Dosimetric evaluation of lung treatment plans produced by the Prowess Panther system using Monte Carlo simulation

被引:5
|
作者
Luong Thi Oanh [1 ,2 ]
Duong Thanh Tai [2 ,3 ]
Truong Thi Hong Loan [2 ]
Trinh Hong Minh [4 ]
Truong Van Minh [5 ]
Chow, James C. L. [6 ]
机构
[1] Nguyen Tat Thanh Univ, Fac Med, Ho Chi Minh, Vietnam
[2] VNUHCM Univ Sci, Fac Phys & Engn Phys, Ho Chi Minh, Vietnam
[3] Dong Nai Gen Hosp, Dept Radiat Oncol, Bien Hoa, Vietnam
[4] Appl Res Inst Psychol & Educ Sci, Bien Hoa, Vietnam
[5] Dong Nai Univ, Dept Sci, Bien Hoa, Vietnam
[6] Univ Hlth Network, Princess Margaret Canc Ctr, Toronto, ON, Canada
来源
关键词
algorithm dose calculation; Monte Carlo simulation; treatment planning system; DOSE CALCULATION ACCURACY; CALCULATION ALGORITHMS; RADIATION-THERAPY; COLLAPSED-CONE; PENCIL-BEAM; WATER; HEAD;
D O I
10.1088/2057-1976/ab367d
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study evaluated the accuracy of lung dose calculation done by the fast photon effective (FPE) and the collapsed cone convolution (CCC) algorithms of the Prowess Panther treatment planning system (TPS) using Monte Carlo (MC) simulation. Materials and methods: A set of treatment plans of test cases including an acrylic phantom, the QUASAR multi-purpose body phantom, and one lung cancer patient, were created the system to assess the accuracy of the FPE and CCC algorithms. The DICOM-RT files of the plans were imported to the EGSnrc-based Monte Carlo simulation for dose calculations. The plans generated by the TPS and Monte Carlo simulation were compared using relative dose error comparison and 3D gamma index. The gamma index, using global methods, was implemented in PTW-VeriSoft with 3%/3 mm, 2%/2 mm criteria. Results: There was a good agreement between Monte Carlo-simulated and TPS-calculated doses for both the QUASAR multipurpose body phantom and one lung cancer patient. However, discrepancies for the FPE algorithm were found to be 10% in the inhomogeneous medium such as the lung. Conclusions: The FPE algorithm may not accurately predict the dose distributions in and near the inhomogeneous structures. Monte Carlo simulation and CCC algorithm are more accurate than the FPE algorithm in calculating the dose in an inhomogeneous medium. The FPE and CCC algorithms must be validated before clinical implementation of the system.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Dosimetric and Radiobiological Comparison of Conventional and Monte Carlo Treatment Plans of Lung Tumors Using SBRT
    Stathakis, S.
    Mavroidis, P.
    Tuazon, B.
    Defoor, D.
    Kirby, N.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3607 - 3607
  • [2] Dosimetric Verification of Jaw-Only IMRT On Prowess Panther System Based On AAPM TG 119: A Monte Carlo Study
    Tai, D.
    Oanh, L.
    Loan, T.
    Chow, J.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E618 - E619
  • [3] Robustness evaluation of production plans using Monte Carlo simulation
    Franke, Susanne
    Franke, Felix
    Riedel, Ralph
    [J]. 10TH CIRP SPONSORED CONFERENCE ON DIGITAL ENTERPRISE TECHNOLOGIES (DET 2020) - DIGITAL TECHNOLOGIES AS ENABLERS OF INDUSTRIAL COMPETITIVENESS AND SUSTAINABILITY, 2021, 54 : 130 - 135
  • [4] Dosimetric evaluation of a monte carlo IMRT treatment planning system with the MIMiC
    Rassiah-Szegedi, P.
    Fuss, M.
    Sheikh-Bagheri, D.
    Szegedi, M.
    Papanikolaoul, N.
    Stathakis, S.
    Salter, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2441 - 2442
  • [5] Dosimetric Evaluation of Radiation Treatment Planning for Simultaneous Integrated Boost Technique Using Monte Carlo Simulation
    Shende, Ravindra
    Dhoble, S. J.
    Gupta, Gourav
    [J]. JOURNAL OF MEDICAL PHYSICS, 2023, 48 (03) : 298 - 306
  • [6] Dosimetric evaluation of a Monte Carlo IMRT treatment planning system incorporating the MIMiC
    Rassiah-Szegedi, P.
    Fuss, M.
    Sheikh-Bagheri, D.
    Szegedi, M.
    Stathakis, S.
    Lancaster, J.
    Papanikolaou, N.
    Salter, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23): : 6931 - 6941
  • [7] Dosimetric evaluation of a dedicated stereotactic linear accelerator using measurement and Monte Carlo simulation
    Heydarian, M.
    Asnaashari, K.
    Allahverdi, M.
    Jaffray, D. A.
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 3943 - 3954
  • [8] Evaluation of Dosimetric Parameters of Two Brachytherapy Ophthalmic Sources Using Monte Carlo Simulation
    Ramon, P.
    Quevedo, A.
    Pianoschi, T.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [9] Evaluation of the Influence of the Dosimetric Leaf Gap on RapidArc Plans Using Gafchromic Film and Monte Carlo Simulations
    Szpala, S.
    Atwal, P.
    Cao, F.
    Kohl, K.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Monte Carlo fluence simulation for prospective evaluation of interstitial photodynamic therapy treatment plans
    Cassidy, Jeffrey
    Betz, Vaughn
    Lilge, Lothar
    [J]. OPTICAL METHODS FOR TUMOR TREATMENT AND DETECTION: MECHANISMS AND TECHNIQUES IN PHOTODYNAMIC THERAPY XXIV, 2015, 9308