DETERMINATION OF TRACE AMOUNTS OF LEAD BY CHELATING ION-EXCHANGE AND ONLINE PRECONCENTRATION IN FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY

被引:26
|
作者
PUROHIT, R [1 ]
DEVI, S [1 ]
机构
[1] MAHARAJA SAYAJIRAO UNIV BARODA,FAC SCI,DEPT CHEM,BARODA 390002,GUJARAT,INDIA
关键词
ATOMIC ABSORPTION SPECTROMETRY; ION EXCHANGE; FLOW SYSTEM; CHELATING RESINS; LEAD; PRECONCENTRATION;
D O I
10.1016/0003-2670(92)85074-G
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Resins synthesized from quinolin-8-ol and resorcinol or hydroquinone, with furfuraldehyde, formaldehyde or benzaldehyde as cross-linking agent, were used for the preconcentration of nanogram amounts of lead. The rate of exchange and activation energy of lead exchange were calculated. Column separations of lead-copper and lead-zinc did not show any cross-contamination. A continuous-flow manifold using resin microcolumns was developed for the preconcentration and determination of lead.
引用
收藏
页码:53 / 60
页数:8