Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Momennezhad, Mehdi [2 ]
Hashemi, Seyed Mohammad [1 ]
机构
[1] Mashhad Univ Med Sci, Med Phys Res Ctr, Mashhad, Iran
[2] Mashhad Univ Med Sci, Imam Reza Hosp, Nucl Med Res Ctr, Mashhad, Iran
关键词
Simulation; Monte Carlo; Linear accelerator; Dosimetry;
D O I
10.1016/j.rpor.2012.01.011
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: Exact knowledge of dosimetric parameters is an essential pre-requisite of an effective treatment in radiotherapy. In order to fulfill this consideration, different techniques have been used, one of which is Monte Carlo simulation. Materials and methods: This study used the MCNP-4C(c) to simulate electron beams from Neptun 10 PC medical linear accelerator. Output factors for 6, 8 and 10 MeV electrons applied to eleven different conventional fields were both measured and calculated. Results: The measurements were carried out by a Wellhofler-Scanditronix dose scanning system. Our findings revealed that output factors acquired by MCNP-4C simulation and the corresponding values obtained by direct measurements are in a very good agreement. Conclusion: In general, very good consistency of simulated and measured results is a good proof that the goal of this work has been accomplished. (c) 2012 Greater Poland Cancer Centre, Poland. Published by Elsevier Urban & Partner Sp. z.o.o. All rights reserved.
引用
收藏
页码:115 / 118
页数:4
相关论文
共 50 条
  • [1] Monte Carlo calculations of electron beam output factors for a medical linear accelerator
    Kapur, A
    Ma, CM
    Mok, EC
    Findley, DO
    Boyer, AL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12): : 3479 - 3494
  • [2] Monte Carlo modeling of electron beams from a NEPTUN 10PC medical linear accelerator
    Jabbari, Nasrollah
    Hashemi-Malayeri, Bijan
    [J]. NUKLEONIKA, 2009, 54 (04) : 233 - 238
  • [3] Modeling 6 MeV Electron Beam From Medical Linear Accelerator Using Monte Carlo Simulation
    Dechsupa, P.
    Tannanonta, C.
    Phaisangittisakul, N.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [4] Benchmarking of electron beam parameters based on Monte Carlo linear accelerator simulation
    Zhang, Fan
    Zhou, Mi
    Liu, Jing
    Yue, Lu
    Deng, Lihua
    Xu, Zhijian
    Wang, Gang
    [J]. TRANSLATIONAL CANCER RESEARCH, 2020, 9 (02) : 577 - 584
  • [5] Monte Carlo Simulation of an Electron Irradiation Device for Medical Application of an Electron Linear Accelerator
    Kyoung Won Jang
    Manwoo Lee
    Heuijin Lim
    Sang Koo Kang
    Sang Jin Lee
    Seung Heon Kim
    Dong Eun Lee
    Hee Chang Kim
    Dong Hyeok Jeong
    [J]. Journal of the Korean Physical Society, 2020, 76 : 588 - 591
  • [6] Monte Carlo Simulation of an Electron Irradiation Device for Medical Application of an Electron Linear Accelerator
    Jang, Kyoung Won
    Lee, Manwoo
    Lim, Heuijin
    Kang, Sang Koo
    Lee, Sang Jin
    Kim, Seung Heon
    Lee, Dong Eun
    Kim, Hee Chang
    Jeong, Dong Hyeok
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2020, 76 (07) : 588 - 591
  • [7] Monte Carlo simulation of a medical linear accelerator for radiotherapy use
    Serrano, B.
    Hachem, A.
    Franchisseur, E.
    Herault, J.
    Marcie, S.
    Costa, A.
    Bensadoun, R. J.
    Barthe, J.
    Gerard, J. P.
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 506 - 509
  • [8] A MULTIPLE SOURCE MONTE CARLO MODEL DEVELOPED FOR SIMULATING THE CLINICAL ELECTRON BEAMS OF NEPTUN 10PC MEDICAL LINAC
    Hashemi, B.
    Jabbari, N.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S401 - S402
  • [9] An intraoral cone system for a Neptun 10PC linear accelerator
    Shokrani, Parvaneh
    Soltani, Minoo
    [J]. NUKLEONIKA, 2011, 56 (04) : 381 - 384
  • [10] Monte Carlo investigation of the effects of an IMRT compensator on linear accelerator beam output
    Niven, D.
    Lewis, J.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2559 - 2559