Implementation of compensator-based intensity modulated radiotherapy with a conventional telecobalt machine using missing tissue approach

被引:0
|
作者
Tagoe, Samuel N. A. [1 ,2 ]
Mensah, Samuel Y. [2 ]
Fletcher, John J. [3 ]
机构
[1] Korle Bu Teaching Hosp, Natl Ctr Radiotherapy & Nucl Med, Accra, Ghana
[2] Univ Cape Coast, Coll Agr & Nat Sci, Sch Phys Sci, Dept Phys, Cape Coast, Ghana
[3] Univ Dev Studies, Dept Appl Phys, Navrongo Campus, Navrongo, Ghana
来源
关键词
telecobalt machine; bolus; compensator; Gafchromic film; intensity modulated;
D O I
10.2478/pjmpe-2018-0024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The present study aimed to generate intensity-modulated beams with compensators for a conventional telecobalt machine, based on dose distributions generated with a treatment planning system (TPS) performing forward planning, and cannot directly simulate a compensator. Materials and Methods: The following materials were selected for compensator construction: Brass, Copper and Perspex (PMMA). Boluses with varying thicknesses across the surface of a tissue-equivalent phantom were used to achieve beam intensity modulations during treatment planning with the TPS. Beam data measured for specific treatment parameters in a full scatter water phantom with a 0.125 cc cylindrical ionization chamber, with a particular compensator material in the path of beams from the telecobalt machine, and that without the compensator but the heights of water above the detector adjusted to get the same detector readings as before, were used to develop and propose a semiempirical equation for converting a bolus thickness to compensator material thickness, such that any point within the phantom would receive the planned dose. Once the dimensions of a compensator had been determined, the compensator was constructed using the cubic pile method. The treatment plans generated with the TPS were replicated on the telecobalt machine with a bolus within each beam represented with its corresponding compensator mounted on the accessory holder of the telecobalt machine. Results: Dose distributions measured in the tissue-equivalent phantom with calibrated Gafchromic EBT2 films for compensators constructed based on the proposed approach, were comparable to those of the TPS with deviation less than or equal to +/- 3% (mean of 2.29 +/- 0.61%) of the measured doses, with resultant confidence limit value of 3.21. Conclusion: The use of the proposed approach for clinical application is recommended, and could facilitate the generation of intensity-modulated beams with limited resources using the missing tissue approach rendering encouraging results.
引用
下载
收藏
页码:171 / 179
页数:9
相关论文
共 50 条
  • [21] Implementation of Machine Learning-Based Treatment Planning Tool for Whole Breast Radiotherapy Using Irregular Surface Compensator Technique
    Yoo, S.
    Sheng, Y.
    Blitzblau, R. C.
    Suneja, G.
    O'Neill, L.
    Morrison, J.
    Catalano, S.
    Yin, F. F.
    Wu, Q. J. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : S94 - S94
  • [22] A deep-learning-based dose verification tool utilizing fluence maps for a cobalt-60 compensator-based intensity-modulated radiation therapy system
    Oh, Kyuhak
    Gronberg, Mary P.
    Netherton, Tucker J.
    Sengupta, Bishwambhar
    Cardenas, Carlos E.
    Court, Laurence E.
    Ford, Eric C.
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 26
  • [23] Compensator Based Intensity Modulated Radiation Therapy Dosimetry Using EDR2 Film
    Srivastava, R.
    De Wagter, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [24] Clinical implementation of adaptive helical tomotherapy:: A unique approach to image-guided intensity modulated radiotherapy
    Welsh, James S.
    Lock, Michael
    Harari, Paul M.
    Tome, Wolfgang A.
    Fowler, Jack
    Mackie, Thomas Rockwell
    Ritter, Mark
    Kapatoes, Jeff
    Forrest, Lisa
    Chappell, Richard
    Paliwal, Bhudatt
    Mehta, Minesh P.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2006, 5 (05) : 465 - 479
  • [25] Long-Term Breast Cancer Patient Outcomes After Adjuvant Radiotherapy Using Intensity-Modulated Radiotherapy or Conventional Tangential Radiotherapy
    Yang, Jen-Fu
    Lee, Meei-Shyuan
    Lin, Chun-Shu
    Chao, Hsing-Lung
    Chen, Chang-Ming
    Lo, Cheng-Hsiang
    Fan, Chao-Yueh
    Tsao, Chih-Cheng
    Huang, Wen-Yen
    MEDICINE, 2016, 95 (11)
  • [26] Clinical Knowledge-Based Estimation of Target Position Displacements Using Machine Learning Architectures in Prostate Intensity Modulated Radiotherapy
    Kai, Y.
    Arimura, H.
    Ninomiya, K.
    Maruyama, M.
    Nakaguchi, Y.
    Shimohigashi, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E550 - E550
  • [27] A shift from simple to sophisticated: using intensity-modulated radiation therapy in conventional nonstereotactic palliative radiotherapy
    Dennis, Kristopher
    Linden, Kelly
    Gaudet, Marc
    CURRENT OPINION IN SUPPORTIVE AND PALLIATIVE CARE, 2023, 17 (01) : 70 - 76
  • [28] Assessment of treatment accuracy for head and neck intensity modulated radiotherapy (IMRT) and conventional radiotherapy using amorphous silicon electronic portal imaging
    Humphreys, M
    Urbano, TG
    Miles, E
    Mubata, C
    Nutting, C
    Harrington, K
    Bidmead, M
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S65 - S65
  • [29] Prediction of Radiomics-Based Machine Learning for Specific Dosimetric Verification of Pelvic Intensity Modulated Radiotherapy
    Ni, Q.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E479 - E480
  • [30] Hypofractionated versus conventional intensity modulated postmastectomy radiotherapy: Toxicity and quality of life in patients with tissue-expander breast reconstruction
    Farina, Ariel
    Alfonso, Dagmar
    Fernandez, Aroldo
    Carvajal, Felipe
    Maria Ciudad, Ana
    Martinez, Celmira
    Perez, Camila
    Gonzalez, Pablo
    Marsiglia, Hugo
    CANCER RESEARCH, 2021, 81 (04)