Dosimetric verification of prowess panter treatment planning system and the evaluation of the clinical acceptance

被引:0
|
作者
Unal Karabey, Aysegul [1 ]
Onay, Ozgehan [1 ]
Karacam, Songul [2 ]
Koca, Sedat [2 ]
Atkovar, Gulyuz [2 ]
机构
[1] Aktif Care Tibbi Cihazlar Ltd Sti, Istanbul, Turkey
[2] Istanbul Univ, Tip Fak, Radyasyon Onkol Anabilim Dali, Istanbul, Turkey
来源
关键词
Leaf transmission; profile; prowess panther treatment planning system; output; percent depth dose;
D O I
10.5505/tjoncol.2012.553
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVES In this study, the aim was to perform quality control of the Prowess Panther treatment planning system by comparing with the measurement data. METHODS For dosimetric measurements, Iba brand cc 13 and FC-65P 0.6cc model (Scanditronix, Wellhofer, Germany), the cylindrical ion chambers, and Iba brand dosel electrometer model, RW3 solid water phantom and RFA-300 water phantom, were used. Measurements were performed with Elekta Synergy linear accelerator. RFA-300 Wellhofer profile measurements were taken using a water phantom in the wedge and open field depth dose, off-axis ratio, and different x, y and diagonal directions. Output measurements were performed with RW3 solid water. RESULTS Open and wedge field percentage depth dose (PDD) and profiles of the treatment planning system were compared with measurements at Dmax, 5 cm, 10 cm, and 15 cm depths. The results were within the boundaries of acceptable BJR and found less than 3%. CONCLUSION Based on the quality control results, the Prowess Panther v5.01 Planning System was within the boundaries of acceptable accuracy in dosimetry.
引用
收藏
页码:76 / 82
页数:7
相关论文
共 50 条
  • [1] Dosimetric verification of clinical radiotherapy treatment planning system
    Kolarevic, Goran
    Jaros, Drazan
    Marosevic, Goran
    Ignjatic, Dejan
    Mirjanic, Dragoljub
    [J]. VOJNOSANITETSKI PREGLED, 2022, 79 (02) : 115 - 124
  • [2] Dosimetric verification of a clinical inverse treatment planning system
    Xing, L
    Findley, D
    Ma, L
    Ma, CM
    Forster, K
    Boyer, AL
    [J]. RADIOLOGY, 1998, 209P : 397 - 397
  • [3] CLINICAL COMMISSIONING AND DOSIMETRIC VERIFICATION OF THE RAYSTATION TREATMENT PLANNING SYSTEM
    Tugrul, Taylan
    [J]. EAST EUROPEAN JOURNAL OF PHYSICS, 2021, (04): : 114 - 119
  • [4] Dosimetric verification of a commercial inverse treatment planning system
    Lei, X
    Curran, B
    Hill, R
    Holmes, T
    Ma, LJ
    Forster, KM
    Boyer, AL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02): : 463 - 478
  • [5] Clinical evaluation and verification of the hyperthermia treatment planning system HyperPlan
    Gellermann, J
    Wust, P
    Stalling, D
    Seebass, M
    Nadobny, J
    Beck, R
    Hege, HC
    Deuflhard, P
    Felix, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04): : 1145 - 1156
  • [6] THE DOSIMETRIC VERIFICATION OF A PENCIL BEAM BASED TREATMENT PLANNING SYSTEM
    KNOOS, T
    CEBERG, C
    WEBER, L
    NILSSON, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (10): : 1609 - 1628
  • [7] An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning
    Yamamoto, T.
    Mizowaki, T.
    Miyabe, Y.
    Takegawa, H.
    Narita, Y.
    Yano, S.
    Nagata, Y.
    Teshima, T.
    Hiraoka, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (07): : 1991 - 2008
  • [8] Dosimetric Verification of MammoSite Irradiation Treatment Planning
    Budde, M.
    Schmolinga, L.
    Ahl-Alhosseini, S.
    Ahl-Alhosseini, A.
    Demirel, C.
    Spaan, B.
    Adamietz, I. A.
    Hermani, H.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 : S98 - S99
  • [9] Dosimetric Verification of stereotactic Radiation Treatment Planning
    Senn, A. M.
    Garbe, S.
    Muedder, T.
    Schoroth, F.
    Roehner, F.
    Koch, D.
    Schmeel, C.
    Wilhelm-Buchstab, T.
    Leitzen, C.
    Heimann, M.
    Tschirner, S.
    Buecheler, B.
    Baumert, B. G.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 : S70 - S70
  • [10] Dosimetric evaluation of lung treatment plans produced by the Prowess Panther system using Monte Carlo simulation
    Luong Thi Oanh
    Duong Thanh Tai
    Truong Thi Hong Loan
    Trinh Hong Minh
    Truong Van Minh
    Chow, James C. L.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (05):