Electrosurgery in Patients With Implantable Electronic Cardiac Devices (Pacemakers and Defibrillators)

被引:6
|
作者
Garcia Bracamonte, B. [1 ]
Rodriguez, J. [2 ]
Casado, R. [3 ]
Vanaclocha, F. [1 ]
机构
[1] Hosp Univ 12 Octubre, Serv Dermatol, Madrid, Spain
[2] Hosp Univ 12 Octubre, Serv Cardiol, Unidad Marcapasos, Madrid, Spain
[3] Hosp Univ 12 Octubre, Serv Anestesia, Madrid, Spain
来源
ACTAS DERMO-SIFILIOGRAFICAS | 2013年 / 104卷 / 02期
关键词
Electrosurgical unit; Electrosurgery; Electromagnetic interference; Pacemaker; Defibrillator;
D O I
10.1016/j.ad.2012.09.012
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
The electrosurgical unit is a very useful tool widely used in dermatology to treat benign and malignant skin lesions and to achieve hemostasis during surgery. However, precautions are required when this technique is used in patients with implantable electronic cardiac devices (IECD), such as pacemakers and defibrillators, because electromagnetic interference produced by the tool may cause such devices to malfunction. Before using electrosurgery in patients with IECDs, it is essential to ascertain the type of implanted device and the patient's level of dependence on it. The location of the skin lesion to be treated with respect to the device should also be assessed. Bipolar pacemakers are more resistant to interference. Appropriate monitoring and the use of bipolar forceps are recommended. (C) 2012 Elsevier Espana, S.L. and AEDV. All rights reserved.
引用
收藏
页码:128 / 132
页数:5
相关论文
共 50 条
  • [31] Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy
    Solan, AN
    Solan, MJ
    Bednarz, G
    Goodkin, MB
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (03): : 897 - 904
  • [32] Small Bowel Capsule Endoscopy in Patients With Implantable Cardiac Defibrillators and Permanent Pacemakers
    Luthra, Anjuli
    Evans, John
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2014, 109 : S556 - S556
  • [33] CARDIAC PACEMAKERS IN ELECTROSURGERY
    KOHLER, FP
    MACKINNE.CC
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1965, 193 (10): : 855 - &
  • [34] Recommendations for cardiacmagnetic resonance imaging in patients with cardiac pacemakers and implantable cardioverter defibrillators
    von Knobelsdorff-Brenkenhoff, F.
    Bauer, W. R.
    Deneke, T.
    Fleck, E.
    Rolf, A.
    Schulz-Menger, J.
    Sommer, P.
    Tillmanns, C.
    Eitel, I.
    KARDIOLOGE, 2019, 13 (02): : 75 - 86
  • [35] Practice Advisory for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Pacemakers and Implantable Cardioverter-Defibrillators 2020: An Updated Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Implantable Electronic Devices (vol 132, pg 225, 2020)
    Apfelbaum, Jeffrey L.
    Schulman, Peter M.
    Mahajan, Aman
    Connis, Richard T.
    Agarkar, Madhulika
    ANESTHESIOLOGY, 2020, 132 (04) : 938 - 938
  • [36] Pacemakers and implantable cardioverter defibrillators
    Allen, M.
    ANAESTHESIA, 2006, 61 (09) : 883 - 890
  • [37] Pacemakers and implantable cardioverter defibrillators
    Rozner, M
    CRITICAL CARE MEDICINE, 2004, 32 (08) : 1809 - 1812
  • [38] Pacemakers and defibrillators in patients with cardiac sarcoidosis
    Solomon, AJ
    Yeager, H
    Tracy, CM
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (06) : 197A - 197A
  • [39] Safety Profile of Liver FibroScan in Patients with Cardiac Pacemakers or Implantable Cardioverter-Defibrillators
    Chan, Yin
    Pranke, Stephanie
    Rashidi, Farid
    Nosib, Shravan
    Worobetz, Lawrence
    CANADIAN JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2017, 2017
  • [40] Cardiac MRI for Patients With Cardiac Implantable Electronic Devices
    Stojanovska, Jadranka
    Runge, Mason
    Mahani, Maryam Ghadimi
    Cronin, Paul P.
    Sayyouh, Mohamed
    Bogun, Frank
    Ibrahim, El-Sayed H.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (02) : 374 - 381