A pilot study of pharyngeal pulse oximetry with the laryngeal mask airway:: A comparison with finger oximetry and arterial saturation measurements in healthy anesthetized patients

被引:11
|
作者
Keller, C
Brimacombe, J
Agrò, F
Margreiter, J
机构
[1] Leopold Franzens Univ Innsbruck, Dept Anaesthesia & Intens Care Med, Innsbruck, Austria
[2] Univ Queensland, Cairns Base Hosp, Dept Anaesthesia & Intens Care, Cairns, Australia
[3] Univ Sch Med LIU, Dept Anaesthesia, Rome, Italy
来源
ANESTHESIA AND ANALGESIA | 2000年 / 90卷 / 02期
关键词
D O I
10.1097/00000539-200002000-00037
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We compared pharyngeal Spo(2) by using the laryngeal mask airway (LMA) to finger Spo(2) and oxygen saturation from arterial blood samples (Sao(2)). We studied 20 hemodynamically stable, well oxygenated, anesthetized patients (ASA physical status I-III, aged 18-80 yr). A single-use pediatric pulse oximeter was attached to the back plate of a size 5 LMA. Pharyngeal and finger Spo(2) (dominant index finger) and Sao(2) (nondominant radial artery) were measured with the cuff volume at 0-40 mL in the neutral position. The intracuff pressure was then set at 60 cm H2O in the neutral position, and readings were taken with the head-neck flexed, extended, and rotated. Sao(2) was the same as pharyngeal Spo(2) at 20 and 30 mL cuff volume, but higher than pharyngeal Spo(2) at all other cuff volumes and head-neck positions (P < 0.04). Sao(2) was always higher than finger Spo(2) (P < 0.01). Pharyngeal Spo(2) was higher than finger Spo(2) at cuff volumes 10-40 mL and in the flexed and rotated head-neck positions (all: P < 0.007), but was lower at 0 cuff volume (P < 0.0001) and similar in the extended head-neck position. There was an increase in pharyngeal Spo(2) between 0 and 10 mL cuff volume (P < 0.0001), but no changes thereafter. Pharyngeal Spo(2) was similar in the flexed, rotated and extended head-neck positions. Pharyngeal Spo(2) agrees more closely with Sao(2) (mean difference < 0.7%) than finger Spo(2) (mean difference > 1.1%) at 10-40 mL cuff volume and in head-neck flexion. The standard error of limits was identical (0.09) for both finger Spo(2) and pharyngeal Spo(2) if data at 0 cuff volume are excluded. We conclude that pharyngeal Spo(2) with the LMA is feasible and generally provides more accurate readings than finger Spo(2), in hemodynamically stable, well oxygenated, anesthetized patients. Implications: Pharyngeal oximetry with the laryngeal mask airway is feasible and generally provides more accurate readings than finger oximetry in hemodynamically stable, well oxygenated, anesthetized patients.
引用
收藏
页码:440 / 444
页数:5
相关论文
共 50 条
  • [1] Pulse oximetry with the laryngeal mask airway
    Agrò, F
    Carassiti, M
    Cataldo, R
    Gherardi, S
    Barzoi, G
    RESUSCITATION, 1999, 43 (01) : 65 - 67
  • [3] Cephalic and digital pulse oximetry in cardiac surgery A comparative pilot study with arterial oximetry
    Fischer, Marc-Olivier
    Fornier, William
    Hanouz, Jean-Luc
    Fellahi, Jean-Luc
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2015, 32 (01) : 60 - 61
  • [4] A comparison of pharyngeal mucosal pressure and airway sealing pressure with the laryngeal mask airway in anesthetized adult patients
    Brimacombe, J
    Keller, C
    ANESTHESIA AND ANALGESIA, 1998, 87 (06): : 1379 - 1382
  • [5] Reflectance pulse oximetry from core body in neonates and infants: Comparison to arterial blood oxygen saturation and to transmission pulse oximetry
    Kugelman A.
    Wasserman Y.
    Mor F.
    Goldinov L.
    Geller Y.
    Bader D.
    Journal of Perinatology, 2004, 24 (6) : 366 - 371
  • [6] Successful use of pharyngeal pulse oximetry with the oropharyngeal airway in severely shocked patients
    Yu, H.
    Liu, B.
    ANAESTHESIA, 2007, 62 (07) : 734 - 736
  • [7] VALIDATION OF ARTERIAL OXYGEN-SATURATION MEASUREMENTS INUTERO USING PULSE OXIMETRY
    HARRIS, AP
    SENDAK, MJ
    CHUNG, DC
    RICHARDSON, CA
    AMERICAN JOURNAL OF PERINATOLOGY, 1993, 10 (03) : 250 - 254
  • [8] Comparison of oxygen saturation by pulse oximetry and co-oximetry during exercise testing in patients with COPD
    McGovern, JP
    Sasse, SA
    Stansbury, DW
    Causing, LA
    Light, RW
    CHEST, 1996, 109 (05) : 1151 - 1155
  • [9] A comparison of the Supreme™ laryngeal mask airway with the Proseal™ laryngeal mask airway in anesthetized paralyzed adult patients: a randomized crossover study
    Tham, Huae Min
    Tan, Su Meng
    Woon, Kwee Lian
    Zhao, Yu Dong
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2010, 57 (07): : 672 - 678
  • [10] Accuracy of a New Pulse Oximetry in Detection of Arterial Oxygen Saturation and Heart Rate Measurements: The SOMBRERO Study
    Marinari, Stefano
    Volpe, Pasqualina
    Simoni, Marzia
    Aventaggiato, Matteo
    De Benedetto, Fernando
    Nardini, Stefano
    Sanguinetti, Claudio M.
    Palange, Paolo
    SENSORS, 2022, 22 (13)