An institutional experience of quality assurance of a treatment planning system on photon beam

被引:3
|
作者
Ozguven, Yildiray [1 ]
Yaray, Kadir [2 ]
Alkaya, Fadime [3 ]
Yucel, Birsen [1 ]
Soyuer, Serdar [2 ]
机构
[1] Cumhuriyet Univ, Fac Med, Dept Radiat Oncol, TR-58140 Sivas, Turkey
[2] Erciyes Univ, Fac Med, Dept Radiat Oncol, TR-38039 Kayseri, Turkey
[3] Medicana Hosp, Dept Radiat Oncol, TR-34520 Istanbul, Turkey
关键词
Treatment planning system; Quality assurance; Local dose deviation; Confidence limit; Photon beam;
D O I
10.1016/j.rpor.2013.10.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: The purpose of the present study is to show the application of the IAEA TRS-430 QA procedures of Eclipse (TM) v7.5 TPS for photon energies. In addition, the trends of the deviations found in the conducted tests were determined. Background: In the past, the lack of complete TPS QA procedures led to some serious accidents. So, QA in the radiotherapy treatment planning process is essential for determination of accuracy in the radiotherapy process and avoidance of treatment errors. Materials and methods: The calculations of TPS and measurements of irradiations of the treatment device were compared in the study. As a result, the local dose deviation values (delta(1) : central beam axis, delta(2): penumbra and build up region, delta(3) : inside field, delta(4) : outside beam edges, delta(50-90): beam fringe, RW50 : radiological width) and their confidence limit values (including systematic and random errors) were obtained. Results: The confidence limit values of delta(4) were detected to increase with expanding field size. The values of delta(1) and delta(3) of hard wedge were larger than open fields. The values of delta(2) and delta(50-90) of the inhomogeneity effect test were larger, especially than other tests of this study. The average deviation was showed to increase with the rise of the wedge angle. The values of delta(3) and delta(4) of lung irradiation were outside tolerance. Conclusions: The QA of TPS was done and it was found that there were no reservations in its use in patient treatment. The trend of the deviations is shown. (C) 2013 Greater Poland Cancer Centre. Published by Elsevier Urban & Partner Sp. z o.o. All rights reserved.
引用
收藏
页码:195 / 205
页数:11
相关论文
共 50 条
  • [11] First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy
    Karius, Andre
    Strnad, Vratislav
    Lotter, Michael
    Kreppner, Stephan
    Bert, Christoph
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2022, 198 (06) : 573 - 581
  • [12] First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy
    Andre Karius
    Vratislav Strnad
    Michael Lotter
    Stephan Kreppner
    Christoph Bert
    [J]. Strahlentherapie und Onkologie, 2022, 198 : 573 - 581
  • [13] AI-based radiotherapy treatment planning quality assurance: A multi-institutional study
    Kalendralis, P.
    Luk, S. M.
    Canters, R.
    Eyssen, D.
    Vaniqui, A.
    Wolfs, C.
    Murrer, L.
    van Elmpt, W.
    Kalet, A. M.
    Dekker, A.
    van Soest, J.
    Fijten, R.
    Zegers, C. M.
    Bermejo, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S243 - S244
  • [14] First experience with the ESTRO-QUASIMODO tests for quality assurance of treatment planning systems
    Gavryienko, M
    Olszewska, AM
    van Vliet-Vroegindeweij, C
    Pidlubna, T
    Mijnheer, BJ
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S109 - S109
  • [15] COMMISSIONING AND QUALITY ASSURANCE OF TREATMENT PLANNING COMPUTERS
    VANDYK, J
    BARNETT, RB
    CYGLER, JE
    SHRAGGE, PC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (02): : 261 - 273
  • [16] Development of Accurate/Advanced Radiotherapy Treatment Planning and Quality Assurance System (ARTS)
    Wu Yi-Can
    Song Gang
    Cao Rui-Fen
    Wu Ai-Dong
    Cheng Meng-Yun
    Tang Zheng-Dao
    Li Gui
    Jing Jia
    Liu Hui
    Li Jia
    Lan Hai-Yang
    Meng Yao
    Zheng Hua-Qing
    Jin Chu-Feng
    Zeng Qin
    Zheng Shan-Liang
    Huang Qun-Ying
    Wu Li-Jun
    Li Guo-Li
    Lin Hui
    Tao Sheng-Xiang
    Shi Chang-Bei
    [J]. CHINESE PHYSICS C, 2008, 32 : 177 - 182
  • [17] Quality Assurance of Personal Radiation Shield for Kilovoltage Photon: A Multicentre Experience
    Bawazeer, Omemh
    [J]. RISK MANAGEMENT AND HEALTHCARE POLICY, 2021, 14 : 1263 - 1270
  • [18] Automated treatment planning and quality assurance in radiation oncology
    Purdie, Thomas G.
    [J]. CLINICAL CANCER RESEARCH, 2021, 27 (05)
  • [19] Automation for treatment planning and quality assurance (actual advancements)
    Deufel, C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 158 : S93 - S93
  • [20] Quality assurance of treatment planning systems:: The QUASIMODO project
    Olszewska, A
    Fiorino, C
    Knöös, T
    Mijnheer, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S73 - S73