Determination of Working Length with Three Different Apex Locators

被引:0
|
作者
Kqiku, Lumnije [1 ]
Hoxha, Veton [2 ]
Stadtler, Peter [1 ]
机构
[1] Univ Dent Clin, Dept Conservat Dent, Auenbruggerpl 6a, A-8036 Graz, Austria
[2] Univ Dent Clin, Dept Conservat Dent & Endodont, Prishtina, Serbia
关键词
endodontics; root canal therapy; dental equipment;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to test the accuracy three different apex locators. A total of 20 extracted single - rooted human teeth were used for this study. The teeth were embedded in an alginate (Alginate Plus, Henry Schein, Melville, NY, USA) model developed for electric working length determination with apex locators. Electronic length measurement was achieved with three different apex locators: Propex Apex Locator (Dentsply/Maillefer, Montigny le Bretonneux, Switzerland), Root ZX (Morita Europe, Dietzenbach, Germany), Elements Diagnostic Unit Apex Locator (Sybron Endo, Glendora, California, USA) using a size 15 K file (VDW GmbH, Munich, Germany). After that the working length was determined radiographically by placing a size 15 file to the root canal, then subtracting 1 mm from apex. Each root canal was prepared in the apical third and the distance between the instrument's tip to the apical foramen was analyzed under stereomicroscope. The results were subjected to statistical analysis with ANOVA using the SPSS 11.0 program. There were no significant differences between the different apex locators used and the radiographic control in determining working lengths. Electronic working length measurement with different apex locators used in this study proved the same results as the radiological control of canal length determination.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 50 条
  • [31] Clinical reproducibility of three electronic apex locators
    Miletic, V.
    Beljic-Ivanovic, K.
    Ivanovic, V.
    INTERNATIONAL ENDODONTIC JOURNAL, 2011, 44 (08) : 769 - 776
  • [32] Effects of three different embedding media on the accuracy of different electronic apex locators: An in vitro study
    Cudal, Hamza
    Kesim, Bertan
    Aslan, Tugrul
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2022, 13 (01): : 80 - 83
  • [33] The use of micro-computed tomography to determine the accuracy of electronic working length with two apex locators
    Suguro, Hisashi
    Nishihara, Anna
    Tamura, Takahito
    Nakamura, Takeshi
    Toyama, Yurika
    Takeichi, Osamu
    JOURNAL OF ORAL SCIENCE, 2021, 63 (02) : 167 - 169
  • [34] An Invitro Evaluation of the Accuracy of Two Electronic Apex Locators to Determine Working Length in the Presence of Various Irrigants
    Ashwini, Prasad B.
    Harshit, Srivastava
    Aastha, Srivastava Arora
    Deepak, Raisingani
    ETHIOPIAN JOURNAL OF HEALTH SCIENCES, 2016, 26 (05) : 457 - 462
  • [35] Coronal preflaring's influence on the accuracy of root canal length determination by electronic apex locators
    Batista, Larissa C. Novaes
    Janini, Ana Cristina Padilha
    Marciano, Marina A.
    Guimaraes, Tatiany G. F. Araujo
    Guimaraes, Bruno Martini
    REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2024, 65 (03): : 115 - 120
  • [36] An ex vivo comparison of three electronic apex locators
    Baginska, Joanna
    Linczuk, Ewa
    Botulinski, Bartlomiej
    Kierklo, Anna
    HEALTHMED, 2012, 6 (10): : 3275 - 3280
  • [37] In vitro evaluation of the accuracy of three electronic apex locators
    Lucena-Martín, C
    Robles-Gijón, V
    Ferrer-Luque, M
    de Mondelo, JMNR
    JOURNAL OF ENDODONTICS, 2004, 30 (04) : 231 - 233
  • [38] Determination of location of root perforations by electronic apex locators
    Fuss, Z
    Assooline, LS
    Kaufman, AY
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1996, 82 (03): : 324 - 329
  • [39] An in vivo comparison of five different electronic apex locators
    Al-Shaher, Abdulaziz A.
    Madi, Samar
    ENDO-ENDODONTIC PRACTICE TODAY, 2011, 5 (01): : 17 - 25
  • [40] Comparison of three different apex locators in determining the working length of mandibular first molar teeth with irreversible pulpitis compared with an intraoral periapical radiograph: A block randomized, controlled, clinical trial
    Vanitha, Sadashivam
    Sherwood, I. Anand