摘要
Processing of Doppler signals produced by pulsed Doppler systems is based on the assumption that the phase of the received high frequency ultrasound signals changes linearly with depth. However, the random spatial distribution of scatterers is not in accordance with this basic assumption. Consequently, averaging af the demodulated signal over an observation window, covering a few periods of the received signal, does not improve the estimate for the instantaneous quadrature components of the Doppler signal originating from a given depth. Hence, the accuracy of the Doppler velocity estimate is independent of the length of the observation window employed. However, splitting the observation window in subsample volumes, each with a length of one period at the emission frequency, and combining the Doppler signals of the subsample volumes at the last stage of signal processing, i.e., mean Doppler frequency estimation using the autocorrelation technique, results in a consideration reduction of the variance of the velocity estimate. Using a computer simulation of the signal processing involved, it is demonstrated that with subsample volume processing the variance of the velocity estimate attains the same variance as is expected for the RF cross correlation technique.
引用
收藏
页码:953 / 965
页数:13
相关论文
共 50 条
  • [41] Time-scale analysis of quadrature Doppler ultrasound signals
    Aydin, N
    Markus, HS
    IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 2001, 148 (01) : 15 - 22
  • [42] Doppler ultrasound monitoring for detection of microembolic signals in peripheral arteries
    Kudo, T
    Inoue, Y
    Sugano, N
    Iwai, T
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2002, 24 (01) : 37 - 42
  • [43] Analysis of Doppler ultrasound signals based on fuzzy similarity index
    Güler, I
    Übeyli, ED
    EXPERT SYSTEMS WITH APPLICATIONS, 2006, 31 (01) : 130 - 136
  • [44] PREDICTIVE ANALOG-TO-DIGITAL CONVERSION OF DOPPLER ULTRASOUND SIGNALS
    BOE, S
    KRISTOFFERSEN, K
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (03) : 260 - 268
  • [45] Simulation and feature analysis for gas emboli Doppler ultrasound signals
    WANG Teng
    WANG Yuanyuan
    Chinese Journal of Acoustics, 2013, 32 (01) : 79 - 89
  • [47] DIGITAL AUDIO TAPE AS A METHOD OF STORING DOPPLER ULTRASOUND SIGNALS
    BUSH, G
    EVANS, DH
    PHYSIOLOGICAL MEASUREMENT, 1993, 14 (03) : 381 - 386
  • [48] SIMULATION OF DOPPLER ULTRASOUND SIGNALS FOR A LAMINAR, PULSATILE, NONUNIFORM FLOW
    WENDLING, F
    JONES, SA
    GIDDENS, DP
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1992, 18 (02): : 179 - 193
  • [49] The information content of Doppler ultrasound signals from the fetal heart
    S. A. Shakespeare
    J. A. Crowe
    B. R. Hayes-Gill
    K. Bhogal
    D. K. James
    Medical and Biological Engineering and Computing, 2001, 39 : 619 - 626
  • [50] Processing of Transcranial Doppler for Assessment of Blood Volume Loss
    Hakimzadeh, Roya
    Ji, Soo-Yeon
    Smith, Rebecca
    Ward, Kevin
    Daneshvar, Kasra
    Najarian, Kayvan
    PROCEEDINGS OF THE 2009 IEEE INTERNATIONAL CONFERENCE ON INFORMATION REUSE AND INTEGRATION, 2008, : 6 - +