Implementation of compensator-based intensity modulated radiotherapy with a conventional telecobalt machine using missing tissue approach

被引:0
|
作者
Tagoe, Samuel N. A. [1 ,2 ]
Mensah, Samuel Y. [2 ]
Fletcher, John J. [3 ]
机构
[1] Korle Bu Teaching Hosp, Natl Ctr Radiotherapy & Nucl Med, Accra, Ghana
[2] Univ Cape Coast, Coll Agr & Nat Sci, Sch Phys Sci, Dept Phys, Cape Coast, Ghana
[3] Univ Dev Studies, Dept Appl Phys, Navrongo Campus, Navrongo, Ghana
来源
关键词
telecobalt machine; bolus; compensator; Gafchromic film; intensity modulated;
D O I
10.2478/pjmpe-2018-0024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The present study aimed to generate intensity-modulated beams with compensators for a conventional telecobalt machine, based on dose distributions generated with a treatment planning system (TPS) performing forward planning, and cannot directly simulate a compensator. Materials and Methods: The following materials were selected for compensator construction: Brass, Copper and Perspex (PMMA). Boluses with varying thicknesses across the surface of a tissue-equivalent phantom were used to achieve beam intensity modulations during treatment planning with the TPS. Beam data measured for specific treatment parameters in a full scatter water phantom with a 0.125 cc cylindrical ionization chamber, with a particular compensator material in the path of beams from the telecobalt machine, and that without the compensator but the heights of water above the detector adjusted to get the same detector readings as before, were used to develop and propose a semiempirical equation for converting a bolus thickness to compensator material thickness, such that any point within the phantom would receive the planned dose. Once the dimensions of a compensator had been determined, the compensator was constructed using the cubic pile method. The treatment plans generated with the TPS were replicated on the telecobalt machine with a bolus within each beam represented with its corresponding compensator mounted on the accessory holder of the telecobalt machine. Results: Dose distributions measured in the tissue-equivalent phantom with calibrated Gafchromic EBT2 films for compensators constructed based on the proposed approach, were comparable to those of the TPS with deviation less than or equal to +/- 3% (mean of 2.29 +/- 0.61%) of the measured doses, with resultant confidence limit value of 3.21. Conclusion: The use of the proposed approach for clinical application is recommended, and could facilitate the generation of intensity-modulated beams with limited resources using the missing tissue approach rendering encouraging results.
引用
下载
收藏
页码:171 / 179
页数:9
相关论文
共 50 条
  • [41] A NEW APPROACH TO COMPUTING NORMAL TISSUE COMPLICATION PROBABILITY OF AN INTENSITY-MODULATED RADIOTHERAPY TREATMENT WITH STEREOTACTIC RADIOTHERAPY BOOST OF NASOPHARYNGEAL CARCINOMA: A CASE STUDY
    Leung, Lucullus H. T.
    Wong, Wicger K. H.
    Cheng, Ashley C. K.
    Kan, Monica W. K.
    Cheung, Joel Y. C.
    Lam, Nelson K. H.
    Yuen, K. T.
    Yau, C. C.
    MEDICAL DOSIMETRY, 2011, 36 (02) : 138 - 144
  • [42] Obtaining normal tissue constraints using intensity modulated radiotherapy (IMRT) in patients with oral cavity, oropharnygeal, and laryngeal carcinoma
    Skinner, W. J.
    Muse, E.
    Yaparpalvi, R.
    Guha, C.
    Garg, M. K.
    Kalnicki, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S472 - S472
  • [43] The normal tissue sparing obtained with simultaneous treatment of pelvic lymph nodes and bladder using intensity-modulated radiotherapy
    Sondergaard, Jimmi
    Hoyer, Morten
    Petersen, Jorgen B.
    Wright, Pauliina
    Grau, Cai
    Muren, Ludvig Paul
    ACTA ONCOLOGICA, 2009, 48 (02) : 238 - 244
  • [44] Using Image-guided Intensity-modulated Radiotherapy on Patients With Head and Neck Soft-tissue Sarcoma
    Elsayad, Khaled
    Stockmann, Denise
    Channaoui, Mohammed
    Scobioala, Sergiu
    Grajda, Aneta
    Berssenbruegge, Hendrik
    Huss, Sebastian
    Moustakis, Christos
    Haverkamp, Uwe
    Kleinheinz, Johannes
    Lenz, Georg
    Wardelmann, Eva
    Eich, Hans Theodor
    IN VIVO, 2019, 33 (04): : 1293 - 1300
  • [45] OBTAINING NORMAL TISSUE CONSTRAINTS USING INTENSITY MODULATED RADIOTHERAPY (IMRT) IN PATIENTS WITH ORAL CAVITY, OROPHARNYGEAL, AND LARYNGEAL CARCINOMA
    Skinner, William K. J.
    Muse, Evan D.
    Yaparpalvi, Ravindra
    Guha, Chandan
    Garg, Madhur K.
    Kalnicki, Shalom
    MEDICAL DOSIMETRY, 2009, 34 (04) : 279 - 284
  • [46] An in silico comparative dosimetric study of postmastectomy locoregional irradiation using intensity-modulated vs 3-dimensional conventional radiotherapy
    Khullar, Pooja
    Garg, Charu
    Sinha, Sujit Nath
    Kaur, Inderjit
    Datta, Niloy Ranjan
    MEDICAL DOSIMETRY, 2018, 43 (04) : 370 - 376
  • [47] Clinical Implementation of Automated Treatment Planning for Rectum Intensity-Modulated Radiotherapy on UIH Using Voxel-Based Dose Prediction and Clinical Optimization Strategies
    Zhong, Y.
    Wang, J.
    Hu, W.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [48] Clinical Implementation of Automated Treatment Planning for Rectum Intensity-Modulated Radiotherapy Using Voxel-Based Dose Prediction and Post-Optimization Strategies
    Zhong, Yang
    Yu, Lei
    Zhao, Jun
    Fang, Yingtao
    Yang, Yanju
    Wu, Zhiqiang
    Wang, Jiazhou
    Hu, Weigang
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [49] Determination of optimal number of beams in direct machine parameter optimization-based intensity modulated radiotherapy for head and neck cases
    Ranganathan, Vaitheeswaran
    Das, K. J. Maria
    JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 129 - 134
  • [50] Using neoadjuvant chemotherapy and replanning intensity-modulated radiotherapy for nasopharyngeal carcinoma with intracranial invasion to protect critical normal tissue
    Niu, Xiaoshuang
    Chang, Xi
    Gao, Yunsheng
    Hu, Chaosu
    Kong, Lin
    RADIATION ONCOLOGY, 2013, 8