CONE-BEAM TRANSMISSION COMPUTED-TOMOGRAPHY FOR NONUNIFORM ATTENUATION COMPENSATION OF SPECT IMAGES

被引:0
|
作者
MANGLOS, SH
BASSANO, DA
THOMAS, FD
机构
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper develops and tests cone-beam transmission computed tomography (CB-CT) for attenuation compensation of SPECT images. CB-CT was implemented on a rotating gamma camera with a point source (1-2 mCi) of Tc-99m, and a light-weight aluminium source holder. A cone-beam collimator may be used but is not required. Since the point source is either located at the collimator focal point, or the camera is uncollimated, CB-CT has excellent sensitivity (at least 150 times that of a parallel-hole, high-resolution collimator). The predicted resolution is equal to the intrinsic gamma camera resolution (3-4 mm), which is much higher than for a high-resolution, parallel-hole collimator (10-20 mm). In the present study, CB-CT provided low noise, high-resolution attenuation maps for use in a nonuniform attenuation-weighted backprojection algorithm. The attenuation compensation accuracy was tested using basic geometries of line sources and nonuniform density models. For the appropriate scaling of the attenuation map, the attenuation compensation was accurate and removed the SPECT image distortion associated with nonuniform attenuation. Attenuation maps acquired either with cone beam collimator or without any collimator were both successful. Using CB-CT, SPECT can thus be made much more accurate without adding unduly to the imaging time, complexity, or cost.
引用
收藏
页码:1813 / 1820
页数:8
相关论文
共 50 条
  • [1] ATTENUATION MAPS FOR SPECT DETERMINED USING CONE BEAM TRANSMISSION COMPUTED-TOMOGRAPHY
    MANGLOS, SH
    BASSANO, DA
    DUXBURY, CE
    CAPONE, RB
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) : 600 - 608
  • [2] Analysis on the use of megavoltage cone-beam computed-tomography images for dose calculation
    Johnson, M.
    Reitz, B.
    Gayou, O.
    Parda, D.
    Miften, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2551 - 2551
  • [3] Motion compensation in extremity cone-beam computed tomography
    Sisniega, Alejandro
    Thawait, Gaurav K.
    Shakoor, Delaram
    Siewerdsen, Jeffrey H.
    Demehri, Shadpour
    Zbijewski, Wojciech
    SKELETAL RADIOLOGY, 2019, 48 (12) : 1999 - 2007
  • [4] Motion compensation in extremity cone-beam computed tomography
    Alejandro Sisniega
    Gaurav K. Thawait
    Delaram Shakoor
    Jeffrey H. Siewerdsen
    Shadpour Demehri
    Wojciech Zbijewski
    Skeletal Radiology, 2019, 48 : 1999 - 2007
  • [5] IMPROVED CONE-BEAM SPECT VIA AN ACCURATE CORRECTION FOR NONUNIFORM PHOTON ATTENUATION
    WELCH, A
    WEBB, S
    FLOWER, M
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07): : 909 - 928
  • [6] Superimposition of Cone-Beam Computed Tomography Images by Joint Embedding
    Pei, Yuru
    Ma, Gengyu
    Chen, Gui
    Zhang, Xiaoyun
    Xu, Tianmin
    Zha, Hongbin
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (06) : 1218 - 1227
  • [7] Accuracy and reliability of stitched cone-beam computed tomography images
    Egbert, Nicholas
    Cagna, David R.
    Ahuja, Swati
    Wicks, Russell A.
    IMAGING SCIENCE IN DENTISTRY, 2015, 45 (01) : 41 - 47
  • [8] Three-dimensional reconstruction of individual cervical vertebrae from cone-beam computed-tomography images
    Shi, Hongjian
    Scarfe, William C.
    Farman, Allan G.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 131 (03) : 426 - 432
  • [9] Cone-beam computed tomography for trauma
    Gupta, Saurabh
    Martinson, James R.
    Ricaurte, Daniel
    Scalea, Thomas M.
    Morrison, Jonathan J.
    JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2020, 89 (03): : E34 - E40
  • [10] Cone-Beam Computed Tomography in Orthodontics
    Abdelkarim, Ahmad
    DENTISTRY JOURNAL, 2019, 7 (03)