Determination of trace amounts of lead in mussels by flow-injection flame atomic-absorption spectrometry coupled with on-line minicolumn preconcentration

被引:14
|
作者
Yebra, MC [1 ]
Enríquez, MF [1 ]
García, A [1 ]
Moreno-Cid, A [1 ]
机构
[1] Univ Santiago, Fac Chem, Dept Analyt Chem Nutr & Bromatol, E-15706 Santiago, Spain
来源
关键词
D O I
10.1007/s002160100729
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A minicolumn packed with poly(aminophosphonic acid) chelating resin incorporated in an on-line preconcentration system for flame atomic-absorption spectrometry was used to determine ultratrace amounts of lead in mussel samples at mug L-1 level. The preconcentrated lead was eluted with hydrochloric acid and injected directly into the nebulizer for atomization in an air-acetylene flame for measurement. The performance characteristics of the determination of lead were: preconcentration factor 26.8 for 1 min preconcentration time, detection limit (3 sigma) in the sample digest was 0.25 mug g(-1) (dry weight) for a sample volume of 3.5 mL and 0.2 g sample (preconcentration time 1 min), precision (RSD) 2.3% for 25 mug L-1 and 2.0% for 50 mug L-1. The sampling frequency was 45 h(-1). The method was highly tolerant of interferences, and the results obtained for the determination of lead in a reference material testify to the applicability of the proposed procedure to the determination of lead at ultratrace level in biological materials such as mussel samples.
引用
收藏
页码:64 / 68
页数:5
相关论文
共 50 条