Dosimetric evaluation of a Monte Carlo IMRT treatment planning system incorporating the MIMiC

被引:8
|
作者
Rassiah-Szegedi, P. [1 ,2 ]
Fuss, M. [3 ]
Sheikh-Bagheri, D. [4 ]
Szegedi, M. [2 ]
Stathakis, S. [1 ,2 ]
Lancaster, J. [2 ]
Papanikolaou, N. [1 ,2 ]
Salter, B. [5 ]
机构
[1] Canc Therapy & Res Ctr S Texas, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, San Antonio, TX 78229 USA
[3] Univ Oregon, Hlth Sci Ctr, Portland, OR USA
[4] Allegheny Hosp, Pittsburgh, PA USA
[5] Univ Utah, Huntsmen Canc Ctr, Salt Lake City, UT 84112 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 23期
关键词
D O I
10.1088/0031-9155/52/23/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The high dose per fraction delivered to lung lesions in stereotactic body radiation therapy (SBRT) demands high dose calculation and delivery accuracy. The inhomogeneous density in the thoracic region along with the small fields used typically in intensity-modulated radiation therapy (IMRT) treatments poses a challenge in the accuracy of dose calculation. In this study we dosimetrically evaluated a pre-release version of a Monte Carlo planning system (PEREGRINE 1.6b, NOMOS Corp., Cranberry Township, PA), which incorporates the modeling of serial tomotherapy IMRT treatments with the binary multileaf intensity modulating collimator (MIMiC). The aim of this study is to show the validation process of PEREGRINE 1.6b since it was used as a benchmark to investigate the accuracy of doses calculated by a finite size pencil beam (FSPB) algorithm for lung lesions treated on the SBRT dose regime via serial tomotherapy in our previous study. Doses calculated by PEREGRINE were compared against measurements in homogeneous and inhomogeneous materials carried out on a Varian 600C with a 6 MV photon beam. Phantom studies simulating various sized lesions were also carried out to explain some of the large dose discrepancies seen in the dose calculations with small lesions. Doses calculated by PEREGRINE agreed to within 2% in water and up to 3% for measurements in an inhomogeneous phantom containing lung, bone and unit density tissue.
引用
收藏
页码:6931 / 6941
页数:11
相关论文
共 50 条
  • [1] Dosimetric evaluation of a monte carlo IMRT treatment planning system with the MIMiC
    Rassiah-Szegedi, P.
    Fuss, M.
    Sheikh-Bagheri, D.
    Szegedi, M.
    Papanikolaoul, N.
    Stathakis, S.
    Salter, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2441 - 2442
  • [2] Dosimetric evaluation of the clinical implementation of the first commercial IMRT Monte Carlo treatment planning system at 6 MV
    Heath, E
    Seuntjens, J
    Sheikh-Bagheri, D
    [J]. MEDICAL PHYSICS, 2004, 31 (10) : 2771 - 2779
  • [3] Dosimetric validation of a commercial Monte Carlo based IMRT planning system
    Grofsmid, Dennis
    Dirkx, Maarten
    Marijnissen, Hans
    Woudstra, Evert
    Heijmen, Ben
    [J]. MEDICAL PHYSICS, 2010, 37 (02) : 540 - 549
  • [4] Dosimetric verification of breast IMRT treatment planning using Monte Carlo method
    Gagne, I.
    Zavgorodni, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S224 - S225
  • [5] Commissioning of Monaco Monte Carlo IMRT Treatment Planning System
    Li, J.
    Doemer, A.
    Cao, J.
    Podder, T.
    Harrison, A.
    Yu, Y.
    Galvin, J.
    Xiao, Y.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [6] Dosimetric verification of IMRT treatment planning using Monte Carlo simulations for prostate cancer
    Yang, J
    Li, J
    Chen, L
    Price, R
    McNeeley, S
    Qin, L
    Wang, L
    Xiong, W
    Ma, CM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05): : 869 - 878
  • [7] Monte Carlo flattening filter free MIMiC based IMRT planning
    Rassiah-Szegedi, P.
    Stathakis, S.
    Szegedi, M.
    Esquivel, C.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2575 - 2575
  • [8] Dosimetric Validation of a Commercial Monte-Carlo Based IMRT Planning
    Grofsmid, D.
    Dirkx, M.
    Woudstra, E.
    Marijnissen, H.
    Heijmen, B.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [9] An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning
    Yamamoto, T.
    Mizowaki, T.
    Miyabe, Y.
    Takegawa, H.
    Narita, Y.
    Yano, S.
    Nagata, Y.
    Teshima, T.
    Hiraoka, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (07): : 1991 - 2008