ONLINE PRECONCENTRATION AND DETERMINATION OF TRACE PLATINUM BY FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY

被引:64
|
作者
CANTARERO, A [1 ]
GOMEZ, MM [1 ]
CAMARA, C [1 ]
PALACIOS, MA [1 ]
机构
[1] UNIV COMPLUTENSE MADRID,FAC QUIM,DEPT QUIM ANALIT,E-28040 MADRID,SPAIN
关键词
ATOMIC ABSORPTION SPECTROMETRY; FLOW INJECTION; ALUMINA; MICROCOLUMNS; PLATINUM; PRECONCENTRATION; WATERS;
D O I
10.1016/0003-2670(94)80264-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The analytical performance of a platinum preconcentration method using a 0.01 M HNO3 carrier solution and an alumina microcolumn is discussed. On-line preconcentration is followed by flame atomic absorption spectrometry (FAAS), and off-line preconcentration by graphite furnace atomic absorption spectrometry (GFAAS). The preconcentration factors were 600 for FAAS (25 mu l elution volume) and 30 for GFAAS (500 mu l elution volume), both with a 15-ml sampling volume. The detection limits in these conditions were 0.02 mg l(-1), and relative standard deviation (R.S.D.) of 9% (0.1 mg 1(-1) solution, n = 5) for FAAS and 0.33 mu g l(-1), and a R.S.D. of 7% (5 mu g l(-1) solution n = 5) for GFAAS. The proposed method is suitable for platinum determination in natural water samples.