ACCURACY OF DISPOSABLE BLOOD-PRESSURE TRANSDUCERS USED IN THE CRITICAL CARE SETTING

被引:7
|
作者
BAILEY, RH
BAUER, JH
YANOS, J
机构
[1] UNIV MISSOURI, DIV PULM CRIT CARE & ENVIRONM MED, COLUMBIA, MO USA
[2] UNIV MISSOURI, DIV NEPHROL, HYPERTENS RES SECT, COLUMBIA, MO USA
[3] UNIV MISSOURI, DEPT INTERNAL MED, COLUMBIA, MO USA
[4] HARRY S TRUMAN MEM VET HOSP, COLUMBIA, MO 65201 USA
[5] CASE WESTERN RESERVE UNIV, DEPT MED, CLEVELAND, OH 44106 USA
关键词
TRANSDUCERS; PRESSURE; BLOOD PRESSURE DETERMINATION; CRITICAL CARE; BLOOD PRESSURE; CENTRAL VENOUS PRESSURE; MANOMETRY; MONITORING; PHYSIOLOGICAL;
D O I
10.1097/00003246-199501000-00029
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To assess the initial accuracy, drift in calibration over several hours, and decline in accuracy over daily use of blood pressure transducers used in the critical care setting. Design: Prospective, criterion standard. Setting: Three critical care units in a university hospital. Subjects: Twenty-seven consecutively available high-pressure transducers were used to measure intra-arterial blood pressures, and 11 consecutively available low-pressure transducers were used to measure central venous and pulmonary arterial blood pressures. Interventions: High-pressure transducers were compared with a mercury column manometer, and low-pressure transducers were compared with a water column manometer at three pressure levels. Data were collected initially after installation of the pressure transducer, at 3 hrs, and at 6 hrs. Data collections were repeated on subsequent days. Measurements and Main Results: a) The majority of pressure transducers are accurate; b) there is no drift in the accuracy of pressure transducers over several hours; c) the accuracy of pressure transducers does not decline over subsequent days. Some transducers overestimated standard pressures by 10 mm Hg and/or underestimated standard pressure by 17 mm Hg. Conclusions: The majority of pressure transducers used in the critical care setting are accurate, although there are some transducers that may compromise patient care. Blood pressure transducers should be checked against a standard manometer upon installation, and daily during use.
引用
收藏
页码:187 / 192
页数:6
相关论文
共 50 条
  • [21] THE EFFECT OF THE DENTAL SETTING ON BLOOD-PRESSURE MEASUREMENT
    DIANGELIS, N
    LUEPKER, RV
    AMERICAN JOURNAL OF PUBLIC HEALTH, 1983, 73 (10) : 1210 - 1212
  • [22] THE INFLUENCE OF THE DENTAL SETTING ON BLOOD-PRESSURE MEASUREMENT
    DIANGELIS, N
    LUEPKER, RV
    CIRCULATION, 1981, 64 (04) : 81 - 81
  • [23] ACCURACY OF INDIRECT BLOOD-PRESSURE MEASUREMENT IN THE ELDERLY
    OCALLAGHAN, WG
    FITZGERALD, DJ
    OMALLEY, K
    OBRIEN, E
    BRITISH MEDICAL JOURNAL, 1983, 286 (6377): : 1545 - 1546
  • [24] THE ACCURACY OF AUSCULTATORY BLOOD-PRESSURE MEASUREMENT IN THE ELDERLY
    NIELSEN, PE
    ACTA MEDICA SCANDINAVICA, 1983, : 39 - 44
  • [25] ACCURACY OF OSCILLOMETRIC BLOOD-PRESSURE MEASUREMENTS IN PREECLAMPSIA
    SHENNAN, A
    DESWIET, M
    HALLIGAN, A
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1995, 172 (04) : 1325 - 1325
  • [26] ACCURACY OF INDIRECT DETERMINATION OF BLOOD-PRESSURE IN DOGS
    GLEN, JB
    RESEARCH IN VETERINARY SCIENCE, 1973, 14 (03) : 291 - 295
  • [27] BLOOD-PRESSURE MEASUREMENT IN PRIMARY CARE
    MASTERS, N
    PRACTITIONER, 1988, 232 (1455) : 1042 - &
  • [28] BLOOD-PRESSURE IN A COMMUNITY SETTING - MEASUREMENT PRACTICES AND CONSEQUENCES
    STUDNEY, DR
    KUNSTAETTER, R
    RABKIN, S
    CLINICAL RESEARCH, 1982, 30 (01): : A44 - A44
  • [29] THE ACCURACY AND PRECISION OF THE ACCUTRACKER AMBULATORY BLOOD-PRESSURE MONITOR
    APPEL, LJ
    WHELTON, PK
    SEIDLER, AJ
    PATEL, AR
    KLAG, MJ
    AMERICAN JOURNAL OF EPIDEMIOLOGY, 1990, 132 (02) : 343 - 354
  • [30] SELF BLOOD-PRESSURE DEVICES ARE NOT FREQUENTLY CHECKED FOR ACCURACY
    GRIM, CM
    GRIM, CE
    PREVENTIVE MEDICINE, 1978, 7 (01) : 79 - 79