SATURATION BY PULSE OXIMETRY - COMPARISON OF THE RESULTS OBTAINED BY INSTRUMENTS OF DIFFERENT BRANDS

被引:46
|
作者
THILO, EH
ANDERSEN, D
WASSERSTEIN, ML
SCHMIDT, J
LUCKEY, D
机构
[1] CHILDRENS HOSP, DEPT RESP THERAPY, DENVER, CO 80218 USA
[2] CHILDRENS HOSP, DEPT PATHOL, DENVER, CO 80218 USA
[3] UNIV COLORADO, SCH MED, DEPT PEDIAT, DENVER, CO 80202 USA
来源
JOURNAL OF PEDIATRICS | 1993年 / 122卷 / 04期
关键词
D O I
10.1016/S0022-3476(05)83549-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
We noticed that arterial oxygen saturation by pulse oximetry (Spo2) was generally lower when determined by the Ohmeda Biox 3700 pulse oximeter than when determined by the Nellcor N-100 pulse oximeter, and we investigated whether this finding was consistent and the reason for the discrepancy. We placed both oximeters simultaneously on 30 infants with indwelling arterial catheters and measured arterial partial pressure of oxygen (Pao2), percentage of fetal hemoglobin, and complete cooximetry, including arterial oxygen saturation (Sao2) with a Radiometer OSM-3 cooximeter, with and without correction for fetal hemoglobin levels, in four samples of blood from each infant during a 12-hour period for a total of 120 samples. The Nellcor SPO2 was consistently higher than the Ohmeda SPO2 by a mean (+/- SD) of 1.61% +/- 2.69% (p <0.001). The Nellcor SPO2 correlated best with functional SaO2 ((oxyhemoglobin(HbO2)/(HbO2 + reduced hemoglobin)) x 100); Ohmedo SPO2 correlated best with fractional SaO2 ((HbO2/(HbO2 + reduced hemoglobin + carboxyhemoglobin + methemoglobin)) X 100), reflecting a fundamental difference in the calibration algorithms used in the two instruments. A desired Pao2 of 50 to 100 mm Hg, is maintained when the range of SPO2 is 90% to 96% for Ohmeda Spo2 and 92% to 98% for Nellcor Spo2 in the neonate, giving a positive predictive value in this study of 94% to 95%. We conclude that Spo2 determined by pulse oximeters of different brands is not interchangeable, and this may be of clinical importance in predicting Pao2 on the basis of Spo2.
引用
收藏
页码:620 / 626
页数:7
相关论文
共 50 条
  • [21] Pulse Oximetry in Bronchiolitis: Have We Reached Saturation?
    Lee, Vivian
    Ruppel, Halley
    Schondelmeyer, Amanda C.
    HOSPITAL PEDIATRICS, 2024, 14 (02) : e107 - e109
  • [22] Racial Disparity in Oxygen Saturation Measurements by Pulse Oximetry
    Jamali, Haya
    Castillo, Lauren T.
    Morgan, Chelsea Cosby
    Coult, Jason
    Muhammad, Janice L.
    Osobamiro, Oyinkansola O.
    Parsons, Elizabeth C.
    Adamson, Rosemary
    ANNALS OF THE AMERICAN THORACIC SOCIETY, 2022, 19 (12) : 1951 - 1964
  • [23] CLINICAL-RESULTS OF FETAL PULSE OXIMETRY MEASURING THE FETAL ARTERIAL OXYGEN-SATURATION
    BUTTERWEGGE, M
    LAMPE, S
    GEBURTSHILFE UND FRAUENHEILKUNDE, 1995, 55 (11) : 634 - 641
  • [24] MONITORING OF ARTERIAL OXYGEN-SATURATION BY PULSE OXIMETRY
    STRIEBEL, HW
    STEINHOFF, U
    KRETZ, FJ
    ANASTHESIOLOGIE & INTENSIVMEDIZIN, 1988, 29 (01): : 8 - 16
  • [25] Baseline oxigen saturation obtained by pulse-oximetry and diagnosis of heart failure in the acute phase of myocardial infarction
    Masip, J
    Vecilla, F
    Gayà, M
    Páez, J
    Betbesé, A
    Manresa, R
    Padró, J
    Cañizares, R
    EUROPEAN HEART JOURNAL, 2002, 23 : 226 - 226
  • [26] Comparison of human smartwatch and transmittance pulse oximetry for evaluating peripheral oxygen saturation in anesthetized dogs
    Senocak, M. G.
    Okur, S.
    Ersoz, U.
    Yanmaz, L. E.
    Kocaman, Y.
    Turgut, F.
    Bedir, A. G.
    Orhun, O. T.
    IRANIAN JOURNAL OF VETERINARY RESEARCH, 2024, 25 (02)
  • [27] Validity Of Heart Rate Obtained Via Pulse Oximetry
    Webb, Kyle M.
    Hanson, Karla J.
    Sheely, Jennifer M.
    Berry, Michael J.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2005, 37 : S232 - S233
  • [28] A Comparison of Soundness Results Obtained by Different Approaches
    Gruhn, Volker
    Lane, Ralf
    BUSINESS PROCESS MANAGEMENT WORKSHOPS, 2009, 2010, 43 : 501 - 512
  • [29] A pilot study of pharyngeal pulse oximetry with the laryngeal mask airway:: A comparison with finger oximetry and arterial saturation measurements in healthy anesthetized patients
    Keller, C
    Brimacombe, J
    Agrò, F
    Margreiter, J
    ANESTHESIA AND ANALGESIA, 2000, 90 (02): : 440 - 444
  • [30] Comparison of transesophageal pulse oximetry (TEPO) with peripheral surface pulse oximetry in CABG patients
    Prielipp, RC
    Scuderi, PE
    Butterworth, JF
    Royster, RL
    Atlee, JL
    ANESTHESIOLOGY, 1996, 85 (3A) : A485 - A485