THE COMPARISON OF INTRAOCULAR PRESSURE MEASUREMENT BY GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR PUFF TONOMETER: A CROSS-SECTIONAL RESEARCH

被引:0
|
作者
Arooj, Bushra [1 ]
Nadeem, Asmarah [1 ]
Ali, Sobia [1 ]
机构
[1] Lahore Gen Hosp, Lahore, Pakistan
来源
关键词
Tonometry; Glaucoma; Goldmann Applanation Tonometer and Air Puff Tonometer;
D O I
10.5281/zenodo.2526307
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
Objective: The aim of this particular research was to determine the accuracy and frequency of intraocular pressure which measured with the help of an Air Puff (non-contact tonometer); moreover, we also confirmed the same by Goldmann applanation tonometer. Material and methods: Our research design was cross-sectional and it was carried out at Ophthalmology Department of Mayo Hospital, Lahore (February to July 2017) on a total of 240 patients. All the patients were in the age bracket of (10 - 70) years including males and females. The selection process included those patients who visited Eye OPD for regular intraocular pressure, refraction, glaucoma, cataract surgery and suspected glaucoma. A complete ocular assessment was mandatory for every individual participating in the research. Results: In the age bracket of (10 - 70) years the mean age was (43.844 +/- 15.434) years and the values of mean IOP was for Goldmann Applanation Tonometer as (18.92 +/- 8.852) mmHg and for Air Puff Tonometer as (21.463 +/- 9.456) mmHg. The different measurement ranges for Air-Puff Tonometer and Goldmann Applanation Tonometer was respectively (10 - 54) mmHg and (11 - 58) mmHg. Conclusion: In order to measure intraocular pressure, the method of Air puff tonometer is non-contact and quick. Air Puff tonometer is effective for screening purposes; whereas, the intraocular pressure measurements are accurately measured through Goldmann applanation tonometer.
引用
收藏
页码:16875 / 16879
页数:5
相关论文
共 50 条
  • [41] Comparison of Goldmann Applanation Tonometer, Noncontact Tonometer, and TonoPen XL for Intraocular Pressure Measurement in Different Types of Glaucomatous, Ocular Hypertensive, and Normal Eyes
    Kim, Na Rae
    Kim, Chan Yun
    Kim, Hyunjoong
    Seong, Gong Je
    Lee, Eun Suk
    CURRENT EYE RESEARCH, 2011, 36 (04) : 295 - 300
  • [42] MEASUREMENT OF INTRAOCULAR-PRESSURE WITH A PNEUMATIC APPLANATION TONOMETER
    WEST, CE
    CAPELLA, JA
    KAUFMAN, HE
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1972, 74 (03) : 505 - &
  • [43] NON-CONTACT (AIR-PUFF) TONOMETER - VARIABILITY AND CORNEAL STRAINING
    MYERS, KJ
    SCOTT, CA
    AMERICAN JOURNAL OF OPTOMETRY AND PHYSIOLOGICAL OPTICS, 1975, 52 (01): : 36 - 46
  • [44] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770
  • [45] Comparison of disposable Goldmann applanation tonometer and ICare ic100 to standards of care Goldmann nondisposable applanation tonometer and Tono-pen XL for Measuring Intraocular Pressure
    Wong, Benny
    Parikh, Deep
    Angert, Mallorie
    Gorski, Matthew
    Shih, Carolyn
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [46] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [47] Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects
    Guler, Mete
    Bilak, Semsettin
    Bilgin, Burak
    Simsek, Ali
    Capkin, Musa
    Reyhan, Ali Hakim
    JOURNAL OF GLAUCOMA, 2015, 24 (08) : 613 - 618
  • [48] Reproducibility of intraocular pressure self-measurement by ICare Home rebound tonometer and comparison with Goldman applanation tonometer
    Gaillard, R.
    Fenolland, J. R.
    Bonnel, S.
    Rosenberg, R.
    Marill, A. F.
    Bostanci, K.
    Giraud, J. M.
    Renard, J. P.
    ACTA OPHTHALMOLOGICA, 2015, 93
  • [49] Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma Patients
    Salim, Sarwat
    Du, Haiming
    Wan, Jim
    JOURNAL OF GLAUCOMA, 2013, 22 (04) : 325 - 329
  • [50] IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry
    Yang Zhang
    Lin Zheng
    Ailing Bian
    Qi Zhou
    International Ophthalmology, 2018, 38 : 697 - 703