共 50 条
ONLINE PRECONCENTRATION AND DETERMINATION OF TRACE PLATINUM BY FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY
被引:64
|作者:
CANTARERO, A
[1
]
GOMEZ, MM
[1
]
CAMARA, C
[1
]
PALACIOS, MA
[1
]
机构:
[1] UNIV COMPLUTENSE MADRID,FAC QUIM,DEPT QUIM ANALIT,E-28040 MADRID,SPAIN
关键词:
ATOMIC ABSORPTION SPECTROMETRY;
FLOW INJECTION;
ALUMINA;
MICROCOLUMNS;
PLATINUM;
PRECONCENTRATION;
WATERS;
D O I:
10.1016/0003-2670(94)80264-5
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The analytical performance of a platinum preconcentration method using a 0.01 M HNO3 carrier solution and an alumina microcolumn is discussed. On-line preconcentration is followed by flame atomic absorption spectrometry (FAAS), and off-line preconcentration by graphite furnace atomic absorption spectrometry (GFAAS). The preconcentration factors were 600 for FAAS (25 mu l elution volume) and 30 for GFAAS (500 mu l elution volume), both with a 15-ml sampling volume. The detection limits in these conditions were 0.02 mg l(-1), and relative standard deviation (R.S.D.) of 9% (0.1 mg 1(-1) solution, n = 5) for FAAS and 0.33 mu g l(-1), and a R.S.D. of 7% (5 mu g l(-1) solution n = 5) for GFAAS. The proposed method is suitable for platinum determination in natural water samples.
引用
收藏
页码:205 / 211
页数:7
相关论文