Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Momennezhad, Mehdi [2 ]
Hashemi, Seyed Mohammad [1 ]
机构
[1] Mashhad Univ Med Sci, Med Phys Res Ctr, Mashhad, Iran
[2] Mashhad Univ Med Sci, Imam Reza Hosp, Nucl Med Res Ctr, Mashhad, Iran
关键词
Simulation; Monte Carlo; Linear accelerator; Dosimetry;
D O I
10.1016/j.rpor.2012.01.011
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: Exact knowledge of dosimetric parameters is an essential pre-requisite of an effective treatment in radiotherapy. In order to fulfill this consideration, different techniques have been used, one of which is Monte Carlo simulation. Materials and methods: This study used the MCNP-4C(c) to simulate electron beams from Neptun 10 PC medical linear accelerator. Output factors for 6, 8 and 10 MeV electrons applied to eleven different conventional fields were both measured and calculated. Results: The measurements were carried out by a Wellhofler-Scanditronix dose scanning system. Our findings revealed that output factors acquired by MCNP-4C simulation and the corresponding values obtained by direct measurements are in a very good agreement. Conclusion: In general, very good consistency of simulated and measured results is a good proof that the goal of this work has been accomplished. (c) 2012 Greater Poland Cancer Centre, Poland. Published by Elsevier Urban & Partner Sp. z.o.o. All rights reserved.
引用
收藏
页码:115 / 118
页数:4
相关论文
共 50 条
  • [21] Monte Carlo simulation of the photon beam characteristics from medical linear accelerators
    Kim, H. K.
    Han, S. J.
    Kim, J. L.
    Kim, B. H.
    Chang, S. Y.
    Lee, J. K.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 510 - 513
  • [22] DETERMINATION OF PRIMARY ELECTRON BEAM PARAMETERS IN MONTE CARLO SIMULATION OF MEDICAL LINAC
    Vaezzadeh, S. A.
    Allahverdi, M.
    Nedaie, H. A.
    Yarahmadi, M.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S518 - S518
  • [23] COMPARISON OF 3-DIMENSIONAL MONTE-CARLO SIMULATION AND THE PENCIL BEAM ALGORITHM FOR AN ELECTRON-BEAM FROM A LINEAR-ACCELERATOR
    MANFREDOTTI, C
    NASTASI, U
    RAGONA, R
    ANGLESIO, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 255 (1-2): : 355 - 359
  • [24] Comparison of graphite calorimeter dosimetry system with Monte Carlo simulation at an industrial electron beam accelerator
    Benny, P. G.
    Selvam, T. Palani
    Sarma, K. S. S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 703 : 98 - 101
  • [25] Validating Small Beam Output Factor of Macro Electron Monte Carlo Dose Calculation
    Yeo, I.
    Nie, W.
    Xu, Q.
    MEDICAL PHYSICS, 2022, 49 (06) : E832 - E832
  • [26] A study on scattered particles from medical linear accelerator head using monte carlo simulation
    Katayose, T. (u5856601@gs.metro-hs.ac.jp), (KEK High Energy Accelerator Research Organization):
  • [27] Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam
    Fukata K.
    Sugimoto S.
    Kurokawa C.
    Saito A.
    Inoue T.
    Sasai K.
    Radiological Physics and Technology, 2018, 11 (2) : 192 - 201
  • [28] MONTE CARLO SIMULATION OF ELECTRON DOSE DISTRIBUTIONS PRODUCED BY THE ELEKTA SYNERGY TM LINEAR ACCELERATOR
    Nedaie, H. A.
    Mosleh Shirazi, M. A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S475 - S475
  • [30] MONTE-CARLO CALCULATIONS OF ELECTRON-BEAM PARAMETERS FOR 3 PHILIPS LINEAR ACCELERATORS
    UDALESMITH, M
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01): : 85 - 105