Measurement verification of dose distributions in pulsed-dose rate brachytherapy in breast cancer

被引:2
|
作者
Mantaj, Patrycja [1 ]
Zwierzchowski, Grzegorz [2 ,3 ]
机构
[1] Greater Poland Canc Ctr, Dept Radiat Protect, Garbary 15, PL-61866 Poznan, Poland
[2] Greater Poland Canc Ctr, Dept Med Phys, PL-61866 Poznan, Poland
[3] Poznan Univ Med Sci, Dept Electroradiol, Poznan, Poland
关键词
Pulse brachytherapy; Implant; Thermoluminescent detectors; Gafchromic EBT dosimetry;
D O I
10.1016/j.rpor.2013.03.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: The aim of the study was to verify the dose distribution optimisation method in pulsed brachytherapy. Background: The pulsed-dose rate brachytherapy is a very important method o f breast tumour treatment using a standard brachytheraphy equipment. The appropriate dose distribution round an implant is an important issue in treatment planning. Advanced computer systems of treatment planning are equipped with algorithms optimising dose distribution. Materials and methods: The wax-paraffin phantom was constructed and seven applicators were placed within it. Two treatment plans (non-optimised, optimised) were prepared. The reference points were located at a distance of 5 mm from the applicators' axis. Thermoluminescent detectors were placed in the phantom at suitable 35 chosen reference points. Results: The dosimetry verification was carried out in 35 reference points for the plans before and after optimisation. Percentage difference for the plan without optimisation ranged from -8.5% to 1.4% and after optimisation from -8.3% to 0.01%. In 16 reference points, the calculated percentage difference was negative (from -8.5% to 1.3% for the plan without optimisation and from -8.3% to 0.8% for the optimised plan). In the remaining 19 points percentage difference was from 9.1% to 1.4% for the plan without optimisation and from 7.5% to 0.01% for the optimised plan. No statistically significant differences were found between calculated doses and doses measured at reference points in both dose distribution non-optimised treatment plans and optimised treatment plans. Conclusions: No statistically significant differences were found in dose values at reference points between doses calculated by the treatment planning system and those measured by TLDs. This proves the consistency between the measurements and the calculations. (C) 2013 Published by Elsevier Urban & Partner Sp. z o.o. on behalf of Greater Poland Cancer Centre.
引用
收藏
页码:139 / 147
页数:9
相关论文
共 50 条
  • [41] A dose verification tool for high-dose-rate interstitial brachytherapy treatment planning in accelerated partial breast irradiation
    Marqa, Mohamad Feras
    Caudrelier, Jean-Michel
    Betrouni, Nacim
    BRACHYTHERAPY, 2012, 11 (05) : 359 - 368
  • [42] Dose volume histogram analysis of MRI-based intracavitary pulsed dose rate brachytherapy in cervical cancer
    Lindegaard, J.
    Tanderup, K.
    Gelineck, J.
    Nielsen, S. K.
    Christensen, J. J.
    Christensen, T. B.
    Nyvang, G-B.
    Jurik, A. G.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S78 - S78
  • [43] Low-dose-rate brachytherapy is superior to high-dose-rate brachytherapy for bladder cancer
    Pos, FJ
    Horenblas, S
    Lebesque, J
    Moonen, L
    Schneider, C
    Sminia, P
    Bartelink, H
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (03): : 696 - 705
  • [44] OUR EXPERIENCE IN LOW-DOSE-RATE BRACHYTHERAPY PREVIOUSLY TO START THE HIGH-DOSE-RATE BRACHYTHERAPY BOOST TECHNIQUE BREAST CANCER
    Castro, J. E.
    Gonzalez Patino, E.
    Salvador Garrido, N.
    Luna Vega, V.
    Varela Pazos, A.
    Anido Herranz, U.
    Lobato Busto, R.
    Porto Vazquez, C.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S201 - S201
  • [45] Sigmoid dose delivered by high-dose-rate brachytherapy versus low-dose-rate brachytherapy for cervical cancer
    Holloway, C. L.
    O'Farrell, D. A.
    Cormack, R. A.
    Viswanathan, A. N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S165 - S165
  • [46] Thermoluminescence dosimetry dose verification of a high dose rate brachytherapy treatment planning in phantom
    Sakulsingharoj, S.
    Tangboonduangjit, P.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S328 - S328
  • [47] An Algorithm to Calculate Biologically Effective Dose Distributions for Low Dose Rate Prostate Brachytherapy
    Guest, D.
    Riegel, A.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [48] CT based dose distribution and optimization for gynaecological pulsed dose rate interstitial brachytherapy
    Brunaud, C
    Ahmad, F
    Peiffert, D
    Metayer, Y
    Aletti, P
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S78 - S78
  • [49] Thermoluminescence dosimetry for in-vivo verification of high dose rate brachytherapy for prostate cancer
    R. Das
    W. Toye
    T. Kron
    S. Williams
    G. Duchesne
    Australasian Physics & Engineering Sciences in Medicine, 2007, 30 : 178 - 184
  • [50] Thermoluminescence dosimetry for in-vivo verification of high dose rate brachytherapy for prostate cancer
    Das, R.
    Toye, W.
    Kron, T.
    Williams, S.
    Duchesne, G.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) : 178 - 184