QUALITY ASSURANCE TOOL FOR HIGH-DOSE-RATE BRACHYTHERAPY

被引:18
|
作者
DEWERD, LA [1 ]
JURSINIC, P [1 ]
KITCHEN, R [1 ]
THOMADSEN, BR [1 ]
机构
[1] UNIV WISCONSIN,DEPT HUMAN ONCOL,MADISON,WI 53706
关键词
HIGH DOSE RATE BRACHYTHERAPY; QUALITY ASSURANCE; IR-192 REMOTE AFTERLOADERS; CALIBRATION;
D O I
10.1118/1.597469
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An insert for a well ionization chamber is described for quality assurance tests for high dose rate remote afterloaders. The use of this insert over time has shown its applicability for efficient QA tests for the high dose rate afterloader. The measurement results are six times greater than the uncertainties involved for the measurements. Checks on position can be made better than 1 mm and dwell times better than 1 s. The overall measurements can be done to better than 1.5%. QA checks can be accomplished in a shorter time period with as great or better accuracy than prior film techniques. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:435 / 440
页数:6
相关论文
共 50 条
  • [1] A quality assurance tool for high-dose-rate brachytherapy
    Rickey, Daniel W.
    Sasaki, David
    Bews, Jeff
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2525 - 2532
  • [2] QUALITY ASSURANCE PROGRAM FOR HIGH-DOSE-RATE BRACHYTHERAPY
    MOWER, HW
    COSTEY, GE
    LAWSON, BJ
    WRIGHT, KA
    [J]. RADIOLOGY, 1992, 185 : 273 - 273
  • [3] Quality assurance measurements for high-dose-rate brachytherapy without film
    Jursinic, Paul A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01): : 246 - 261
  • [4] Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy
    Yogo, Katsunori
    Matsushita, Akihiro
    Tatsuno, Yuya
    Shimo, Takahiro
    Hirota, Seiko
    Nozawa, Marika
    Ozawa, Shuichi
    Ishiyama, Hiromichi
    Yasuda, Hiroshi
    Nagata, Yasushi
    Hayakawa, Kazushige
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [5] Fluorescent screen for high-dose-rate (HDR) brachytherapy quality assurance
    Lightstone, AW
    [J]. MEDICAL DOSIMETRY, 2005, 30 (03) : 143 - 144
  • [6] Imaging Cherenkov emission for quality assurance of high-dose-rate brachytherapy
    Katsunori Yogo
    Akihiro Matsushita
    Yuya Tatsuno
    Takahiro Shimo
    Seiko Hirota
    Marika Nozawa
    Shuichi Ozawa
    Hiromichi Ishiyama
    Hiroshi Yasuda
    Yasushi Nagata
    Kazushige Hayakawa
    [J]. Scientific Reports, 10
  • [7] Quality assurance of treatment plans for interstitial and intracavitary high-dose-rate brachytherapy
    Das, Rupak K.
    Bradley, Kristin A.
    Nelson, Ian A.
    Patel, Rakesh
    Thomadsen, Bruce R.
    [J]. BRACHYTHERAPY, 2006, 5 (01) : 56 - 60
  • [9] A quality assurance device for measuring afterloader performance and transit dose for nasobiliary high-dose-rate brachytherapy
    Deufel, Christopher L.
    Mullins, John P.
    Zakhary, Mark J.
    [J]. BRACHYTHERAPY, 2018, 17 (04) : 726 - 731
  • [10] HIGH-DOSE-RATE BRACHYTHERAPY
    SPEISER, BL
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 : 116 - 116