Dosimetric evaluation of a Monte Carlo IMRT treatment planning system incorporating the MIMiC

被引:8
|
作者
Rassiah-Szegedi, P. [1 ,2 ]
Fuss, M. [3 ]
Sheikh-Bagheri, D. [4 ]
Szegedi, M. [2 ]
Stathakis, S. [1 ,2 ]
Lancaster, J. [2 ]
Papanikolaou, N. [1 ,2 ]
Salter, B. [5 ]
机构
[1] Canc Therapy & Res Ctr S Texas, San Antonio, TX 78229 USA
[2] Univ Texas Hlth Sci Ctr San Antonio, San Antonio, TX 78229 USA
[3] Univ Oregon, Hlth Sci Ctr, Portland, OR USA
[4] Allegheny Hosp, Pittsburgh, PA USA
[5] Univ Utah, Huntsmen Canc Ctr, Salt Lake City, UT 84112 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 23期
关键词
D O I
10.1088/0031-9155/52/23/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The high dose per fraction delivered to lung lesions in stereotactic body radiation therapy (SBRT) demands high dose calculation and delivery accuracy. The inhomogeneous density in the thoracic region along with the small fields used typically in intensity-modulated radiation therapy (IMRT) treatments poses a challenge in the accuracy of dose calculation. In this study we dosimetrically evaluated a pre-release version of a Monte Carlo planning system (PEREGRINE 1.6b, NOMOS Corp., Cranberry Township, PA), which incorporates the modeling of serial tomotherapy IMRT treatments with the binary multileaf intensity modulating collimator (MIMiC). The aim of this study is to show the validation process of PEREGRINE 1.6b since it was used as a benchmark to investigate the accuracy of doses calculated by a finite size pencil beam (FSPB) algorithm for lung lesions treated on the SBRT dose regime via serial tomotherapy in our previous study. Doses calculated by PEREGRINE were compared against measurements in homogeneous and inhomogeneous materials carried out on a Varian 600C with a 6 MV photon beam. Phantom studies simulating various sized lesions were also carried out to explain some of the large dose discrepancies seen in the dose calculations with small lesions. Doses calculated by PEREGRINE agreed to within 2% in water and up to 3% for measurements in an inhomogeneous phantom containing lung, bone and unit density tissue.
引用
收藏
页码:6931 / 6941
页数:11
相关论文
共 50 条
  • [41] Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system
    Cadman, P
    Bassalow, R
    Sidhu, NPS
    Ibbott, G
    Nelson, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16): : 3001 - 3010
  • [42] Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the electron Monte Carlo dose algorithm in the Eclipse treatment planning system
    Ali, Imad
    Alsbou, Nesreen
    Ahmad, Salahuddin
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (01):
  • [43] Experimental and Monte Carlo evaluation of Eclipse treatment planning system for lung dose calculations
    Mesbahi, Asghar
    Thwaites, David I.
    Reilly, Andrew J.
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2006, 11 (03) : 123 - 133
  • [44] Evaluation of a commercial electron treatment planning system based on Monte Carlo techniques (eMC)
    Pemler, Peter
    Besserer, Juergen
    Schneider, Uwe
    Neuenschwander, Hans
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2006, 16 (04): : 313 - 329
  • [45] Evaluation of clinical IMRT treatment planning using the GATE Monte Carlo simulation platform for absolute and relative dose calculations
    Benhalouche, S.
    Visvikis, D.
    Le Maitre, A.
    Pradier, O.
    Boussion, N.
    [J]. MEDICAL PHYSICS, 2013, 40 (02)
  • [46] Dosimetric evaluation of the PEREGRINE IMRT MC treatment planning system at 6 MV for small fields in heterogeneous media
    Sheikh-Bagheri, D
    Cocheci, C
    Heath, E
    Seuntjens, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2016 - 2016
  • [47] Quantitative dosimetric verification of an IMRT planning and delivery system
    Low, DA
    Mutic, S
    Dempsey, JF
    Gerber, RL
    Bosch, WR
    Perez, CA
    Purdy, JA
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 49 (03) : 305 - 316
  • [48] Clinical Implementation and Evaluation of Monte Carlo Treatment Planning for CyberKnife
    Sharma, S.
    Ott, J.
    Williams, J.
    Dickow, D.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [49] Prone accelerated partial breast irradiation with MERT plus IMRT using a Monte Carlo treatment planning system.
    Jimenez-Ortega, E.
    Brualla, L.
    Ureba, A.
    Baeza, J. A.
    Gonzalez, A.
    Barbeiro, A. R.
    Rosello, J.
    Lopez-Torrecilla, J.
    Ferrer, J.
    Leal, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S199 - S200
  • [50] Comparison of PRIMO Monte Carlo code and Eclipse treatment planning system in calculation of dosimetric parameters in brain cancer radiotherapy
    Altuwayrish, Ali
    Ghorbani, Mahdi
    Bakhshandeh, Mohsen
    Roozmand, Zahra
    Hoseini-Ghahfarokhi, Mojtaba