Determination of Working Length with Three Different Apex Locators

被引:0
|
作者
Kqiku, Lumnije [1 ]
Hoxha, Veton [2 ]
Stadtler, Peter [1 ]
机构
[1] Univ Dent Clin, Dept Conservat Dent, Auenbruggerpl 6a, A-8036 Graz, Austria
[2] Univ Dent Clin, Dept Conservat Dent & Endodont, Prishtina, Serbia
关键词
endodontics; root canal therapy; dental equipment;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to test the accuracy three different apex locators. A total of 20 extracted single - rooted human teeth were used for this study. The teeth were embedded in an alginate (Alginate Plus, Henry Schein, Melville, NY, USA) model developed for electric working length determination with apex locators. Electronic length measurement was achieved with three different apex locators: Propex Apex Locator (Dentsply/Maillefer, Montigny le Bretonneux, Switzerland), Root ZX (Morita Europe, Dietzenbach, Germany), Elements Diagnostic Unit Apex Locator (Sybron Endo, Glendora, California, USA) using a size 15 K file (VDW GmbH, Munich, Germany). After that the working length was determined radiographically by placing a size 15 file to the root canal, then subtracting 1 mm from apex. Each root canal was prepared in the apical third and the distance between the instrument's tip to the apical foramen was analyzed under stereomicroscope. The results were subjected to statistical analysis with ANOVA using the SPSS 11.0 program. There were no significant differences between the different apex locators used and the radiographic control in determining working lengths. Electronic working length measurement with different apex locators used in this study proved the same results as the radiological control of canal length determination.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 50 条
  • [21] Effect of Prefllaring on Accuracy of Endodontic Length Determination by Four Electronic Apex Locators
    Izadi, Arash
    Bayandari, Maliheh
    Kazeminejad, Ezatolah
    JOURNAL OF DENTISTRY INDONESIA, 2022, 29 (03) : 165 - 171
  • [22] Integrated electronic apex locators for determining working length: a systematic review of in-vitro studies
    La Rosa, Giusy Rita Maria
    Angjelova, Angela
    Jovanova, Elena
    Apostolska, Sonja
    Santos, Joao Miguel
    Pedulla, Eugenio
    ODONTOLOGY, 2024,
  • [23] An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen
    de Vasconcelos, Bruno Carvalho
    do Vale, Taina Macedo
    Teixeira de Menezes, Antonio Sergio
    Pinheiro-Junior, Elilton Cavalcante
    Vivacqua-Gomes, Nilton
    Bernardes, Ricardo Affonso
    Hungaro Duarte, Marco Antonio
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2010, 110 (02): : E57 - E61
  • [24] Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters
    Ebrahim, A. K.
    Wadachi, R.
    Suda, H.
    AUSTRALIAN DENTAL JOURNAL, 2006, 51 (03) : 258 - 262
  • [25] Efficacy of electronic apex locators in comparison with intraoral radiographs in working length determination- a systematic review and meta-analysis
    Kaur, Gurveen
    Thomas, Anchu Rachel
    Samson, Renu Sarah
    Varghese, Eby
    Ponraj, Ratna Rachel
    Nagraj, Sumanth Kumbargere
    Shrivastava, Deepti
    Algarni, Hmoud Ali
    Siddiqui, Amna Yusuf
    Alothmani, Osama S.
    Srivastava, Kumar Chandan
    BMC ORAL HEALTH, 2024, 24 (01)
  • [26] Comparison of various current electronic apex locators to determine the working length using the clearing technique
    Altunbas, D.
    Kustarci, A.
    Arslan, D.
    Er, K.
    Kocak, S.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2015, 18 (03) : 359 - 363
  • [27] In Vitro Comparison of Three Electronic Apex Locators
    Guise, George M.
    Goodell, Gary G.
    Imamura, Glen M.
    JOURNAL OF ENDODONTICS, 2010, 36 (02) : 279 - 281
  • [28] Accuracy of different electronic apex locators in determination of minimum Root perforation diameter
    Koc, Simay
    Kustarci, Alper
    Er, Kursat
    AUSTRALIAN ENDODONTIC JOURNAL, 2023, 49 : 179 - 186
  • [29] Determination of the Accuracy of 5 Electronic Apex Locators in the Function of Different Employment Protocols
    Oliveira, Tiago Nepomuceno
    Vivacqua-Gomes, Nilton
    Bernardes, Ricardo Affonso
    Vivan, Rodrigo Ricci
    Hungaro Duarte, Marco Antonio
    Vasconcelos, Bruno Carvalho
    JOURNAL OF ENDODONTICS, 2017, 43 (10) : 1663 - 1667
  • [30] A critical analysis of research methods and experimental models to study working length determination and the performance of apex locators - A narrative review with recommendations for the future
    ElAyouti, Ashraf
    Connert, Thomas
    Dummer, Paul
    Loest, Claus
    INTERNATIONAL ENDODONTIC JOURNAL, 2022, 55 : 281 - 294