, 12 (04) : 541 - 546
  • [26] THE EFFECT OF BEAD DIAMETER ON THE ACCURACY OF 2 CURRENT TECHNIQUES USED TO QUANTIFY BONE INGROWTH IN POROUS-COATED IMPLANTS
    CAJA, VL
    MORONI, A
    EGGER, EL
    GOTTSAUNERWOLF, F
    CHAO, EYS
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1994, 5 (01) : 29 - 32
  • [27] DUAL-ENERGY X-RAY ABSORPTIOMETRY MEASUREMENT OF BONE-MINERAL DENSITY AROUND POROUS-COATED CEMENTLESS FEMORAL IMPLANTS - METHODS AND PRELIMINARY-RESULTS
    KILGUS, DJ
    SHIMAOKA, EE
    TIPTON, JS
    EBERLE, RW
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 279 - 287
  • [28] A MODEL FOR THE IMPLANT-BONE INTERFACE CHARACTERISTICS OF POROUS DENTAL IMPLANTS
    COOK, SD
    KLAWITTER, JJ
    WEINSTEIN, AM
    JOURNAL OF DENTAL RESEARCH, 1982, 61 (08) : 1006 - 1009
  • [29] Superior sealing effect of hydroxyapatite in porous-coated implants - Experimental studies on the migration of polyethylene particles around stable and unstable implants in dogs
    Rahbek, O
    Kold, S
    Bendix, K
    Overgaard, S
    Soballe, K
    ACTA ORTHOPAEDICA, 2005, 76 (03) : 375 - 385
  • [30] THE EXPERIMENTAL-DETERMINATION OF TRABECULAR BONE COMPRESSIVE MECHANICAL-PROPERTIES AND THEIR RELATIONSHIP TO MEASURED CT VALUES
    DICKIE, D
    GOLDSTEIN, SA
    FLYNN, MJ
    CIARELLI, MJ
    KU, JL
    KAPPER, M
    MEDICAL PHYSICS, 1986, 13 (04) : 577 - 577